AirTouch N20 recallZ-0540-2024 · ResMed Ltd. · initiated Nov 20, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ResMed Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Masks non-invasive interface for channeling airflow to patients have magnets that have potential magnetic interference with active medical implants that interact with magnets, and metallic implants/objects containing ferromagnetic material so contraindications and warnings about safe distance to medical devices/implants are being updated for patients using masks or anyone in close physical contact
FDA root cause category: Device Design
What the company is doing
On 11/20/23, "Urgent Field Safety Notices" and FAQs were placed on the firms website at www.resmed.com/magnetupdate, and these notices were emailed to healthcare providers, and distributors who were asked to do the following: 1) Provide a copy of this notice and the physician letter to prescribing physicians, and/or any other relevant healthcare professionals, notifying them of the updated labeling (contraindications and warning). 2) Provide a copy of the patient letter to all patients currently using an affected mask with magnets, notifying them of the updated labeling (contraindications and warning). Note, patients are instructed in the patient letter to contact their mask provider for a replacement mask if they are now contraindicated. 3) Provide a replacement mask for contraindicated patients to an alternative mask without magnets in a timely manner. Where an alternative mask is not available, inform patients to consult with their physician. 4) Instruct patients to consult their physician and/or manufacturer of their implant / other medical device if they require additional information on the potential adverse effects of magnetic fields for their particular device (as described in warnings). 5) For alternative mask options, please contact your ResMed or Customer Service representative. 6) Complete and return the customer acknowledgement form via email to [email protected] For any questions, please contact your local ResMed contact or visit www.resmed.com/contact.
Recall details
- Recall number
- Z-0540-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 20, 2023
- Posted by the FDA
- Dec 21, 2023
- Recalling firm
- ResMed Ltd.
- Quantity
- 424,309
- Product code
- BZD
- FDA submissions
- K161978
- Distribution
- US nationwide distribution including Puerto Rico, Guam, and Virgin Islands. OUS (foreign) countries of: AE, AR, AT, AU, BB, BD, BE, BG, BH, BM, BR, CA, CH, CL, CN, CO, CW, CY, CZ, DE, DK, DO, DZ, EC, EE, EG, ES, FI, FO, FR, GB, GF, GG, GL, GP, GR, GT, GY, HK, HN, HR, HU, ID, IE, IL, IM, IN, IR, IS, IT, JE, JO, JP, KE, KG, KR, KW, LB, LK, LT, LV, LY, MA, MC, MM, MQ, MT, MU, MV, MX, MY, NA, NC, NG, NL, NO, NP, NZ, OM, PA, PE, PF, PH, PK, PL, PM, PT, QA, RE, RO, RS, RU, SA, SD, SE, SG, SI, SK, SV, TH, TN, TR, TT, TW, UY, VE, VN, WF, YT, ZA
Product description
AirTouch N20 Nasal Mask and User Guide
Code information (models, lots, serial numbers)
All mask lots used with User Guide 638218/version 2020-02, and prior. UDI-DI/GTIN: 619498639069, 619498639045, 619498639052, 619498639021, 619498639038, 619498639007, 619498639076, 619498639014, 619498639540, 619498639205, 619498639229, 619498639113, 619498639090, 619498639120, 619498639212
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- AirTouch N20 63900Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63901Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63902Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63903Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63904Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63905Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63906Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63907Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
- AirTouch N20 63954Resmed Pty Ltd · Ventilator, Non-Continuous (Respirator)1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93499).
- Z-0536-2024 · Open, ClassifiedAirFit N10 Nasal Masks and User Guide
- Z-0537-2024 · Open, ClassifiedAirFit F20 Full Face Mask and User Guide
- Z-0538-2024 · Open, ClassifiedAirTouch F20 Full Face Mask and User Guide
- Z-0539-2024 · Open, ClassifiedAirFit N20 Nasal Mask and User Guide
- Z-0541-2024 · Open, ClassifiedAirFit F30 Full Face Mask and User Guide
- Z-0542-2024 · Open, ClassifiedAirFit F30i Full Face Mask and User Guide
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page