DRX-Revolution Mobile X-ray System recallZ-0289-2024 · Carestream Health, Inc. · initiated Oct 2, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Carestream Health, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a potential for unexpected failure of the electrical components within the Carestream Health Inc. generator installed in your DRX-Revolution Mobile X-ray System(s). This can lead to temporary and self contained thermal overload within the generator. Should an event like this occur, the system will become inoperable and loud noise, burnt smell and smoke may be detectable.
FDA root cause category: Component design/selection
What the company is doing
An MEDICAL DEVICE FIELD CORRECTIVE ACTION notification letter dated 9/27/23 was sent to customers. Actions To Be Taken by User Facility: " Inform all personnel that utilize the system of the potential issue. " Should an event of this nature occur: o Move the system away from patients/staff. o The system will be inoperable. o Call the Carestream Customer Care Center in the U.S. at 1-800-328-2910, your local Carestream Service support number or Carestream Health Authorized Representative to place a service call. " If an Adverse Health Event is encountered, notify FDA MedWatch o MedWatch: Use the MedWatch Online Reporting Form to report adverse events or call 800-332-1088. o If outside the US, notify the local competent authority in your region. " Review and complete the Consignee Notification Acknowledgement form as instructed. Corrective Action to Be Taken by Carestream: You will be contacted by a Carestream Service Engineer or Carestream Health Authorized Representative to schedule a time for your system(s) to be serviced. At the scheduled time, a Carestream Service Engineer or Carestream Health Authorized Representative will inspect your DRX-Revolution Mobile X-ray System and replace generator component(s) which will resolve the identified Carestream generator issue. If you have any questions or concerns please contact the Carestream Customer Care Center in the U.S. at 1-800-328-2910, available 7 days per week on a 24-hour basis. Outside of the U.S., please call your local Carestream Service support number. If you have distributed the device outside your facility, please alert your customer(s) of this Field Corrective Action and contact the Carestream Customer Care Center as instructed above.
Recall details
- Recall number
- Z-0289-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 2, 2023
- Posted by the FDA
- Nov 9, 2023
- Recalling firm
- Carestream Health, Inc.
