Skip to content

AU/DxC AU Chemistry Transferrin REF: OSR6152 recallZ-0309-2024 · Beckman Coulter Biomedical Division · initiated Sep 6, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Biomedical Division or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The firm has identified that the Transferrin reagent does not meet the Instructions for Use (IFU) reagent on-board stability claims, which may cause control failures and/or erroneous patient samples results. The Complaint handling unit has received a total of 2 safety cases (CASE-2023-02027212 and CASE-2023-02045278) and 99 quality cases with the failure mode attributed to Transferrin reagent (OSR6152 / lot 2573) associated with reagent blank shifts for the Reagent 2 (R2), imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023. Reagent Blank shifts for R2, imprecision, calibration failures, erratic Quality Control (QC) recoveries, and/or erroneous high/low (serum, plasma, urine) patient results. The issue was confirmed through internal testing by Beckman Coulter on June 30, 2023.

FDA root cause category: Nonconforming Material/Component

What the company is doing

On 9/6/23, recall notices were mailed and emailed to customers who were asked to do the following: 1) Discontinue use of affected transferrin. 2) Contact firm for replacement lots. 3) No retrospective review of past patient results is required but this may be carried out at the discretion of the Laboratory Director. 4) Share this information with laboratory staff and retain the notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, provide them a copy of this letter. 5) Complete and return the response form via email to Email: [email protected] Questions can be directed to the Customer Support Center via website: http://www.beckmancoulter.com or phone: 1-800-854-3633 in the United States.

Recall details

Recall number
Z-0309-2024
Status
Open, Classified
Initiated by the company
Sep 6, 2023
Posted by the FDA
Nov 16, 2023
Recalling firm
Beckman Coulter Biomedical Division
Quantity
2,754
Product code
DDD
FDA submissions
K963139
Distribution
Worldwide - US Nationwide distribution including in the states of NY, CA, MS, GA, PA, SC, WA, TN, AL, AK, OR, MO, KS, FL, LA, OH, TX, NJ, MI, WV, MD, IL, RI, VA, NC, IN, OK, NV, NE, WY, KY, WI, ND, MA, NH, UT, CO, GU, PR, AR, AZ, ME and the countries of Canada, Netherlands, United Kingdom, Sweden, Germany, France, Switzerland, Tunisia, Australia, Belgium, Costa Rica, Uruguay, Turkey, South Africa, New Zealand, Italy, Thailand, Ghana, Russia, Hungary, Czech Republic, Slovakia, India, Austria, Poland, Romania, Ireland, Spain, Portugal, Mexico, Vietnam, Korea, Mozambique, Kazakhstan, China, Malaysia, Singapore, Kenya, Brazil, Namibia, Egypt, Panama.

Product description

AU/DxC AU Chemistry Transferrin REF: OSR6152 Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. Turbidimeters measure the reduction of incident light due to reflection, absorption, or scatter. In the procedure, the measurement of the decrease in light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction, is the basis of this assay.

Code information (models, lots, serial numbers)

UDI-DI: 15099590010492, LOT: 2573, Expiration: 01 August 2024

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page