Sensica UO System recallZ-0554-2024 · C.R. Bard Inc · initiated Nov 15, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from C.R. Bard Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.

FDA root cause category: Under Investigation by firm

What the company is doing

BD (CR Bard) issued an URGENT Medical Device Product Advisory to its consignees on 11/15/2023, via FedEx and email. The notice explained the problem with the device, potential hazard, and requested that the consignee perform Operating System and Sensica Software settings changes per the instructions provided on all devices within your facility s control using the attached directions.

Recall details

Recall number
Z-0554-2024
Status
Open, Classified
Initiated by the company
Nov 15, 2023
Posted by the FDA
Dec 14, 2023
Recalling firm
C.R. Bard Inc
Quantity
148 units
Product code
EXS
Distribution
US Nationwide distribution.

Product description

Sensica Urine Output System, Catalog Number SCCS1002

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007, BMGPRMYS020, BMGPRMYS026, BMGPRMYS028, BMGPRMYS030, BMFWRMSBT4, BMFWRMSJ45, BMFWRMSJ49, BMFZRMS104, BMGNRMS085, BMFSRMSH98, BMFSRMSJ01, BMFVRMSQ62, BMFVRMSQ66, BMFVRMSQ68, BMFYRMSJ41, BMFYRMSJ49, BMFZRMS067, BMFZRMS070, BMFZRMS080, BMFZRMS113, BMFZRMS117, BMGNRMS070, BMFURMSF02, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ89, BMFVRMSQ91, BMFWRSJ46, BMGPRMYS011, BMGPRMYS018, BMGPRMYS022 , BMGPRMYS023, BMGPRMYS027, BMFYRMSJ35, BMFYRMSJ37, BMFYRMSJ54, BMFYRMSJ58, BMGNRMS079, BMGNRMSG69, BMGNRMSG74, BMFSRMSH94, BMFSRMSH99, BMFSRMSJ38, BMFSRMSJ44, BMFTRMS138, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFQRMS248, BMFQRMS250, BMFRRMS646, BMFRRMS647, BMGPRMYS008, BMGPRMYS009, BMGPRMYS010, BMGPRMYS013, BMGPRMYS014, BMGPRMYS016, BMFZRMS073, BMFSRMSH78, BMFSRMSH79, BMFSRMSH80, BMFSRMSH81, BMFSRMSH82, BMFSRMSH83, BMFSRMSH93, BMFSRMSJ00, BMFSRMSJ14, BMFSRMSJ15, BMFQRMS246, BMFWRMSJ39, BMFWRMSJ41, BMFSRMSJ36, BMFSRMSJ37, BMFSRMSJ42, BMFTRMS106, BMFTRMS109, BMFTRMS114, BMFTRMS142, BMFURMSF01, BMFURMSF05, BMFSRMSH95, BMFSRMSH97, BMFTRMS120, BMFTRMS122, BMFTRMS123, BMFTRMS135, BMFTRMS136, BMFTRMS137, BMGPRMYS002, BMGPRMYS012, BMGPRMYS019 , BMGPRMYS029, BMFPRMYS009, BMFQRMS244, BMFQRMS272, BMFQRMS276, BMFQRMS286, BMFQRMS293, BMFSRMSJ43, BMFTRMS121, BMGNRMS067, BMGNRMS076, BMFVRMSQ58, BMFVRMSQ70, BMFVRMSQ73, BMFVRMSQ74, BMFVRMSQ90, BMFVRMSR17, BMFWRMSJ27, BMFWRMSJ35, BMFURMSF06, BMFURMSF09, BMFWRMSJ36, BMFWRMSJ51, BMFZRMS068, BMFQRMS294, BMFZRMS105, BMFSRMSJ20, BMFTRMS140, BMFVRMSQ63, BMFVRMSQ64, BMFVRMSQ65, BMFVRMSQ72, BMFVRMSQ75, BMFVRMSQ76, BMFVRMSQ77, BMFVRMSQ78, BMFVRMSQ80, BMFYRMSJ39, BMFZRMS081, BMFZRMS116, BMGNRMS015, BMGNRMS029, BMGNRMS073

Devices named in this recall