Why it was recalled
BD (C.R. Bard) has received user complaints reporting that when the device was connected to WiFi/Internet, the time displayed on the Sensica device changed during a patient monitoring session which resulted in 1) incorrect time displayed on the monitoring screen, 2) incorrect hour block timestamps on the monitoring screen, 3) incorrect hour timestamps on data screens, and 4) incorrect catheter use time.
FDA root cause category: Under Investigation by firm
What the company is doing
BD (CR Bard) issued an URGENT Medical Device Product Advisory to its consignees on 11/15/2023, via FedEx and email. The notice explained the problem with the device, potential hazard, and requested that the consignee perform Operating System and Sensica Software settings changes per the instructions provided on all devices within your facility s control using the attached directions.
Recall details
- Recall number
- Z-0554-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 15, 2023
- Posted by the FDA
- Dec 14, 2023
- Recalling firm
- C.R. Bard Inc
- Quantity
- 148 units
- Product code
- EXS
- Distribution
- US Nationwide distribution.
Product description
Sensica Urine Output System, Catalog Number SCCS1002
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
UDI/DI 00801741215773, Serial numbers: BMGPRMYS001 , BMGPRMYS003, BMGPRMYS004, BMGPRMYS005, BMGPRMYS006, BMGPRMYS007, BMGPRMYS020, BMGPRMYS026, BMGPRMYS028, BMGPRMYS030, BMFWRMSBT4, BMFWRMSJ45, BMFWRMSJ49, BMFZRMS104, BMGNRMS085, BMFSRMSH98, BMFSRMSJ01, BMFVRMSQ62, BMFVRMSQ66, BMFVRMSQ68, BMFYRMSJ41, BMFYRMSJ49, BMFZRMS067, BMFZRMS070, BMFZRMS080, BMFZRMS113, BMFZRMS117, BMGNRMS070, BMFURMSF02, BMFURMSF04, BMFURMSF14, BMFURMSF21, BMFVRMSQ51, BMFVRMSQ52, BMFVRMSQ54, BMFVRMSQ79, BMFVRMSQ83, BMFVRMSQ89, BMFVRMSQ91, BMFWRSJ46, BMGPRMYS011, BMGPRMYS018, BMGPRMYS022 , BMGPRMYS023, BMGPRMYS027, BMFYRMSJ35, BMFYRMSJ37, BMFYRMSJ54, BMFYRMSJ58, BMGNRMS079, BMGNRMSG69, BMGNRMSG74, BMFSRMSH94, BMFSRMSH99, BMFSRMSJ38, BMFSRMSJ44, BMFTRMS138, BMFSRMSJ13, BMFSRMSJ27, BMFSRMSJ33, BMFSRMSJ34, BMFQRMS248, BMFQRMS250, BMFRRMS646, BMFRRMS647, BMGPRMYS008, BMGPRMYS009, BMGPRMYS010, BMGPRMYS013, BMGPRMYS014, BMGPRMYS016, BMFZRMS073, BMFSRMSH78, BMFSRMSH79, BMFSRMSH80, BMFSRMSH81, BMFSRMSH82, BMFSRMSH83, BMFSRMSH93, BMFSRMSJ00, BMFSRMSJ14, BMFSRMSJ15, BMFQRMS246, BMFWRMSJ39, BMFWRMSJ41, BMFSRMSJ36, BMFSRMSJ37, BMFSRMSJ42, BMFTRMS106, BMFTRMS109, BMFTRMS114, BMFTRMS142, BMFURMSF01, BMFURMSF05, BMFSRMSH95, BMFSRMSH97, BMFTRMS120, BMFTRMS122, BMFTRMS123, BMFTRMS135, BMFTRMS136, BMFTRMS137, BMGPRMYS002, BMGPRMYS012, BMGPRMYS019 , BMGPRMYS029, BMFPRMYS009, BMFQRMS244, BMFQRMS272, BMFQRMS276, BMFQRMS286, BMFQRMS293, BMFSRMSJ43, BMFTRMS121, BMGNRMS067, BMGNRMS076, BMFVRMSQ58, BMFVRMSQ70, BMFVRMSQ73, BMFVRMSQ74, BMFVRMSQ90, BMFVRMSR17, BMFWRMSJ27, BMFWRMSJ35, BMFURMSF06, BMFURMSF09, BMFWRMSJ36, BMFWRMSJ51, BMFZRMS068, BMFQRMS294, BMFZRMS105, BMFSRMSJ20, BMFTRMS140, BMFVRMSQ63, BMFVRMSQ64, BMFVRMSQ65, BMFVRMSQ72, BMFVRMSQ75, BMFVRMSQ76, BMFVRMSQ77, BMFVRMSQ78, BMFVRMSQ80, BMFYRMSJ39, BMFZRMS081, BMFZRMS116, BMGNRMS015, BMGNRMS029, BMGNRMS073
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Sensica UO System SCCS1002C. R. Bard, Inc. · URINOMETER, ELECTRICAL2 recalls289 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page