- Quantity
- 2402 systems
- Product code
- IZL
- FDA submissions
- K191025
- Distribution
- US Nationwide. Global Distribution.
Product description
The DRX-Revolution Mobile X-Ray System is a diagnostic mobile x-ray system utilizing digital radiography (DR) technology. the system consists of a self-contained x-ray generator, image receptor(s), imaging display and software for acquiring medical diagnostic images outside of standard stationary x-ray room. The DRX-Revolution system incorporates a flat-panel detector that can be used wirelessly for exams such as in-bed chest projections. The system can also be used to expose CR phosphor screens or films.
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
UDI-DI: 60889971019394, Serial Numbers: 1243, 3231, 3232, 3233, 3245, 3246, 3247, 3248, 3249, 3250, 3251, 3252, 3254, 3257, 3260, 3263, 3264, 3265, 3266, 3267, 3270, 3272, 3277, 3285, 3312, 3313, 3317, 3318, 3320, 3321, 3322, 3323, 3324, 3328, 3334, 3465, 3470, 3477, 3489, 3490, 3495, 3496, 3502, 3503, 3506, 3509, 3510, 3511, 3512, 3518, 3521, 3543, 3546, 3548, 3549, 3551, 3552, 3556, 3557, 3563, 3565, 3566, 3572, 3578, 3586, 3587, 3588, 3610, 3611, 3612, 3614, 3617, 3645, 3649, 3650, 3657, 3658, 3660, 3663, 3664, 3671, 3672, 3673, 3686, 3700, 3701, 3703, 3704, 3705, 3706, 3707, 3708, 3709, 3710, 3711, 3712, 3713, 3714, 3715, 3716, 3717, 3721, 3724, 3725, 3730, 3745, 3749, 3750, 3751, 3752, 3753, 3760, 3761, 3762, 3763, 3764, 3766, 3772, 3773, 3776, 3777, 3782, 3783, 3784, 3785, 3793, 3794, 3795, 3798, 3799, 3801, 3802, 3809, 3810, 3814, 3815, 3821, 3826, 3827, 3828, 3829, 3830, 3831, 3833, 3834, 3860, 3864, 3867, 3872, 3875, 3876, 3877, 3878, 3888, 3897, 3898, 3909, 3910, 3911, 3912, 3925, 3937, 3950, 3953, 3961, 3964, 3969, 3974, 3993, 3994, 3995, 3996, 4012, 4030, 4213, 4735, 7616, 818, 3981, 3997, 3998, 4007, 4013, 4014, 4015, 4017, 4021, 4029, 4035, 4045, 4062, 4063, 4070, 4071, 4072, 4073, 4089, 4090, 4097, 4102, 4103, 4116, 4131, 4132, 4133, 4134, 4135, 4136, 4137, 4138, 4139, 4140, 4141, 4142, 4143, 4144, 4145, 4146, 4147, 4148, 4149, 4151, 4152, 4153, 4160, 4161, 4162, 4163, 4164, 4165, 4166, 4167, 4168, 4169, 4171, 4172, 4173, 4175, 4176, 4177, 4178, 4179, 4180, 4181, 4182, 4183, 4184, 4185, 4186, 4187, 4188, 4189, 4190, 4191, 4192, 4193, 4194, 4195, 4196, 4197, 4198, 4199, 4200, 4201, 4202, 4203, 4204, 4205, 4206, 4208, 4210, 4211, 4212, 4214, 4215, 4217, 4218, 4219, 4220, 4223, 4224, 4225, 4226, 4227, 4228, 4229, 4230, 4231, 4234, 4235, 4236, 4237, 4240, 4241, 4242, 4243, 4244, 4245, 4246, 4247, 4248, 4249, 4250, 4251, 