Quantum TTC recallZ-0459-2024 · Wilson-Cook Medical Inc. · initiated Nov 3, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Wilson-Cook Medical Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Nonconforming devices while manufactured correctly, do not comply to their design requirements. Potential harms associated with balloon detachment are endoscopic retrieval of an impacted or non-impacted object, bleeding caused by the retrieval device, or the incident can occur without injury and the object is left to pass naturally.
FDA root cause category: Under Investigation by firm
What the company is doing
Cook Medical issued an URGENT: MEDICAL DEVICE REMOVAL notice on 11/03/2023 by mail. The notice explained the problem with the device, risk to health, and requested the following actions be taken: Customer: Cease use, quarantine, and return all affected devices. If devices were further distributed, those to whom the devices were sent must be notified. Distributor: Quarantine all affected devices, notify your customers to whom you distributed devices. Return all devices to Cook Medical.
Recall details
- Recall number
- Z-0459-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 3, 2023
- Posted by the FDA
- Dec 1, 2023
- Recalling firm
- Wilson-Cook Medical Inc.
- Quantity
- 553 devices
- Product code
- FGE
- FDA submissions
- K171223
- Distribution
- US
Product description
QUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-4X3, Order Number G22654; used to dilate strictures of the biliary tree
Code information (models, lots, serial numbers)
UDI/DI 00827002226548 (each), 10827002226545 (case), Lot Numbers: W4720848, W4721792, W4721795, W4722159, W4722981, W4723304, W4723846, W4724692, W4737576, W4738717, W4742525, W4743193, W4743605, W4743606, W4743607, W4743608, W4744022, W4744441, W4748043, W4748946, W4749748, W4750264, W4750265, W4750910, W4751772, W4751906, W4751914, W4753360, W4753361, W4753362, W4753365, W4753816, W4755138, W4756901, W4756903, W4756905, W4757732, W4758647, W4759511, W4759976, W4760715, W4761738, W4761740, W4761935, W4762150
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Quantum TTC G22654Cook Incorporated · CATHETER, BILIARY, DIAGNOSTIC1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93326).
- Z-0457-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3, Order Number G22657; used to dilate strictures of the biliary tree
- Z-0458-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-10X3-E, Order Number G22667; used to dilate strictures of the biliary tree
- Z-0460-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3, Order Number G22655; used to dilate strictures of the biliary tree
- Z-0461-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-6X3-E, Order Number G22765; used to dilate strictures of the biliary tree
- Z-0462-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-8X3, Order Number G22656; used to dilate strictures of the biliary tree
- Z-0463-2024 · Open, ClassifiedQUANTUM TTC BILIARY BALLOON DILATOR, REF QBD-8X3-E, Order Number G22766; used to dilate strictures of the biliary tree
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page