CereLink recallZ-0548-2024 · Integra LifeSciences Corp. · initiated Oct 5, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Integra LifeSciences Corp. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm is expanding the existing recall RES 90457 for the Cerelink, initiated in June 2022, to include all CereLink ICP Extension Cables. The updated corrective action includes a design change to the extension cable to address the "out of range readings" issue.
FDA root cause category: Device Design
What the company is doing
On October 5, 2023, the firm issued an updated letter, intended to replace the notification sent on June 22, 2022 for the out-of-range readings issue (RES 90457). As an output of the root cause and failure investigation, the design of the extension cable is being modified to incorporate a solution to address the "out of range readings" issue. The firm instructed customers to return all existing CereLink ICP Extension Cables in their inventory. This includes all CereLink ICP Extension Cables included with the monitor AND all CereLink ICP Extension Cables provided separately. Customers should check their inventory and quarantine any affected product. Once the firm receives the customer's response form, the firm will provide instructions to return the product. Customers with questions about this recall should contact their Account Manager.
Recall details
- Recall number
- Z-0548-2024
- Status
- Open, Classified
- Initiated by the company
- Oct 5, 2023
- Posted by the FDA
- Dec 14, 2023
- Recalling firm
- Integra LifeSciences Corp.
- Quantity
- 4941
- Product code
- GWM
- FDA submissions
- K210993, K232890
- Distribution
- US distribution nationwide to the following states: AL AR AZ CA CO CT DC DE FL GA IA KY LA MA ME MN ND NE NH NJ NY OK PA RI TN TX UT VA WI WV OUS distribution to Australia/New Zealand, Canada, and Europe/Middle East/Africa.
Product description
CereLink ICP Extension Cable (Extension Cable), Model No. 826845 Used with Codman CereLink Intracranial Pressure (ICP) Monitor, Model No. 826820/826820P. Cable is also provided with monitor as part of the same SKU.
Code information (models, lots, serial numbers)
UDI-DI 10381780520665 All lots distributed
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- CereLink™ 826845Integra Lifesciences Production Corporation · Device, monitoring, intracranial pressure1 recall2 deaths16 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.