Monarch Platform REF MON-000006 100-240V- 50/60Hz 1440w recallZ-0450-2024 · Auris Health Inc · initiated Oct 18, 2023
Class II Open, Classified All recalls posted in December 2023
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Auris Health Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
FDA root cause category: Software design
What the company is doing
On October 19, 2023, Auris Health, Inc. issued an "URGENT: MEDICAL DEVICE CORRECTION" notification to affected consignees. Auris asked consignees to take the following actions: Should an image flip be detected in the Virtual Bronchoscope view while navigating toward the target, please perform the following steps to correct the image orientation: 1. Retract the MONARCH Bronchoscope outside the patient 2. Undock the MONARCH Bronchoscope from the arms 3. Clear any fault pop-up messages 4. Re-dock the MONARCH Bronchoscope to the arms 5. Re-initialize Navigation and proceed with the case 6. Please review this letter carefully and share it with anyone in your facility that needs to be informed. 7. Please complete all fields of the attached Acknowledgement Form confirming receipt of this notification and return the form via email to [email protected]. If you have any questions regarding this letter or need to report any customer complaints, please contact the MONARCH Support team at +1 (800) 434-0032 or via email at [email protected].
Recall details
- Recall number
- Z-0450-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 18, 2023
- Posted by the FDA
- Dec 1, 2023
- Recalling firm
- Auris Health Inc
- Quantity
- 110 units In total
- Product code
- EOQ
- FDA submissions
- K193534
- Distribution
- US Distribution: AK, AL, AZ, CA, CO, CT, DC, DE, FL, GA, IL, IN, KY, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OR, PA, SC, SD, TN, TX, WI & WV.
Product description
Monarch Platform REF MON-000006 100-240V- 50/60Hz 1440w
Code information (models, lots, serial numbers)
Software version 2.2.4
Other devices under the same FDA submission
1 device cleared under K193534 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Monarch Bronchoscope MBR-000211-AAuris Health, Inc. · Bronchoscope (Flexible Or Rigid)12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93320).
- Z-0448-2024 · Open, ClassifiedMonarch Platform REF MON-000005-01, 100-240V- 50/60Hz 1440w
- Z-0449-2024 · Open, ClassifiedMonarch Platform REF MON-000005-01R, 100-240V- 50/60Hz 1440w
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page