4252, 4253, 4254, 4255, 4256, 4257, 4259, 4260, 4262, 4263, 4267, 4268, 4269, 4270, 4271, 4272, 4275, 4276, 4277, 4278, 4279, 4280, 4281, 4320, 4321, 4322, 4323, 4324, 4326, 4327, 4328, 4329, 4405, 4424, 4445, 4446, 4447, 4448, 4449, 4556, 4557, 4558, 4559, 4560, 4561, 4562, 4563, 4564, 4565, 4566, 4567, 4568, 4569, 4570, 4571, 4572, 4573, 4574, 4575, 4576, 4577, 4578, 4579, 4580, 4581, 4582, 4583, 4584, 4585, 4586, 4587, 4588, 4589, 4590, 4591, 4592, 4593, 4594, 4595, 4596, 4597, 4598, 4599, 4600, 4601, 4602, 4603, 4604, 4605, 4606, 4607, 4608, 4609, 4610, 4611, 4612, 4613, 4614, 4615, 4616, 4617, 4618, 4619, 4620, 4621, 4622, 4623, 4624, 4625, 4626, 4627, 4628, 4629, 4630, 4631, 4632, 4633, 4634, 4635, 4636, 4637, 4638, 4639, 4640, 4641, 4642, 4643, 4644, 4645, 4646, 4647, 4648, 4649, 4650, 4651, 4652, 4653, 4654, 4655, 4656, 4657, 4658, 4659, 4660, 4661, 4662, 4663, 4664, 4665, 4666, 4667, 4668, 4669, 4670, 4671, 4672, 4673, 4674, 4675, 4676, 4677, 4678, 4679, 4680, 4681, 4682, 4683, 4684, 4685, 4686, 4687, 4688, 4689, 4690, 4691, 4692, 4694, 4695, 4696, 4697, 4698, 4699, 4700, 4701, 4702, 4703, 4704, 4705, 4706, 4707, 4708, 4709, 4713, 4714, 4715, 4716, 4717, 4718, 4719, 4720, 4721, 4722, 4723, 4724, 4725, 4726, 4728, 4729, 4730, 4736, 4737, 4738, 4739, 4740, 4741, 4742, 4743, 4744, 4745, 4746, 4749, 4750, 4751, 4752, 4753, 4755, 4756, 4757, 4758, 4759, 4760, 4761, 4762, 4763, 4764, 4765, 4766, 4767, 4768, 4769, 4770, 4771, 4772, 4773, 4774, 4775, 4776, 4777, 4778, 4779, 4780, 4781, 4782, 4783, 4790, 4791, 4792, 4793, 4794, 4797, 4798, 4800, 4801, 4802, 4803, 4804, 4805, 4806, 4807, 4808, 4809, 4810, 4811, 4812, 4813, 4814, 4815, 4816, 4817, 4818, 4819, 4820, 4821, 4823, 4824, 4825, 4829, 4832, 4833, 4834, 4835, 4836, 4837, 4838, 4839, 4840, 4841, 4842, 4843, 4844, 4845, 4846, 4847, 4848, 4849, 4850, 4851, 4852, 4853, 4854, 4855, 4856, 4857, 4859, 4865, 4866, 4867, 4868, 4869, 4870, 4871, 4872, 4873, 4874, 4875, 4876, 4878, 4879, 4880, 4881, 4882, 4883, 4884, 4886, 7361, 7378, 7379, 7380, 7384, 7385, 8541, 3115, 3116, 3117, 3180, 3236, 3237, 3238, 3239, 3240, 3241, 3242, 3243, 3244, 3253, 3255, 3256, 3258, 3261, 3262, 3268, 3269, 3271, 3273, 3274, 3275, 3276, 3278, 3281, 3283, 3286, 3287, 3288, 3289, 3290, 3291, 3292, 3295, 3296, 3297, 3298, 3299, 3300, 3301, 3302, 3303, 3304, 3305, 3306, 3307, 3308, 3309, 3310, 3314, 3315, 3316, 3319, 3330, 3331, 3333, 3335, 3466, 3467, 3468, 3469, 3471, 3472, 3473, 3474, 3475, 3476, 3478, 3479, 3484, 3485, 3491, 3492, 3493, 3494, 3497, 3498, 3499, 3504, 3505, 3513, 3514, 3515, 3516, 3517, 3519, 