RAPIDPOINT 500e Blood Gas System recallZ-0422-2024 · Siemens Healthcare Diagnostics Inc · initiated Sep 26, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Siemens Healthcare Diagnostics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There are potential drug interferences from Perhexiline Maleate or Atomoxetine Hydrochloride that may cause falsely elevated sodium results reported on the RAPIDPoint 500 and RAPIDPoint 500e Blood Gas Systems. The erroneous sodium levels with positive bias may lead to unrecognized hyponatremia and/or the inappropriate treatment of hypernatremia may lead to iatrogenic hyponatremia, resulted in a delay in patient diagnosis and optimal patient management.
FDA root cause category: Under Investigation by firm
What the company is doing
The firm issued an Urgent Medical Device Correction notice on 9/26/2023 to customers asking that customers review the notice with the facility's Medical Director to determine the appropriate course of action for previously generated results, complete and return the Field Correction Effectiveness Check Form within 30 days, forward this notice to customers if your firm is a distributor, and retain the notification in your system's Operator's Guide and forward to those who may have received product. The Field Correction Effectiveness Check Form is to be returned via email to [email protected] or by fax to 312-275-7795. Questions are to be directed to your local Siemens Healthineers Technical Support Representative.
Recall details
- Recall number
- Z-0422-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 26, 2023
- Posted by the FDA
- Dec 1, 2023
- Recalling firm
- Siemens Healthcare Diagnostics Inc
- Quantity
- 8,749 units
- Product code
- JGS
- FDA submissions
- K192240
- Distribution
- Domestic: Nationwide Distribution. Foreign: Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Colombia, Congo, Republic, Costa Rica, Cote dIvoire, Croatia, Curacao, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Fiji, Finland, France, French, Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Islamic Republic of, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Latvia, Lebanon, Lesotho, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mexico, Moldova, Montenegro, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Palestinian Territory, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic Korea, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom Great Britain, Uruguay, & Vietnam.
Product description
RAPIDPOINT 500e Blood Gas System
Code information (models, lots, serial numbers)
Material No. 11416751, 11416752, 11416754, & 11416755; UDI-DI: 00630414286150, 00630414286167, 00630414286174, & 00630414286143; All Units and Software Versions.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- RAPIDPoint® 500e Blood Gas System 11416751Siemens Healthcare Diagnostics Inc. · ELECTRODE, ION SPECIFIC, CALCIUM2 recalls7 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93240).
- Z-0423-2024 · Open, ClassifiedRAPIDPOINT 500 Blood Gas System
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page