Dover URINE METER ADD-A-FOLEY TRAY, 200 ML, NEEDLE SAMPLING… recallZ-1046-2024 · Cardinal Health 200, LLC · initiated Dec 4, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Convenience kits were manufactured and distributed with components which were subsequently recalled by Nurse Assist.
FDA root cause category: Nonconforming Material/Component
What the company is doing
Cardinal Health Issued an URGENT KIT/TRAY COMPONENT RECALL NOTICE ON 12/04/2023 via FedEx next day delivery. The notice explained the issue, potential risk to health, and requested the following actions be taken: If you locate affected product within your facility, please do the following: 1. QUARANTINE affected kits/trays. 2. ALERT your clinicians of this action. 3. AFFIX the enclosed WARNING LABEL to the front of each kit/tray so that it is clearly visible to clinicians, instructing them to remove and discard the recalled component(s). 4. NOTE if you have transferred the affected kits/trays to other departments or facilities within your system, please provide a copy of this notice and recall acknowledgement form to them. 5. DISTRIBUTOR if you have further distributed affected kits/trays, please notify your customers of this recall. 6. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of over labeled product via email [email protected] or fax to 614.652.9648. Additional labels will be sent to your facility upon request. For field action concerns, please contact the field action team at: [email protected] or call 800-292-9332. On 12/21/2023, Cardinal Health issued an URGENT: EXPANSION KIT/TRAY COMPONENT RECALL notice stating the following: As an update to Cardinal Healths initial communication on December 05, 2023, additional kit/trays are being included as part of the component recall. If you did not receive the initial communication on December 05, 2023, you were not impacted by the initial scope of the event. Instructions below are unchanged. Please reference Attachment 1 - Updated Impacted Products V2 for additional products impacted. The notice explained the issue, potential risk to health, and requested the same actions be taken as listed above in the 12/04/2023 letter.
Recall details
- Recall number
- Z-1046-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 4, 2023
- Posted by the FDA
- Feb 16, 2024
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- 435060 units
- Product code
- KNX
- Distribution
- US and EMEA, Japan, Latin America
Product description
Dover URINE METER ADD-A-FOLEY TRAY, 200 ML, NEEDLE SAMPLING PORT, DRAIN TUBE, PREP TRAY; Dover UNIVERSAL TRAY, 1000 ML BASIN, PVP SWAB STICKS, PREPPING COMPONENTS, 30 CC PRE-FILLED SYRINGE; Dover URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, A/R, DRAIN TUBE, PREP TRAY, 30 ML SYRINGE; Dover HYDROGEL COATED LATEX FOLEY TRAY 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN TUBE; Dover 100% SILICONE FOLEY TRAY, NEEDLE PORT, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover 100% SILICONE FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover SILICONE COATED LATEX FOLEY TRAY, 18 FR (6.0 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN TUBE; Dover SILICONE COATED LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN TUBE; Dover HYDROGEL COATED LATEX FOLEY TRAY, LL, 18 FR (6.0 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT, CSD; Dover HYDROGEL COATED LATEX FOLEY TRAY, LL, 16 FR (5.3 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT, CSD; Dover 100% SILICONE FOLEY TRAY, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG, CATH SECUREMENT; Dover 100% SILICONE FOLEY TRAY, LUER LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG, CATH SECUREMENT; Dover URINE METER ADD-A-FOLEY TRAY, 400 ML, TOP VENT, LUER-LOCK SAMPLING, DRAIN SPOUT, PREP TRAY; Dover 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2-WAY, CATH SECUREMENT; Dover 100% SILICONE SILVER-HYDROGEL COATED FOLEY TRAY, 14 FR (4.7 MM), 5CC, 2-WAY, 2000ML DRAIN BAG; Dover 100% SILICONE FOLEY TRAY, LUER LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX FOLEY TRAY, NEEDLELESS, 16 FR (5.3 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover 100% SILICONE FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, CATH SECUREMENT; Dover 100% SILICONE FOLEY TRAY, LUER LOCK, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover 100% SILICONE FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN TUBE; Dover HYDROGEL COATED LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN SPOUT; Dover 100% SILICONE FOLEY TRAY, NEEDLELESS, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG; Dover URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, DRAIN TUBE, PREP TRAY; Dover URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLE SAMPLING, DRAIN TUBE, PREP TRAY, 30 ML LUBE SYRINGE; Dover 100% SILICONE FOLEY TRAY, NEEDLE PORT, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX FOLEY TRAY, 18 FR (6.0 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX FOLEY TRAY, 14 FR (4.7 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX FOLEY TRAY, LUER LOCK, 14 FR (4.7 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover 100% SILICONE FOLEY TRAY, NEEDLE PORT, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN TUBE; Dover HYDROGEL COATED LATEX FOLEY TRAY, 16 FR (5.3 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN SPOUT ; Dover HYDROGEL COATED LATEX FOLEY TRAY, 18 FR (6.0 MM), 5 CC, 2000 ML DRAINAGE BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, NEEDLELESS, 16 FR (5.3 MM), 5 CC, 2000 ML BAG; Dover HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, LUER LOCK, 18 FR (6.0 MM), 5 CC, 2000 ML BAG; Dover 100% SILICONE 200 ML U/M FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover URINE DRAINAGE ADD-A-FOLEY TRAY, NEEDLELESS SAMPLING, A/R CHAMBER, DRAIN SPOUT, PREP TRAY; Dover HYDROGEL COATED LATEX FOLEY TRAY, NEEDLELESS, 14 FR (4.7 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX FOLEY TRAY, NEEDLELESS, 18 FR (6.0 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT; Dover" 100% SILICONE FOLEY TRAY, NEEDLELESS, 18 FR (6.0 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover HYDROGEL COATED LATEX TRAY, LL, 14 FR (4.7 MM), 5 CC, 2000 ML BAG, DRAIN SPOUT, CSD; Dover 100% SILICONE SILVERHYDROGEL FOLEY TRAY; Dover Silicone 100% 16FR/18F; Dover 100% Silver Hydro tray
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
a) 6140, 30884521015214 (cs), 10884521015210 (ea), UDI/DI Lot Numbers: 2206723264, 2310813264, 2213047664, 2213301264, 2211624264 2205319164, 2209519564, 2218504964; b) 6148LL, UDI/DI 30884521015528 (cs), 10884521015524 (ea), Lot Numbers: 2210903364, 2308701264, 2310107364, 2212900264, 2212900364; c) 6153, UDI/DI 30884521015252 (cs), 10884521015258 (ea), Lot Numbers: 2206007764, 2222830664, 2233605264, 2300125764, 2300125864, 2300132664, 2314302964, 2205201164, 2206007864, 2213047864, 2223545264, 2304402364, 2317731064, 2211628564, 2205329464, 2205328764, 2213742064, 2216533964; d) 6154, UDI/DI 20884521015262 (cs), 10884521015265 (ea), Lot Numbers: 2233605364, 2206011164, 2309042764, 2217922664; e) 8949, UDI/DI 30884521015399 (cs), 10884521015395 (ea), Lot Numbers: 2206903964, 2206904064, 2313507764, 2313507864, 2322322564, 2223546964, 2306613764, 2205201264, 2310108164, 2205320964, 2209526764; f) PP14SD, UDI/DI 20884521029672 (cs), 10884521029675 (ea), Lot Numbers: 2301030764, 2216500164, 2309039464, 2314220064, 2203931264, 2317729764, 2200119164, 2200427964, 2201132564, 2215117664; g) PP16SD, UDI/DI 20884521029696 (cs), 10884521029699 (ea), Lot Numbers: 2213738864, 2313806164, 2321617264, 2303818764, 2304400164, 2301023064, 2301023164, 2307110464, 2308004064, 2310901664, 2310901864, 2314306964, 2318416964, 2319218264, 2203928664, 2210901564, 2209524064, 2216500264, 2216533864, 2304400264, 2302000364, 2302401364, 2302401464, 2313100464, 2313806064, 2215111764, 2215111864, 2216801464, 2305903064, 2134217964, 2200119064, 2200428064, 2201132664, 2204633564, 2204633664, 2208814464, 2215823864; h) PP18SD, UDI/DI 20884521029719 (cs), 10884521029712 (ea), Lot Numbers: 2303822164, 2310807864, 2212321464, 2317729964, 2203931364 i) 3718, Lot Numbers: 2303815264, 2307202164, 2209534664, 2223544564, 2219402664, 2220726564, 2206727764, 2201131264, 2203933064, 2209525464, 2210232964, 2217921964 j) 3716, 2203930364, 2206718464, 2213046664, 2222114064, 2222114164, 2222830064, 2303815064, 2303815164, 2308002764, 2307109164, 2307109264, 2308002564, 2314901264, 2314901364, 2206718364, 2209530364, 2214452164, 2215825964, 2218612564, 2304401864, 2304701764, 2305203064, 2308002664, 2308700864, 2308719164, 2211621264, 2211621364, 2210900964, 2210901064, 2216502664, 2310107164, 2207404064, 2310301864, 2200120364, 2200120464, 2200429564, 2200429664, 2203930464, 2203930564, 2204635764, 2204635864, 2205317964, 2205318064, 2206024964, 2206025064, 2208812264, 2209101864, 2212301864, 2212301964, 2216806264, 2217921864 k) PP18ULD, Lot Numbers: 2215117764, 2312205864 l) PP16ULD, Lot Numbers: 2209514764, 2209514864, 2214449264, 2217919664, 2308700364, 2311521264, 2317729864, 2313805964, 2322321364, 2208814564, 2215823964 n) 8324, Lot Number: 2210221064 o) 7020LL, Lot Numbers: 2303824064, 2210220864, 2223546364, 2232613764, 2312912164, 2209526064, 2215826864 p) 6080, Lot Number: 2214452564 q) PP16XSD, Lot Numbers: 2301030964, 2204633764, 2207402364, 2219310764, 2302405264, 2312205764, 2217919764 r) 6144ICLL, UDI/DI 30884521009121 (cs), 10884521009127 (ea) Lot Numbers: 2216802364, 2208800564, 2307111564 s) 6148LL, Lot Numbers: 2210903364, 2308701264, 2310107364, 2212900264, 2212900364 t) 6155, Lot Numbers: 2205319564, 2205328664, 2216802264, 2318417664, 2213047964, 2312208664, 2312208864, 2313100564, 2313100664, 2210903464, 2310107464 u) 3532, Lot Numbers: 2213741364, 2306612564, 2314901664 v) 6014, Lot Numbers: 2211030964, 2308700964, 2309042564, 2308900664, 2216806564, 2205318964, 2215119764
Devices named in this recall
9 devices in the FDA device registry whose identifier appears in the recall's code information.
- Dover 6144ICLLCardinal Health 200, LLC · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall2 reports
- Dover PP14SDCardinal Health 200, LLC · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall2 reports
- Dover PP16SDCardinal Health 200, LLC · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall1 report
- Dover 6140Cardinal Health, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall0 reports
- Dover 6153Cardinal Health 200, LLC · Urinary Drainage Collection Kit1 recall0 reports
- Dover 6154Cardinal Health 200, LLC · Urinary Drainage Collection Kit1 recall0 reports
- Dover 8949Cardinal Health, Inc. · Collector, Urine, (And Accessories) For Indwelling Catheter1 recall0 reports
- Dover 6148LLCardinal Health, Inc. · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall0 reports
- Dover PP18SDCardinal Health 200, LLC · COLLECTOR, URINE, (AND ACCESSORIES) FOR INDWELLING CATHETER1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93551).
- Z-1043-2024 · Open, ClassifiedArgyle" Suction Catheter Tray with Chimney Valve Sterile Saline 18 Fr/Ch (6.0mm); Argyle" Suction Catheter Tray with Chimney Valve Sterile Saline, 6 Fr/Ch (2.0 mm); Argyle" Graduat…
- Z-1045-2024 · Open, ClassifiedDover" 100% SILICONE 400 ML U/M FOLEY TRAY, NEEDLE PORT, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover" HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, NEEDLE, 16 FR (5.3 MM), 5 CC, …
- Z-1047-2024 · Open, ClassifiedDover" HYDROGEL COATED LATEX 400 ML U/M FOLEY TRAY, LL, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG, CSD; Dover" 100% SILICONE SILVER-HYDROGEL COATED 400ML U/M TEMP SENSE FOLEY TRAY, …
- Z-1048-2024 · Open, ClassifiedDover" 100% SILICONE 400 ML U/M FOLEY TRAY, COUD TIP, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, COUD TIP, 18 FR (6.0 MM), 5 CC, 2000 ML D…
- Z-1050-2024 · Open, ClassifiedDover" 100% SILICONE FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover" 100% SILICONE 400 ML U/M FOLEY TRAY, LUER LOCK, 16 FR (5.3 MM), 5 CC, 2000 …
- Z-1051-2024 · Open, ClassifiedDover" 100% SILICONE 400ML U/M TEMP SENSE FOLEY TRAY, LL, 18 FR (6.0 MM), 5CC, 2000ML DRAIN BAG, CSD
- Z-1052-2024 · Open, ClassifiedDover" 100% SILICONE FOLEY TRAY, NEEDLELESS, 14 FR (4.7 MM), 5 CC, 2000 ML DRAIN BAG, DRAIN SPOUT; Dover" HYDROGEL COATED LATEX FOLEY CATHETER KIT, 5 CC, 2-WAY, 14 FR (4.7 MM), 10 …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page