3522, 3523, 3524, 3525, 3527, 3528, 3529, 3530, 3539, 3547, 3550, 3553, 3554, 3555, 3558, 3559, 3560, 3561, 3567, 3568, 3569, 3570, 3571, 3573, 3574, 3575, 3577, 3579, 3580, 3581, 3582, 3583, 3584, 3585, 3589, 3590, 3591, 3592, 3593, 3594, 3595, 3596, 3597, 3598, 3599, 3600, 3601, 3602, 3603, 3604, 3605, 3606, 3607, 3613, 3615, 3616, 3618, 3619, 3620, 3621, 3622, 3623, 3624, 3625, 3626, 3627, 3628, 3629, 3630, 3631, 3632, 3633, 3634, 3635, 3636, 3637, 3638, 3639, 3640, 3641, 3642, 3643, 3644, 3646, 3647, 3648, 3651, 3652, 3653, 3654, 3655, 3656, 3659, 3665, 3666, 3667, 3668, 3669, 3670, 3674, 3675, 3676, 3677, 3678, 3679, 3680, 3681, 3682, 3683, 3684, 3685, 3687, 3688, 3689, 3690, 3691, 3692, 3693, 3694, 3695, 3696, 3698, 3699, 3722, 3723, 3726, 3727, 3728, 3729, 3731, 3732, 3733, 3734, 3735, 3736, 3737, 3738, 3739, 3740, 3741, 3742, 3743, 3744, 3754, 3755, 3757, 3758, 3759, 3765, 3767, 3769, 3770, 3771, 3778, 3779, 3780, 3786, 3787, 3788, 3789, 3790, 3791, 3792, 3796, 3797, 3800, 3803, 3804, 3805, 3806, 3807, 3808, 3811, 3812, 3817, 3818, 3819, 3820, 3822, 3823, 3824, 3825, 3837, 3838, 3839, 3840, 3841, 3842, 3843, 3844, 3845, 3846, 3847, 3848, 3849, 3850, 3851, 3852, 3853, 3854, 3855, 3856, 3857, 3858, 3859, 3865, 3866, 3868, 3869, 3870, 3871, 3873, 3874, 3879, 3884, 3885, 3886, 3887, 3889, 3890, 3891, 3892, 3893, 3894, 3895, 3896, 3899, 3900, 3901, 3902, 3903, 3904, 3905, 3906, 3907, 3913, 3914, 3915, 3916, 3917, 3918, 3919, 3920, 3921, 3922, 3923, 3924, 3926, 3927, 3928, 3929, 3930, 3931, 3932, 3933, 3934, 3935, 3936, 3938, 3939, 3940, 3941, 3942, 3943, 3944, 3947, 3948, 3949, 3952, 3954, 3955, 3956, 3957, 3958, 3959, 3960, 3965, 3966, 3967, 3968, 3970, 3971, 3972, 3973, 3975, 3976, 3977, 3983, 3984, 3990, 3992, 4006, 4009, 4010, 4027, 4031, 4032, 4034, 4041, 4042, 4043, 4044, 7344, 7443, 8575, 197, 1103, 1684, 1728, 1924, 1925, 2451, 3363, 3978, 3979, 3980, 3982, 3985, 3986, 3987, 3988, 3989, 3991, 3999, 4000, 4001, 4002, 4003, 4004, 4005, 4011, 4018, 4019, 4020, 4022, 4024, 4025, 4026, 4028, 4033, 4036, 4037, 4038, 4039, 4040, 4047, 4048, 4049, 4050, 4051, 4052, 4053, 4054, 4055, 4056, 4057, 4059, 4060, 4061, 4064, 4065, 4066, 4067, 4068, 4069, 4074, 4075, 4077, 4078, 4079, 4080, 4081, 4082, 4083, 4085, 4086, 4088, 4091, 4092, 4093, 4094, 4095, 4096, 4098, 4099, 4100, 4101, 4104, 4106, 4107, 4108, 4109, 4110, 4111, 4112, 4113, 4114, 4120, 4121, 4122, 4123, 4124, 4125, 4126, 4127, 4128, 4129, 4130, 4154, 4155, 4156, 4157, 4158, 4159, 4216, 4221, 4222, 4232, 4233, 4238, 4239, 4258, 4264, 4265, 4266, 4273, 4274, 4282, 4283, 4284, 4285, 4286, 4287, 4288, 4289, 4290, 4291, 4292, 4293, 4294, 4295, 4296, 4297, 4298, 4299, 4300, 4301, 4302, 4303, 4304, 4305, 4306, 4307, 4308, 4309, 4310, 4311, 4312, 4313, 4314, 4315, 4316, 4317, 4318, 4319, 4330, 4331, 4332, 4333, 4334, 4335, 4337, 4338, 4339, 4340, 4341, 4342, 4343, 4344, 4345, 4346, 4347, 4348, 4349, 4350, 4351, 4352, 4353, 4354, 4355, 4356, 4357, 4358, 4359, 4360, 4361, 4362, 4363, 4364, 4365, 4366, 4367, 4368, 4369, 4370, 4371, 4372, 4373, 4374, 4375, 4376, 4377, 4378, 4379, 4380, 4381, 4382, 4383, 4384, 4385, 4386, 4387, 4388, 4389, 4390, 4391, 4392, 4393, 4394, 4395, 4396, 4397, 4398, 4399, 4400, 4401, 4402, 4403, 4406, 4407, 4408, 4409, 4410, 4411, 4412, 4413, 4414, 4415, 4416, 4417, 4418, 4419, 4420, 4421, 4422, 4423, 4425, 4426, 4427, 4428, 4429, 4430, 4431, 4432, 4433, 4434, 4435, 4436, 4437, 4438, 4439, 4440, 4441, 4442, 4443, 4444, 4450, 4451, 4452, 4453, 4454, 4455, 4456, 4457, 4458, 4460, 4461, 4462, 4463, 4464, 4465, 4466, 4467.4468, 4469, 4470, 4471, 4472, 4473, 4474, 4475, 4476, 4477, 4478, 4479, 4480, 4481, 4482, 4483, 4484, 4485, 4486, 4487, 4488, 4489, 4490, 4491, 4492, 4493, 4494, 4495, 4496, 4497, 4498, 4499, 4500, 4501, 4502, 4503, 4504, 4505, 4506, 4507, 4508, 4509, 4510, 4511, 4512, 4513, 4514, 4515, 4516, 4517, 4518, 4519, 4520, 4521, 4522, 4523, 4524, 4525, 4526, 4527, 4529, 4530, 4531, 4532, 4533, 4534, 4535, 4536, 4537, 4538, 4539, 4540, 4541, 4542, 4543, 4544, 4545, 4546, 4547, 4548, 4549, 4550, 4551, 4552, 4553, 4554, 4555, 4727, 4747, 4748, 4754, 4784, 4785, 4786, 4787, 4788, 4789, 4795, 4799, 4827, 4830, 4831, 7067, 7079, 800672; UDI-DI: 60889978618897, Serial Numbers: 7001, 7016, 7021, 7031, 7032, 7054, 7060, 7075, 7084, 7085, 7093, 7094, 7095, 7114, 7115, 7116, 7118, 7119, 7143, 7144, 7145, 7146, 7147, 7148, 7149, 7150, 7152, 7153, 7154, 7155, 7156, 7157, 7158, 7160, 7161, 7167, 7168, 7169, 7170, 7171, 7172, 7173, 7174, 7180, 7181, 7182, 7184, 7185, 7186, 7203, 7204, 7208, 7209, 7223, 7224, 7225, 7233, 7234, 7235, 7236, 7237, 7243, 7244, 7245, 7246, 7247, 7248, 7249, 7250, 7251, 7268, 7269, 7270, 7281, 7282, 7283, 7285, 7286, 7287, 7292, 7294, 7299, 7301, 7303, 7304, 7312, 7313, 7314, 7315, 7319, 7320, 7321, 7322, 7323, 7330, 7331, 7355, 7356, 7357, 7358, 7359, 7360, 7362, 7363, 7365, 7366, 7367, 7368, 7372, 7381, 7383, 7392, 7393, 7405, 7428, 7431, 7437, 7446, 7447, 7448, 7463, 7464, 7491, 7505, 7506, 7514, 7517, 7518, 7534, 7535, 7536, 7537, 7538, 7539, 7540, 7541, 7542, 7543, 7544, 7545, 7546, 7547, 7548, 7554, 7560, 7561, 7562, 7563, 7572, 7573, 7574, 7575, 7576, 7577, 