Olympus recallZ-0936-2024 · Olympus Corporation of the Americas · initiated Nov 10, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Product balloon issues: inflating, deflating and/or retrieving the devices; reports of bursting and leaking and events involving foreign body in patient and prolonged procedure.
FDA root cause category: Process control
What the company is doing
Olympus issued Urgent Medical Device Removal Letter on 11/10/23. Firm issued a Correction letter dated December 1, 2023 to the letter of November 10, 2023 including Additional Lot numbers . Letter states reason for recall, health risk and action to take: 1. Inspect your inventory and identify any models and lots subject to this action. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and cease use of the affected model/lots. Any models and lots that are not impacted can remain in inventory and be used. The model/lot number can be found on the box or pouch in Figure 1. 2. Call your Olympus customer service representative at 1-800-848-9024, option 2, with the quantity, model, and lot number of affected devices. Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification: a. Go to https://olympusamerica.com/recall. b. Enter the recall number: 0433 c. Complete the form as instructed. 4. If you have distributed these devices outside your facility, please notify your customers of this matter immediately. Your notification to your customers may be enhanced by including a copy of this letter. Olympus requests that you report complaints to our Technical Assistance Center (TAC) at 1-800-848-9024, option 1.
Recall details
- Recall number
- Z-0936-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 10, 2023
- Posted by the FDA
- Jan 31, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 5232 units
- Product code
- FGE
- FDA submissions
- K180086
- Distribution
- Nationwide. Foreign: AU BR BR CA DE HK IN JP SG
Product description
EZDilate Fixed Wire Balloon 13.5-14.5-15.5. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-1580
Code information (models, lots, serial numbers)
UDI-DI: 00821925033191 Lot Numbers: 360910, 360911, 364014, 364158, 364338, 367646, 367647, 367648, 383363, 383846, 383847, 384058
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Olympus BD-400P-1580Gyrus ACMI, LLC · Dilator, esophageal2 recalls4 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93508).
- Z-0933-2024 · Open, ClassifiedEZDilate Fixed Wire Balloon 6-7-8. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-0880
- Z-0934-2024 · Open, ClassifiedEZDilate Fixed Wire Balloon 8.5-9.5-10.5. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-1080
- Z-0935-2024 · Open, ClassifiedEZDilate Fixed Wire Balloon 11-12-13. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-1380
- Z-0937-2024 · Open, ClassifiedEZDilate Fixed Wire Balloon 16-17-18. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-1880
- Z-0938-2024 · Open, ClassifiedEZDilate Fixed Wire Balloon 18-19-20. Indicated for endoscopic dilation of strictures of the esophagus in adults and adolescents (>12 years). Model: BD-400P-2080
- Z-0939-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 6-7-8. indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years) Model: BD-410X-0855
- Z-0940-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 8.5-9.5-10.5. indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years) Model: BD-410X-1055
- Z-0941-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 11-12-13. Indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years). Model: BD-410X-1355
- Z-0942-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 13.5-14.5-15.5. Indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years). Model: BD-410X-1555
- Z-0943-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 16-17-18. Indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years). Model: BD-410X-1855
- Z-0944-2024 · Open, ClassifiedEZDilate Wire Guided Balloon 18-19-20. Indicated for endoscopic dilation of strictures in the alimentary tract in adults and adolescents (>12 years). Model: BD-410X-2055
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page