Skip to content

EVIS EXERA III GASTROINTESTINAL VIDEOSCOPE recallZ-0288-2024 · Olympus Corporation of the Americas · initiated Oct 2, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.

FDA root cause category: Reprocessing Controls

What the company is doing

The firm distributed recall notifications via mail to consignees dated 10/02/2023. Consignees are asked to inspect inventory for affected devices, ensure that all reprocessing personnel are aware and knowledgeable of the recall notice and its contents, and acknowledge receipt of the recall notification by returning the enclosed response form via email to [email protected] or by fax at 800-913-4345. Customers with any questions about the acknowledgement form are to call 866-912-9544. Customers with any questions about the recall are to contact Cynthia Ow by phone at 647-999-3203 or by email at [email protected].

Recall details

Recall number
Z-0288-2024
Status
Open, Classified
Initiated by the company
Oct 2, 2023
Posted by the FDA
Nov 9, 2023
Recalling firm
Olympus Corporation of the Americas
Quantity
1687 units
Product code
FDF
FDA submissions
K112680
Distribution
Nationwide Distribution.

Product description

EVIS EXERA III Gastrointestinal Videoscope

Code information (models, lots, serial numbers)

Model No. GIF-1TH190; UDI-DI: 04953170343360; All Serial Numbers.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page