7578, 7579, 7580, 7581, 7582, 7583, 7585, 7586, 7588, 7589, 7590, 7591, 7592, 7594, 7595, 7597, 7600, 7602, 7603, 7604, 7605, 7606, 7607, 7608, 7609, 7610, 7611, 7623, 7624, 7625, 7629, 7634, 7643, 7644, 7650, 7651, 7665, 7666, 7667, 7668, 7669, 7673, 7686, 7687, 7688, 7689, 7691, 7692, 7693, 7694, 7714, 7722, 7723, 7724, 7725, 7727, 7728, 7729, 7763, 7764, 7765, 7766, 7778, 7779, 7780, 7781, 7783, 7784, 7785, 7786, 7787, 7788, 7789, 7790, 7791, 7792, 7796, 7801, 7812, 7829, 7830, 7831, 7832, 7833, 7834, 7835, 7848, 7849, 7850, 7851, 7868, 7869, 7870, 7885, 7886, 7889, 7890, 7891, 7901, 7902, 7904, 7907, 7909, 7910, 7911, 7917, 7919, 7921, 7937, 7938, 7946, 7947, 7952, 7953, 7954, 7967, 7969, 7974, 7975, 7976, 7977, 7996, 8005, 8006, 8007, 8008, 8009, 8010, 8017, 8018, 8021, 8059, 8071, 8089, 8102, 8117, 20003, 20004, 20005, 20006, 20007, 20008, 20009, 20010, 20011, 20012, 20013, 20014, 20016, 20019, 20020, 20021, 20023, 20025, 20026, 20027, 20029, 20031, 20032, 20034, 20037, 20038, 20042, 20044, 20048, 20049, 20050, 7006, 7007, 7009, 7010, 7011, 7012, 7013, 7014, 7015, 7017, 7018, 7019, 7020, 7022, 7023, 7024, 7025, 7026, 7027, 7028, 7029, 7030, 7033, 7034, 7035, 7036, 7037, 7038, 7039, 7040, 7041, 7042, 7043, 7044, 7045, 7046, 7047, 7048, 7049, 7050, 7051, 7052, 7053, 7055, 7056, 7057, 7058, 7059, 7061, 7062, 7063, 7064, 7065, 7066, 7068, 7069, 7070, 7071, 7072, 7073, 7074, 7076, 7077, 7078, 7080, 7081, 7082, 7083, 7086, 7087, 7088, 7089, 7090, 7091, 7092, 7096, 7097, 7098, 7099, 7100, 7101, 7102, 7103, 7104, 7105, 7106, 7107, 7108, 7109, 7110, 7111, 7112, 7113, 7117, 7122, 7123, 7124, 7125, 7126, 7127, 7128, 7129, 7130, 7131, 7132, 7133, 7134, 7135, 7136, 7137, 7138, 7139, 7140, 7141, 7142, 7151, 7159, 7162, 7163, 7164, 7165, 7166, 7175, 7176, 7177, 7178, 7187, 7188, 7189, 7190, 7191, 7192, 7193, 7194, 7195, 7196, 7197, 7198, 7199, 7200, 7201, 7202, 7210, 7211, 7212, 7213, 7214, 7215, 7216, 7217, 7218, 7219, 7220, 7221, 7222, 7227, 7228, 7229, 7230, 7231, 7232, 7239, 7240, 7241, 7242, 7252, 7253, 7254, 7255, 7256, 7258, 7259, 7260, 7261, 7262, 7263, 7264, 7265, 7266, 7267, 7271, 7272, 7273, 7274, 7275, 7276, 7277, 7278, 7279, 7288, 7289, 7290, 7291, 7295, 7296, 7297, 7298, 7305, 7306, 7307, 7308, 7309, 7310, 7311, 7316, 7317, 7318, 7324, 7325, 7326, 7327, 7328, 7329, 7332, 7333, 7334, 7335, 7336, 7337, 7338, 7339, 7340, 7341, 7342, 7343, 7345, 7346, 7347, 7348, 7349, 7350, 7351, 7352, 7353, 7354, 7369, 7373, 7374, 7375, 7376, 7377, 7386, 7387, 7388, 7389, 7390, 7391, 7394, 7395, 7396, 7397, 7398, 7399, 7400, 7401, 7402, 7403, 7404, 7406, 7407, 7408, 7409, 7410, 7411, 7412, 7413, 7414, 7415, 7416, 7418, 7420, 7421, 7422, 7423, 7424, 7425, 7426, 7427, 7429, 7430, 7432, 7433, 7434, 7435, 7436, 7439, 7440, 7441, 7442, 7444, 7445, 7450, 7451, 7452, 7453, 7454, 7455, 7456, 7457, 7458, 7459, 7460, 7461, 7462, 7465, 7466, 7467, 7468, 7469, 7470, 7471, 7472, 7473, 7474, 7475, 7476, 7477, 7478, 7479, 7480, 7481, 7482, 7483, 7484, 7485, 7486, 7487, 7488, 7489, 7490, 7492, 7493, 7494, 7496, 7497, 7498, 7499, 7500, 7501, 7503, 7504, 7507, 7508, 7509, 7510, 7511, 7512, 7513, 7515, 7516, 7519, 7520, 7521, 7522, 7523, 7524, 7526, 7527, 7528, 7529, 7530, 7531, 7532, 7533, 7549, 7550, 7551, 7552, 7553, 7555, 7556, 7557, 7558, 7559, 7564, 7565, 7566, 7567, 7568, 7569, 7570, 7584, 7587, 7593, 7612, 7613, 7614, 7618, 7619, 7620, 7621, 7626, 7627, 7628, 7631, 7632, 7633, 7635, 7637, 7638, 7639, 7640, 7641, 7645, 7647, 7648, 7649, 7652, 7653, 7654, 7656, 7657, 7659, 7660, 7661, 7662, 7663, 7670, 7672, 7674, 7675, 7676, 7677, 7678, 7679, 7680, 7681, 7682, 7683, 7695, 7696, 7697, 7698, 7699, 7700, 7701, 7702, 7703, 7704, 7705, 7706, 7707, 7708, 7709, 7710, 7711, 7712, 7713, 7715, 7716, 7717, 7718, 7719, 7731, 7732, 7733, 7734, 7735, 7736, 7737, 7738, 7739, 7741, 7742, 7743, 7744, 7745, 7746, 7747, 7748, 7749, 7750, 7751, 7752, 7753, 7754, 7755, 7757, 7758, 7759, 7760, 7761, 7762, 7767, 7768, 7770, 7771, 7772, 7773, 7774, 7775, 7776, 7777, 7793, 7794, 7795, 7797, 7798, 7799, 7800, 7802, 7803, 7804, 7805, 7806, 7807, 7808, 7809, 7810, 7811, 7813, 7814, 7815, 7816, 7817, 7818, 7836, 7838, 7839, 7840, 7841, 7842, 7843, 7844, 7845, 7846, 7847, 7853, 7854, 7855, 7856, 7857, 7858, 7859, 7860, 7861, 7862, 7863, 7864, 7865, 7866, 7871, 7872, 7874, 7875, 7876, 7877, 7878, 7879, 7880, 7881, 7882, 7883, 7884, 7887, 7888, 7892, 7893, 7894, 7895, 7896, 7898, 7899, 7900, 7915, 7916, 7922, 7923, 7924, 7925, 7926, 7927, 7928, 7929, 7930, 7931, 7932, 7933, 7934, 7935, 7936, 7955, 7966, 7968, 7979, 7980, 7981, 7982, 7983, 7984, 7985, 7986, 7987, 7988, 7989, 7990, 7991, 7993, 7994, 7995, 7997, 7998, 7999, 8000, 8001, 8002, 8003, 8022, 8023, 8024, 8025, 8026, 8027, 8028, 8029, 8030, 8031, 8032, 8033, 8034, 8035, 8036, 8037, 8039, 8040, 8041, 8042, 8043, 8044, 8045, 8060, 8061, 8062, 8065, 8066, 8077, 8078, 8079, 8080, 8081, 8082, 8083, 8084, 8091, 8092, 8094, 8095, 8097, 8105
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile2 recalls12 reports
- DRX-Revolution Mobile X-ray SystemCarestream Health, Inc. · System, x-ray, mobile1 recall16 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.