Skip to content

RANDOX Total Bilirubin recallZ-0501-2024 · Randox Laboratories Ltd. · initiated Oct 11, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Randox has had reports of elevated patient results using Plasma (lithium heparin) samples with Total Bilirubin reagent (TBil) Vanadate Oxidation Method.

FDA root cause category: Device Design

What the company is doing

Consignees were emailed an Urgent Device Correction letter on 10/11/2023. The notice instructed consignees to only use Total Bilirubin Vanadate Oxidation Method on Serum samples, to review results generated with Plasma samples in like with the clinical profiles of the patient, discuss the contents of the recall notice with your facility's Medical Director, and to return the provided response form completed to [email protected] within 5 working days. The recall notice is to be further distributed to all affected customers and those within consignee organizations, if further distributed. A second recall notification, dated 4/26/24, was sent to consignees identifying the reason for the product issue to be intralipid interference. Reagent IFUs have been updated with sample specific Plasma (lithium heparin) interference claims for intralipids, which can be found on www.randox.com. Consignees are asked to complete and return the provided response form to [email protected] within five working days.

Recall details

Recall number
Z-0501-2024
Status
Open, Classified
Initiated by the company
Oct 11, 2023
Posted by the FDA
Dec 6, 2023
Recalling firm
Randox Laboratories Ltd.
Quantity
1644 units
Product code
JFM
FDA submissions
K152344
Distribution
US Nationwide distribution in the states of AR, GA, NC, TN, & PR.

Product description

RANDOX Total Bilirubin.

Code information (models, lots, serial numbers)

Catalog No. BR8377, BR9766, & BR4061; GTIN: 05055273214772, 05055273208337, & 05055273214765; All Batches.

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page