VITROS recallZ-0546-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Oct 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT PRODUCT CORRECTION NOTIFICATION letter dated 10/10/23 was sent to customers. REQUIRED ACTIONS " Review the Customer Actions for the Condition Codes described in this notification. " Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than October 18, 2023. " Save this notification with your User Documentation or post this notification by each VITROS 3600/4600/5600/XT 3400/XT 7600 System until this issue is resolved. " If your laboratory has experienced the issue described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization (Previously Ortho Care). Resolution QuidelOrtho is investigating root cause and will communicate again after root cause has been determined. A software update to address this issue is currently in development. Contact Information We apologize for any inconvenience this may cause your laboratory. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.
Recall details
- Recall number
- Z-0546-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 10, 2023
- Posted by the FDA
- Dec 13, 2023
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Quantity
- 51 units
- Product code
- JJE
- FDA submissions
- K182063
- Distribution
- US Nationwide. Global Distribution.
Product description
VITROS XT 3400 Chemistry System Product Code 6844458 Running Software Version 3.8.0
Code information (models, lots, serial numbers)
Product Code: 6844458, UDI-DI: 10758750031986; Serial Numbers: J34500140 J34500113 J34500180 J34500114 J34500257 J34500116 J34500289 J34500117 J34500319 J34500123 J34500386 J34500128 J34500422 J34500129 J34500447 J34500130 J34500473 J34500139 J34500474 J34500190 J34500510 J34500261 J34500539 J34500269 J34500547 J34500271 J34500579 J34500292 J34500588 J34500301 J34500622 J34500337 J34500656 J34500356 J34500777 J34500363 J34500818 J34500515 J34500887 J34500520 J34500893 J34500536 J34500896 J34500582 J34500911 J34500609 J34500931 J34500628 J34500658 J34500711 J34500793
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VITROS 6844458Ortho-Clinical Diagnostics, Inc. · ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE4 recalls2 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93366).
- Z-0543-2024 · Open, ClassifiedVITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
- Z-0544-2024 · Open, ClassifiedVITROS 4600 Chemistry System Product Code 6802445 and Refurbished Product Code 6900440 Running Software Version 3.8.0
- Z-0545-2024 · Open, ClassifiedVITROS 5600 Integrated System Product Code 6802413 and Refurbished Product Code 6802915 Running Software Version 3.8.0
- Z-0547-2024 · Open, ClassifiedVITROS XT 7600 Integrated System Product Code 6844461 Running Software Version 3.8.0
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page