GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++… recallZ-0359-2024 · Instrumentation Laboratory · initiated Sep 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.

FDA root cause category: Under Investigation by firm

What the company is doing

An URGENT MEDICAL DEVICE RECALL notification letter dated 9/29/23 was sent to customers. Mandatory Customer Actions Please take the following immediate actions: " Check your inventory for the GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Discontinue use and destroy any GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to Werfen Regulatory at the fax number or e-mail address listed below. " Contact Werfen Customer Service at 1-800-955-9525 to order replacement PAKs (cartridges). " Share this information with your laboratory staff and follow your internal procedures. " Forward this notification to all affected locations within your facility. " Retain a copy of this notification for your records. Mandatory Response Tracking Record " Return the completed and signed Mandatory Response Tracking Form to: " Fax No.: 1-781-861-4207 Or " E-mail: [email protected] Contact Information for Questions " For technical questions, contact the Technical Support Center at 1-800-678-0710. " For replacement cartridges, contact Customer Service at 1-800-955-9525.

Recall details

Recall number
Z-0359-2024
Status
Open, Classified
Initiated by the company
Sep 29, 2023
Posted by the FDA
Nov 21, 2023
Recalling firm
Instrumentation Laboratory
Quantity
2448 units
Product code
CHL
FDA submissions
K161818, K992834
Distribution
US States: MA, MN, NY. India.

Product description

GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Part No. 00024315009: UDI-DI: 08426950082491, Lot Numbers: 323442F, 323142D, 323042K, 323142A, 323342K, 326142H, 323742C, 325442D, 325442E, 323042E, 324342A, 323042G, 323342H, 323542E, 324542A, 323542A, 324342B, 324042F, 323442H, 326442C, 325442F, 325942E, 326342B, 326142F, 325842C, 326442F; Part No. 00024330009: UDI-DI: 08426950082521, Lot Numbers: 323152A, 323152B, 324252B, 323552I, 323652I, 326152H, 323752B, 323352G, 323352H, 323052G, 323552A, 324352B, 324552A, 324052F, 323052E, 323352I, 323552F, 323352K, 325452F, 326352BC; Part No. 00024345009: UDI-DI: 08426950082552, Lot Numbers: 323462F, 323062K, 323162A, 323162C, 323562F, 323562I, 323662I, 323762D, 323762G, 323762J, 323862A, 323462C, 325762B, 323662D, 324462C, 324562A, 323362F, 325462F, 326362B; Part No. 00024360009: UDI-DI: 08426950082576, Lot Numbers: 323072J, 323072K, 323672I, 323772J, 324172D, 324372B, 323572A, 324072F, 325472F;