Baxter Vascular Probe, Product Codes recallZ-0659-2024 · Baxter Healthcare Corporation · initiated Dec 6, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Cardiovascular Surgery marketing brochure (US-AS46-220001) that was distributed for the products listed below contains inaccurate content that does not align with some of the products Instructions for Use (IFU) or Prescribing Information (PI).
FDA root cause category: Under Investigation by firm
What the company is doing
Baxter issued an IMPORTANT PRODUCT INFORMATION to its consignees on 12/06/2023 via USPS first class mail. The notice explained the issue, potential hazard involved, and requested the following: Actions to be Taken by Customers: 1. Healthcare providers may continue to safely use the products listed above while following the associated Instructions for Use. 2. Cease any further distribution of the Cardiovascular Surgery marketing brochure for the products listed above. All copies of this marketing brochure should be promptly discarded. The brochure reference number, US-AS46-220001, can be found at the bottom of the last page of the brochure. 3. If you received this communication directly from Baxter, please acknowledge receipt by responding on our customer portal at https://BaxterFieldActionCustomerPortal.onprocess.com, even if you do not have any inventory. Log in to the portal using the account number listed on the enclosed reply form instruction sheet. Acknowledging receipt of this notification will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to the supplier according to their instructions. 5. If you distributed a copy of this brochure to other facilities or departments within your institution, please forward a copy of this communication to them. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any copies of the brochure to other facilities, please distribute this notification to customers and check the associated box on the customer portal. For general questions, contact Baxter Healthcare Center for Service at 888-229-0001, betwe
Recall details
- Recall number
- Z-0659-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 6, 2023
- Posted by the FDA
- Jan 5, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Product code
- DWP
- FDA submissions
- K130896
- Distribution
- US Nationwide distribution.
Product description
Baxter Vascular Probe, Product Codes: a) 7081015: 8CM, 1- 1.5 MM GLOBAL, b) 7081520: 8CM, 1.5-2 MM GLOBAL, c) 7082025: 8CM, 2-2.5 MM GLOBAL, d) 7151015: 15CM, 1-1.5 MM GLOBAL, e) 7151520: 15CM, 1.5- 2 MM GLOBAL, f) 7152025: 15CM, 2- 2.5 MM GLOBAL, g) 7451015: 45CM, 1- 1.5 MM GLOBAL, h) 7451520: 45CM, 1.5- 2 MM GLOBAL, i) 7452025: 45CM, 2- 2.5 MM GLOBAL, j) 7081015ES: 8CM, 1-1.5 MM GLOBAL, k) 7081520ES: 8CM, 1.5- 2 MM GLOBAL, l) 7082025ES: 8CM, 2-2.5 MM GLOBAL, m) 7151015ES: 15CM, 1-1.5 MM GLOBAL, n) 7151520ES: 15CM, 1.5- 2 MM GLOBAL, o) 7152025ES: 15CM, 2-2.5 MM GLOBAL, p) 7451015ES: 45CM, 1-1.5 MM GLOBAL, q) 7451520ES: 45CM, 1.5-2 MM GLOBAL, r) 7452025ES: 45CM, 2-2.5 MM GLOBAL
Code information (models, lots, serial numbers)
a) 7081015: UDI/DI 0338-9564-17, b) 7081520: UDI/DI 0338-9564-18, c) 7082025: UDI/DI 0338-9564-19, d) 7151015: UDI/DI 0338-9564-12, e) 7151520: UDI/DI 0338-9564-11, f) 7152025: UDI/DI 0338-9564-13, g) 7451015: UDI/DI 0338-9564-14, h) 7451520: UDI/DI 0338-9564-15, i) 7452025: UDI/DI 0338-9564-16, j) 7081015ES: UDI/DI 0338-9564-10, k) 7081520ES: UDI/DI 0338-9564-26, l) 7082025ES: UDI/DI 0338-9564-27, m) 7151015ES: UDI/DI 0338-9564-21, n) 7151520ES: UDI/DI 0338-9564-20, o) 7152025ES: UDI/DI 0338-9564-22, p) 7451015ES: UDI/DI 0338-9564-24, q) 7451520ES: UDI/DI 0338-9564-23, r) 7452025ES: UDI/DI 0338-9564-25, All serial numbers
Other devices under the same FDA submission
18 devices cleared under K130896 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- Vascular Probe 7081015Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7081520Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7082025Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7151015Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7151520Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7152025Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7451015Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7451520Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe 7452025Synovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe Extra Supple 7081015ESSynovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe Extra Supple 7081520ESSynovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
- Vascular Probe Extra Supple 7082025ESSynovis Life Technologies, Inc. · DILATOR, VESSEL, SURGICAL1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93503).
- Z-0656-2024 · Open, ClassifiedBaxter Flo-Rester, Disposable internal vessel occluder, distributed in the following sizes: a) Product Code 50100: 12MM, 1.00 MM GLOBAL; b) Product Code 50150: 12MM, 1.50 MM GLO…
- Z-0657-2024 · Open, ClassifiedBaxter Coseal Premix, Surgical Sealant, Product Codes: a) 934070; b) 934071; c) 934072
- Z-0660-2024 · Open, ClassifiedBaxter Preveleak Surgical Sealant, Product Codes: ADS201808
- Z-0663-2024 · Open, ClassifiedBaxter Flo-Thru Intraluminal Shunt, Product Codes: a) Product Code FT12100, 1.00 GLOBAL; b) Product Code FT12125, 1.25 GLOBAL; c) Product Code FT12150, 1.50 GLOBAL; d) Product …
- Z-0664-2024 · Open, ClassifiedPeri-Guard Repair Patch, Product Codes: a) PG1016 : 10 X 16 CMUS, TS b) PG1225: 12 X 25 CMUS, TS c) PG0404: 4 X 4 CMUS, TS d) PG0608: 6 X 8CMUS, TS e) PG0814: 8 X 14 CMUS, TS
- Z-0665-2024 · Open, ClassifiedPeri-Guard Repair Patch, Product Codes: a) PC1016NBIO: 10X16CMUS, b) PC1225NBIO: 12X25CMUS, c) PC0404NBIO: 4X4CMUS, d) PC0608NBIO: 6X8CMUS, e) PC0814NBIO: 8X14CMUS
- Z-0666-2024 · Open, ClassifiedVascu-Guard Peripheral Vascular Patch (US), Product Codes: a) VG0108: 0.8 X 8 CM US, TS; b) VG0110: 1 X 10 CM US, TS; c) VG0106: 1 X 6 CM US, TS; d) VG0209: 2 X 9 CM US, TS
- Z-0667-2024 · Open, ClassifiedBaxter FloSeal Hemostatic Matrix NL, 5 mL, Product Codes: ADS201883
- Z-0668-2024 · Open, ClassifiedPERI-GUARD SUPPLE, Product Codes: a) SPG1016: 10X16 CM US, TS; b) SPG0404: 4 X 4 CM US, TS; c) SPG0608: 6 X 8 CM US, TS; d) SPG0814: 8 X 14 CM US, TS; e) PC1016SNBIO: 10X…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page