FilmArray Pneumonia Panel (Pneumo) recallZ-0994-2024 · BioFire Diagnostics, LLC · initiated Dec 14, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from BioFire Diagnostics, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is a potential for false positive results when using Pneumonia panel.
FDA root cause category: Under Investigation by firm
What the company is doing
On December 14, 2023, Biomerieux, issued a "Urgent Medical Recall" to affected consignees via: FedEx . Biomerieux asked consignees to take the following actions: 1. Distribute this information to all appropriate personnel in your laboratory, retain a copy in your files, and forward this information to all parties that may use this product, including others to whom you may have transferred our product. 2. Use another method to confirm positive results for seasonal Coronavirus from the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus prior to reporting the result (e.g., consider results from testing a nasopharyngeal swab with a different BIOFIRE respiratory panel such as the BIOFIRE Respiratory Panel 2.1). Alternatively, the BIOFIRE Pneumonia Panel or BIOFIRE Pneumonia Panel plus seasonal Coronavirus positive result should not be reported. 3. Complete the Acknowledgement Form in Attachment A (on the following page) and return it to bioMrieux to confirm receipt of this notice. It is important that you return the acknowledgement form to bioMrieux even if you determine that this urgent product correction notice does not impact your facility.
Recall details
- Recall number
- Z-0994-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 14, 2023
- Posted by the FDA
- Feb 6, 2024
- Recalling firm
- BioFire Diagnostics, LLC
- Quantity
- 274,290 units
- Product code
- QBH
- FDA submissions
- K180966
- Distribution
- Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Argentina, Bolivia, Brazil, Chili, China, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Hong Kong, Honduras, Japan, South Korea, Mexico, Nicaragua, Panama, Peru, Philippines, Paraguay, Singapore, El Salvador, Thailand, Taiwan, Uruguay, Venezuela.
Product description
FilmArray Pneumonia Panel (Pneumo), REF: FIT-ASY-0144, CED 2797, IVD, Rx Only (30 test kit)
Code information (models, lots, serial numbers)
UDI: 00815381020178/ All unexpired lots
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0144Biofire Diagnostics, LLC · Respiratory Panel5 recalls7 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K180966 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- FilmArray® Pneumonia Panel (Pneumo) RFIT-ASY-0145Biofire Diagnostics, LLC · Respiratory Panel2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93592).
- Z-0993-2024 · Open, ClassifiedFilmArray Pneumonia Panel Plus (Pneumoplus), REF: RFIT-ASY-0143, CED 2797, IVD, Rx Only (30 test kit)
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page