Skip to content

Xstar Kojo Safety Slit Knife, 2.5mm 45 Bevel Up, Part… recallZ-0633-2024 · Beaver Visitec International, Inc. · initiated Oct 10, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beaver Visitec International, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

One lot of Part Number 373025, Xstar Kojo Safety Slit Knife 2.5mm 45 bevel up, incorrectly contains a bevel down blade instead of a bevel up blade.

FDA root cause category: Under Investigation by firm

What the company is doing

An "URGENT Field Safety Notice" dated 10/10/23 was sent to customers. Actions Required by You Our records show you received one or more affected product. You can recognize the affected product by determining whether the lot number printed on the device label is one of the affected lots. PLEASE perform the following actions: 1. Immediately examine your inventory and determine if the blades on the knives are bevel-up as intended or bevel down. Only destroy the knives in your inventory that have the incorrect bevel orientation (bevel down). If you have further distributed this product, please notify your customers by initiating a sub-recall. Consider all potential users of this product in your user supply chain. You are encouraged to use a copy of this recall notification letter when contacting your customers. 2. Review the enclosed Business Reply Form (BRF) and indicate whether you have the affected product in your possession. 3. Please enter the following information on the enclosed BRF: part number(s), lot number(s) and quantity(ies) of device(s) destroyed. 4. Return the completed BRF and evidence of destruction of affected product to BVI by emailing the attached to [email protected]. BVI values your business and apologizes for the inconvenience this may have caused. Please direct any questions regarding credit to our customer service department by email and include in the subject line: Master Case PIR 00487979. Country Customer Service Email UK [email protected] FR [email protected] DE/ AT [email protected] IT [email protected] ROW [email protected]

Recall details

Recall number
Z-0633-2024
Status
Open, Classified
Initiated by the company
Oct 10, 2023
Posted by the FDA
Jan 4, 2024
Recalling firm
Beaver Visitec International, Inc.
Quantity
2160 units
Product code
HNN
Distribution
International Distribution to Japan.

Product description

Xstar Kojo Safety Slit Knife, 2.5mm 45 Bevel Up, Part Number 373025

Code information (models, lots, serial numbers)

UDI-DI: 30886158000373; Lot Number: 6049877

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page