Skip to content

LITe DECOMPRESSION recallZ-0610-2024 · Stryker Spine · initiated Nov 1, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Spine or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Product arm was manufactured with the incorrect clamp component. As a result, the affected product arm cannot attach to the arm post shaft.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Stryker issued Urgent Medical Device Correction on 11/20/2023 notifying customers of removal of lot 227764 of the LITe Decompression Tube Snake Arm, catalog number 48080230. A portion of one lot of the LITe Decompression Tube Snake Arm was manufactured with the incorrect clamp component. As a result, the affected Snake Arms cannot attach to the arm post shaft. Customers are to review inventory and remove units from points of use for return to Stryker, share this information with appropriate personnel and return the Business Reply form to [email protected]. If you have any questions or concerns, please contact Regulatory Compliance at (201) 749-8090.

Recall details

Recall number
Z-0610-2024
Status
Open, Classified
Initiated by the company
Nov 1, 2023
Posted by the FDA
Dec 21, 2023
Recalling firm
Stryker Spine
Quantity
9
Product code
LXH
Distribution
US: MD, MI, NM, TX & OUS: Canada & France.

Product description

LITe Decompression Snake Arm, REF 48080230

Code information (models, lots, serial numbers)

UDI-DI (GTIN): 07613327263909, lot # 227764.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates