BD PosiFlush recallZ-0736-2024 · Becton Dickinson & Company · initiated Nov 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Becton Dickinson & Company or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An incorrect stopper was used in nine lots of 10mL BD PosiFlush Normal Saline Flush Syringes with General Pump Compatibility. The use of BD PosiFlush Prefilled Flush Syringes with this stopper causes an increase in injection force, which has the potential to trigger alarms if used with a syringe pump.
FDA root cause category: Under Investigation by firm
What the company is doing
An URGENT: Medical Device Product Advisory dated 11/16/23 was sent to customers. Please Take the Following Actions: 1. Immediately review your inventory for the affected product. a. If the product is used for syringe pumps, destroy all affected product subject to this field action following your institution's process for destruction. b. If the product is used for manual application, the product can continue to be used per IFU. 2. If the affected product was previously used on a patient, no further actions are needed. 3. If the affected product is currently in use with a syringe pump, replace the syringe with a non-impacted lot. Please note that healthcare facilities and providers should rely on their syringe pump manufacturer's IFU for the specific syringes that are compatible with their syringe pumps. 4. Share this field action notification with all users of the product within your facility network to ensure they are also aware of this field action. 5. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material so that BD may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement if product is used on syringe pumps and has been discarded. 6. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA 1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852-9787 Distributors Please Take the Following Actions: 1. Immediately review your inventory for the affected product. Destroy all unused product subject to the product advisory following your institution s process for destruction. 2. Identify all customers within your distribution network that purchased any affected product as defined in this notification. Provide
Recall details
- Recall number
- Z-0736-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 16, 2023
- Posted by the FDA
- Jan 17, 2024
- Recalling firm
- Becton Dickinson & Company
- Quantity
- 5,085,120 units
- Product code
- NGT
- FDA submissions
- K161552
- Distribution
- US Nationwide distribution including in the states of AL, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX,UT, VA, WA, WI, WV.
Product description
PosiFlush Prefilled Saline Syringe with General Pump Compatibility (10mL Syringe with 10mL Saline Fill), Catalog Number 306547
Code information (models, lots, serial numbers)
UDI-DI: 30382903065470; Lot Numbers: 3207548 3207555 3214826 3226386 3214828 3226388 3250302 3250303 3250304
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BD PosiFlush 306547Becton, Dickinson And Company · Catheter,intravascular,therapeutic,short-term less than 30 days2 recalls1 death331 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K161552 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- BD PosiFlush™ 306499Becton, Dickinson And Company · Saline, vascular access flush45 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page