SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use… recallZ-0624-2024 · Randox Laboratories Ltd. · initiated Nov 6, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Randox Laboratories Ltd. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A decrease in the reactivity of HSV Type 1/2 lgM in the Serology ToRCH lgM Positive Control when tested on the DiaSorin Liaison XL.

FDA root cause category: Process control

What the company is doing

Randox issued Urgent Medical Device Correction Letter on 11/6/23 to the US Distributor Randox Laboratories-US, Ltd. , Kearneysville, West Virginia, for issuance to the user level. Letter states reason for recall, health risk and action to take: As the control material is still testing positive for HSV Type 1/2 lgM, there is no risk to health. If the control did test negative for this analyte, the affected run should be discarded, and the samples re-analysed. Action to be taken: " Discuss the contents of this notice with your Medical Director. " Complete and return the response form 12187-QA to [email protected] within five working days. " Please discard all copies of the IFU and download the latest version from www.randox.com.alth risk and action to take: (

Recall details

Recall number
Z-0624-2024
Status
Open, Classified
Initiated by the company
Nov 6, 2023
Posted by the FDA
Jan 2, 2024
Recalling firm
Randox Laboratories Ltd.
Quantity
1 unit
Product code
OHQ
Distribution
NY, WV

Product description

SEROLOGY ToRCH lgM 2022 POSITIVE CONTROLI Intended for use with in vitro assays for determination of IgM antibodies to Cytomegalovirus (CMV), IgM antibodies to Rubella Virus, IgM antibodies to Toxoplasma gondii, IgM antibodies to Herpes Simplex Virus Type 1 and 2. Catalog Number: SR10349

Code information (models, lots, serial numbers)

GTIN: 05055273216424 Bath/Lot Number: 157SR