BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF:… recallZ-0451-2024 · BioFire Diagnostics, LLC · initiated Oct 17, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Routine quality control testing of affected blood culture identification panel with affected control panel may need to be amended, due to control panel/instrument characteristic changes, that lead C. tropicalis to be amplified in some portion of test runs, but then it's reported as "Not Detected" because the assay amplicon Tm is outside the acceptable limit assigned for calling positive results.

FDA root cause category: Process control

What the company is doing

On 10/17/23, "URGENT: MEDICAL DEVICE CORRECTION" notices were mailed to customers who were asked to do the following: 1) Review a complete list of BIOFIRE Quality Control Materials and references, for Example IQCP Risk Analysis can be found in BIOFIRE QC Tech Note (FLM1-PRT-0251), available from bioMrieux customer support. 2) Assess your Individualized Quality Control Plan (IQCP) and prepare for supplemental C. tropicalis testing if a Candida tropicalis Not Detected result is reported when testing with the MMQCI FilmArray BCID2 Control Panel M416 and/or amend testing plans to include alternate Quality Control materials such as those in FLM1-PRT-0251. 3) Complete and return the via email to [email protected] Contact customer support at [email protected] or via telephone by dialing 1.800.735.6544 and selecting option 5 for Product Technical Support with any questions or concerns.

Recall details

Recall number
Z-0451-2024
Status
Open, Classified
Initiated by the company
Oct 17, 2023
Posted by the FDA
Dec 1, 2023
Recalling firm
BioFire Diagnostics, LLC
Quantity
32,910
Product code
PEN
FDA submissions
K193519
Distribution
Worldwide Distribution: US (nationwide): DE, OH, NY, IL, KY, FL, GA, KS, CA, WI, NC,, AK, PA, MN, ND, IN, AR, CO, UT, TX, AL, NJ, MD, SD, LA, OR, MA, ID, NE, MT, VA, IA, MO, WV, TN, MI, WY, SC, AZ, WA, NH, OK, HI, MS, DC, VT, CT, NM, RI, ME, NV; and OUS (foreign) to countries of: Algeria, Angola, Armenia, Bahrain, Bolivia, Bosnia- Herzegovina, Brunei, Bulgaria, Costa Rica, Croatia, Cyprus, Dominican Republic, Dutch Antilles, Ecuador, El Salvador, Estonia, Gabon, Georgia, Guam, Guatemala, Honduras, Indonesia, Ireland, Iceland, Israel, Ivory Coast, Jordan, Kosovo, Kuwait, Latvia, Lebanon, Lithuania, Macao, Macedonia, Malaysia, Mali, Mongolia, Morocco, New Zealand, Nicaragua, Oman, Pakistan, Panama, Paraguay, Peru, Qatar, Romania, Saudi Arabia, Serbia , Slovenia, Tunisia, Uganda, United Arab Emirate, Uruguay, Venezuela, Vietnam, Zimbabwe, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Czech Republic, Spain, France, United Kingdom, Greece, India, Italia, Japan, Korea, Netherlands, Poland, Portugal, Sweden, Singapore, Thailand, Turkey, Taiwan, Switzerland, Austria, Denmark, Finland, Germany, Belgium, Hungary, South Africa, Philippines, Norway, Mexico

Product description

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, Inc (MMQCI) FilmArray BCID2 Control Panel M416, which is referenced in Quality Control Materials for use with BIOFIRE FILMARRAY and BIOFIRE SPOTFIRE Panels - FLM1-PRT-0251

Code information (models, lots, serial numbers)

REF/UDI-DI: RFIT-ASY-0147/00815381020338. FLM1-PRT-0251-16 and prior.