Skip to content

FreeStyle Libre 3 App, part of Continuous Glucose… recallZ-0734-2024 · Abbott Diabetes Care, Inc. · initiated Feb 9, 2023

Class II Completed All recalls posted in January 2024

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Diabetes Care, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

If using affected glucose monitoring app on Android 13 Operating System, extended periods of signal loss may be experienced, due to app not connected, which could impact ability to receive glucose reading/alarms, which could lead to undetected low or high glucose, which could result in delayed treatment: not taking insulin (for high glucose), or not taking glucose (for low glucose) when required.

FDA root cause category: Software design

What the company is doing

On 2/9/23, an Important Message/Communication was sent to customers informing them of the following: If you experience extended periods of signal loss, either of the following actions should allow reconnection: 1) Turn your smartphone's Bluetooth OFF and then ON 2) Turn your smartphone OFF and then ON To be notified if extended periods of signal loss occur: 1) Open your FreeStyle Libre 3 app, navigate to Alarms, and configure your signal loss alarm to be ON with "Override Do Not Disturb" enabled. 2) Please refer to www.FreeStyleAndroid.com for more information on how to configure alarms. On 2/27/23, another Important Message/Communication was sent to customers informing them that updated software version 3.4.2 was available to fix the connection issue. Questions can be directed to Customer Care team at 1-855-632-8658. The team is available 7 days a week from 8 AM to 8 PM Eastern Time, excluding holidays.

Recall details

Recall number
Z-0734-2024
Recall class
Class II
Status
Completed
Initiated by the company
Feb 9, 2023
Posted by the FDA
Jan 16, 2024
Recalling firm
Abbott Diabetes Care, Inc.
Quantity
45,173
Product code
QLG
FDA submissions
K213996
Distribution
US Nationwide distribution in the states of NY, CA, NJ, UT, PA, MS, OH, TX, MN, FL, LA, NE, MI, VA, NC, MA, IN, SC, IA.

Product description

FreeStyle Libre 3 App, part of Continuous Glucose Monitoring System, when used on the Android 13 Operating System

Code information (models, lots, serial numbers)

UDI-DI: 00357599816001, Software v.3.4.0 and below

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page