Skip to content

EXALT Model D recallZ-0397-2024 · Boston Scientific Corporation · initiated Sep 27, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Poor image quality due to fluid ingress in the lens.

FDA root cause category: Process control

What the company is doing

Notification of Urgent Medical Device Product Removal was issued via FedEx on Oct 3, 2023. The notice advises customers to immediately discontinue use and segregate affected product, and to complete and return the Reply Verification and Tracking Form (RVTF) even if you do not have any affected product. Customers are to post the notice near any affected product and notify end users. If further distributed, notify any customers to which product has been further distributed to. Reach out to your local Boston Scientific representative with any questions. The notice is batch / lot specific. Boston Scientific has implemented corrective changes and will make replacement EXALT Model D scopes available as soon as possible.

Recall details

Recall number
Z-0397-2024
Status
Open, Classified
Initiated by the company
Sep 27, 2023
Posted by the FDA
Nov 21, 2023
Recalling firm
Boston Scientific Corporation
Quantity
US: 5,709 devices & OUS: 180 devices
Product code
FDT
FDA submissions
K193202
Distribution
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA VT, WA, WI & WV. The countries of Australia, Belgium, Canada, Czech Republic, Denmark, France, Germany, Great Britain, Italy, New Zealand, Spain, Sweden, Switzerland.

Product description

EXALT Model D Single-Use Duodenoscope, 2 pack: Outer box UPN# M00542421, Inner box UPN # M00542420 & Outer box UPN# M0054242CE1, Inner box UPN # M0054242CE0.

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

UDI-DI (GTIN): 08714729993605, Outer box UPN# M00542421, Inner box UPN # M00542420, Lot / Batch # 31118200, 31537337, 31118202, 31538040, 31129470, 31544823, 31145980, 31544825, 31150536, 31590918, 31150537, 31591865, 31150538, 31607291, 31175587, 31619981, 31193809, 31627631, 31193810, 31636279, 31232040, 31643793, 31239963, 31676406, 31248263, 31683921, 31251918, 31692257, 31269307, 31697145, 31275008, 31706116, 31283026, 31710191, 31305347, 31718478, 31322088, 31726052, 31334176, 31729292,31338402, 31734350, 31347461, 31738755, 31348990, 31738756, 31357937, 31754162, 31360721, 31762358, 31366683, 31764634, 31373783, 31767053, 31409362, 31774421, 31416272, 31783669, 31424741, 31826665, 31429697, 31828975, 31430274, 31846317, 31440192, 31858409, 31447120, 31874720, 31454624, 31915555, 31458181, 31939608, 31462880, 32033589, 31472232, 32047032, 31474875, 32082692, 31537336. UDI-DI (GTIN): 08714729995753, Outer box UPN# M0054242CE1, Inner box UPN # M0054242CE0, Lot /Batch # 31118201, 31331654 31129503, 31817723, 31265893.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93281).

Sources and updates