GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++… recallZ-0359-2024 · Instrumentation Laboratory · initiated Sep 29, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Instrumentation Laboratory or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Internal testing identified that the GEM Premier PAKs (cartridges) may exhibit an over-recovery (i.e., positive bias) of pO2 results in the low range (at pO2 levels less than 60 mmHg) that are outside of labeled performance specifications for pO2. The positive bias may lead to false positive for medical decision levels at pO2 < 60 mmHg, which may result in failure to detect and/or adequately treat hypoxemia and lead to a delay in treatment. The quality control procedures for the Non-iQM GEM Premier 3000 may not detect the non-conformance.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 9/29/23 was sent to customers. Mandatory Customer Actions Please take the following immediate actions: " Check your inventory for the GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Discontinue use and destroy any GEM Premier 3000 Non-iQM PAK (cartridge) lot numbers listed in Appendix A. " Document the destruction on the Mandatory Tracking Response Record and return the completed and signed form to Werfen Regulatory at the fax number or e-mail address listed below. " Contact Werfen Customer Service at 1-800-955-9525 to order replacement PAKs (cartridges). " Share this information with your laboratory staff and follow your internal procedures. " Forward this notification to all affected locations within your facility. " Retain a copy of this notification for your records. Mandatory Response Tracking Record " Return the completed and signed Mandatory Response Tracking Form to: " Fax No.: 1-781-861-4207 Or " E-mail: [email protected] Contact Information for Questions " For technical questions, contact the Technical Support Center at 1-800-678-0710. " For replacement cartridges, contact Customer Service at 1-800-955-9525.
Recall details
- Recall number
- Z-0359-2024
- Status
- Open, Classified
- Initiated by the company
- Sep 29, 2023
- Posted by the FDA
- Nov 21, 2023
- Recalling firm
- Instrumentation Laboratory
- Quantity
- 2448 units
- Product code
- CHL
- FDA submissions
- K161818, K992834
- Distribution
- US States: MA, MN, NY. India.
Product description
GEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Glucose and Lactate, Part Numbers 00024315009, 00024330009, 00024345009, 00024360009
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Part No. 00024315009: UDI-DI: 08426950082491, Lot Numbers: 323442F, 323142D, 323042K, 323142A, 323342K, 326142H, 323742C, 325442D, 325442E, 323042E, 324342A, 323042G, 323342H, 323542E, 324542A, 323542A, 324342B, 324042F, 323442H, 326442C, 325442F, 325942E, 326342B, 326142F, 325842C, 326442F; Part No. 00024330009: UDI-DI: 08426950082521, Lot Numbers: 323152A, 323152B, 324252B, 323552I, 323652I, 326152H, 323752B, 323352G, 323352H, 323052G, 323552A, 324352B, 324552A, 324052F, 323052E, 323352I, 323552F, 323352K, 325452F, 326352BC; Part No. 00024345009: UDI-DI: 08426950082552, Lot Numbers: 323462F, 323062K, 323162A, 323162C, 323562F, 323562I, 323662I, 323762D, 323762G, 323762J, 323862A, 323462C, 325762B, 323662D, 324462C, 324562A, 323362F, 325462F, 326362B; Part No. 00024360009: UDI-DI: 08426950082576, Lot Numbers: 323072J, 323072K, 323672I, 323772J, 324172D, 324372B, 323572A, 324072F, 325472F;
Devices named in this recall
4 devices in the FDA device registry whose identifier appears in the recall's code information.
- GEM Premier 3000 - BG HCT ELECT. GL - 150 Tests 00024315009Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 3000 - BG HCT ELECT GL - 300 Tests 00024330009Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 3000 - BG HCT ELECT. GL - 450 Tests 00024345009Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
- GEM Premier 3000 - BG HCT ELECT. GL - 600 Tests 00024360009Instrumentation Laboratory Company · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93313).
- Z-0357-2024 · Open, ClassifiedGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Part Numbers 00024307504, 00024330004, 00024345004
- Z-0358-2024 · Open, ClassifiedGEM Premier 3000 PAK - pH, pCO2, pO2, Hct, Na+, K+, Ca++, Part Numbers 00024307507, 00024315007, 00024330007, 00024345007, 00024360007
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.