Spetzler-Malis recallZ-0250-2024 · Stryker Corporation · initiated Sep 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An error was identified on the expiration date of the product label. The shelf life for the products impacted is 36 months (3 years) however the product label represents a 54 month (4.5 year) shelf life.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT Medical Device Recall notification letter was sent to customers on 9/11/23. Actions to be taken 1. Immediately check your inventory to locate and quarantine any affected products found at your facility. 2. Sign and return the enclosed Business Response Form (BRF) by email to [email protected], even if affected product is no longer in inventory. 3. Upon receipt of the completed BRF, a shipping label will be provided to return any recalled product on-hand and Stryker will contact you to arrange for replacement of your product(s). 4. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. Please contact Jordan Anderson at (269) 800-1347 or [email protected] with questions or concerns. For reporting any serious adverse events or product quality problems contact Stryker at 1-800-253-3210, ext. 5555 8:00 A.M. to 7:00 P.M. (Eastern Time), Monday Friday or by email at [email protected].
Recall details
- Recall number
- Z-0250-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 11, 2023
- Posted by the FDA
- Nov 7, 2023
- Recalling firm
- Stryker Corporation
- Quantity
- 2,321 total units
- Product code
- GEI
- FDA submissions
- K121426
- Distribution
- US Nationwide. Australia, Canada, Japan, Korea, Singapore.
Product description
Disposable Spetzler-Malis Bipolar forceps, 23CM IM DISP BAYONET 0.5MM TIP, Catalog Number 6770-230-005
Code information (models, lots, serial numbers)
GTIN: 7613327300161; Lot Numbers: 200515009, 201008013, 201015011, 201126013
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Spetzler-Malis 6770-230-005Stryker Corporation · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES1 recall1 report
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93211).
- Z-0239-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-180-005
- Z-0240-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 18CM SM DISP BAYONET 1.5MM TIP, Catalog Number 6760-180-015
- Z-0241-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-200-005
- Z-0242-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 0.5MM TIP, Catalog Number 6760-230-005
- Z-0243-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.0MM TIP, Catalog Number 6760-230-010
- Z-0244-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SM DISP BAYONET 1.5MM TIP, Catalog Number 6760-230-015
- Z-0245-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 18CM IM DISP BAYONET 0.5MM TIP, Catalog Number 6770-180-005
- Z-0246-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 18CM IM DISP BAYONET 1.0MM TIP, Catalog Number 6770-180-010
- Z-0247-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 18CM IM DISP BAYONET 1.5MM TIP, Catalog Number 6770-180-015
- Z-0248-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM IM DISP BAYONET 0.5MM TIP, Catalog Number 6770-200-005
- Z-0249-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM IM DISP BAYONET 1.5MM TIP, Catalog Number 6770-200-015
- Z-0251-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM IM DISP BAYONET 1.0MM TIP, Catalog Number 6770-230-010
- Z-0252-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM SLM DISP BAYONET 0.5MM, Catalog Number 6780-200-005
- Z-0253-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM SLM DISP BAYONET 1.0MM, Catalog Number 6780-200-010
- Z-0254-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 20CM SLM DISP BAYONET 1.5MM, Catalog Number 6780-200-015
- Z-0255-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SLM DISP BAYONET 0.5MM, Catalog Number 6780-230-005
- Z-0256-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SLM DISP BAYONET 1.0MM, Catalog Number 6780-230-010
- Z-0257-2024 · Open, ClassifiedDisposable Spetzler-Malis Bipolar forceps, 23CM SLM DISP BAYONET 1.5MM, Catalog Number 6780-230-015
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page