Injection Needle, REF: 25207 recallZ-0571-2024 · Karl Storz Endoscopy · initiated Oct 24, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Karl Storz Endoscopy or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The efficacy of the manual cleaning process cannot be assured for the affected injection needles intended for general surgery and visceral surgery, which if used, may expose the patient to a higher risk of infection.
FDA root cause category: Device Design
What the company is doing
On 10/24/23, recall notices were mailed to representatives for medical product safety, users, operators, distributors who were asked to do the following: 1) Discontinue use of all affected Injection Needles. 2) Pass on the Recall Notice to all users of the affected products and all other persons who need to be aware within your organization. 3) If you have or may have distributed the products listed, please identify and promptly notify those recipients, or provide the recalling firm with a list of customers who received/may have received the products listed. 4) Contact Customer Support to arrange for the return of affected devices by calling 800-421-0837. 5) Return the completed Customer Reply Form via email to [email protected] For all other questions regarding this communication, please send an email to [email protected]
Recall details
- Recall number
- Z-0571-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 24, 2023
- Posted by the FDA
- Dec 18, 2023
- Recalling firm
- Karl Storz Endoscopy
- Quantity
- 1
- Product code
- GCF
- FDA submissions
- K945209
- Distribution
- US: MI, NH, SC, PA, FL, NY, NC, MS, MO, CA, NJ, IL, NV, OK, CO, OH, WA, WI, VA, IN, IA, AR, TX, WV, MD, MN, GA, MA, NM, VT, AK, OR, ID, LA, KS, DC, KY, HI OUS: Mexica, Canada, Germany
Product description
Injection Needle, REF: 25207
Code information (models, lots, serial numbers)
UDI-DI: 04048551085776. All lot numbers
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- n.a. 25207Karl Storz GmbH & Co. KG · PROCTOSCOPE1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93328).
- Z-0572-2024 · Open, ClassifiedInjection Needle, 3 Fr., REF: 27030N; Injection Cannula, 6 Fr., REF: 27182A
- Z-0573-2024 · Open, ClassifiedInjection Needle, LUER-lock,0.8mm, REF: 26175PD
- Z-0574-2024 · Open, ClassifiedInjection Needle, single cvd. 70 degrees, REF: 738750; Injection Needle, single cvd. 80 degrees, REF: 738752; Injection Needle, bayonet, 13 cm, REF: 738756; Injection Needle, angul…
- Z-0575-2024 · Open, ClassifiedInjection Needle, curved 23 cm, REF: 8598A; Injection Needle, straight, 23 cm, REF: 8598B; Injection Needle, distal 45-degrees, 23 cm, REF: 8598D
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page