BD Recanalization System, REF: BDRECANSYSTEM recallZ-0676-2024 · Bard Peripheral Vascular Inc · initiated Nov 14, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.
FDA root cause category: Process design
What the company is doing
On 11/14/23 recall notices were mailed and emailed to customers who were asked to do the following: 1) Discontinue use of affected product and immediately quarantine until ready to return for service activities. 2) Share this notification with all device users within your facility and network to ensure they are aware of this field action. This notification should also be shared with any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to [email protected] 4) Recalling firm will contact customers to coordinate product return for service. Customers with questions can call the firm's North American Regional Complaint Center at 1-844-823-5433, Say "Recall" when prompted M-F 8am - 5pm CT, or email [email protected]
Recall details
- Recall number
- Z-0676-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 14, 2023
- Posted by the FDA
- Jan 5, 2024
- Recalling firm
- Bard Peripheral Vascular Inc
- Quantity
- 403
- Product code
- PDU
- FDA submissions
- K203363
- Distribution
- US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN
Product description
BD Recanalization System, REF: BDRECANSYSTEM
Code information (models, lots, serial numbers)
Show the full code information (4k characters)
UDI-DI: 00801741202421. Serial Numbers: BSX000001, BSX000002, BSX000003, BSX000004, BSX000005, BSX000006, BSX000007, BSX000008, BSX000009, BSX000010, BSX000011, BSX000012, BSX000013, BSX000014, BSX000015, BSX000016, BSX000017, BSX000018, BSX000019, BSX000020, BSX000021, BSX000022, BSX000023, BSX000024, BSX000025, BSX000026, BSX000027, BSX000028, BSX000029, BSX000030, BSX000031, BSX000032, BSX000033, BSX000034, BSX000035, BSX000036, BSX000037, BSX000038, BSX000039, BSX000040, BSX000041, BSX000042, BSX000043, BSX000044, BSX000045, BSX000046, BSX000047, BSX000048, BSX000049, BSX000050, BSX000051, BSX000052, BSX000053, BSX000054, BSX000055, BSX000056, BSX000057, BSX000058, BSX000059, BSX000060, BSX000061, BSX000062, BSX000063, BSX000064, BSX000065, BSX000066, BSX000067, BSX000068, BSX000109, BSX000110, BSX000111, BSX000112, BSX000113, BSX000114, BSX000115, BSX000116, BSX000117, BSX000118, BSX000119, BSX000120, BSX000121, BSX000122, BSX000123, BSX000124, BSX000125, BSX000126, BSX000127, BSX000128, BSX000129, BSX000130, BSX000131, BSX000132, BSX000133, BSX000134, BSX000135, BSX000136, BSX000137, BSX000138, BSX000139, BSX000140, BSX000141, BSX000142, BSX000143, BSX000144, BSX000145, BSX000146, BSX000147, BSX000148, BSX000149, BSX000150, BSX000151, BSX000152, BSX000153, BSX000154, BSX000155, BSX000156, BSX000157, BSX000158, BSX000159, BSX000160, BSX000161, BSX000162, BSX000163, BSX000164, BSX000165, BSX000166, BSX000167, BSX000168, BSX000169, BSX000170, BSX000171, BSX000172, BSX000173, BSX000174, BSX000175, BSX000176, BSX000177, BSX000178, BSX000179, BSX000180, BSX000181, BSX000182, BSX000183, BSX000184, BSX000185, BSX000186, BSX000187, BSX000188, BSX000189, BSX000190, BSX000191, BSX000192, BSX000193, BSX000194, BSX000195, BSX000196, BSX000197, BSX000198, BSX000199, BSX000200, BSX000201, BSX000202, BSX000203, BSX000204, BSX000205, BSX000206, BSX000207, BSX000208, BSX000209, BSX000210, BSX000211, BSX000212, BSX000213, BSX000214, BSX000215, BSX000216, BSX000217, BSX000218, BSX000219, BSX000220, BSX000221, BSX000222, BSX000223, BSX000224, BSX000225, BSX000226, BSX000227, BSX000228, BSX000229, BSX000230, BSX000231, BSX000232, BSX000233, BSX000234, BSX000235, BSX000236, BSX000237, BSX000238, BSX000239, BSX000240, BSX000241, BSX000242, BSX000243, BSX000244, BSX000245, BSX000246, BSX000247, BSX000248, BSX000249, BSX000250, BSX000251, BSX000252, BSX000253, BSX000254, BSX000255, BSX000256, BSX000257, BSX000258, BSX000259, BSX000260, BSX000261, BSX000262, BSX000263, BSX000264, BSX000265, BSX000266, BSX000267, BSX000268, BSX000269, BSX000270, BSX000271, BSX000272, BSX000273, BSX000274, BSX000275, BSX000276, BSX000277, BSX000278, BSX000279, BSX000280, BSX000281, BSX000282, BSX000283, BSX000284, BSX000285, BSX000286, BSX000287, BSX000288, BSX000289, BSX000290, BSX000291, BSX000292, BSX000293, BSX000294, BSX000295, BSX000296, BSX000297, BSX000298, BSX000299, BSX000300, BSX000301, BSX000302, BSX000303, BSX000304, BSX000305, BSX000306, BSX000307, BSX000308, BSX000309, BSX000310, BSX000311, BSX000312, BSX000313, BSX000314, BSX000315, BSX000316, BSX000317, BSX000318, BSX000319, BSX000320, BSX000321, BSX000322, BSX000323, BSX000324, BSX000325, BSX000326, BSX000327, BSX000328, BSX000330, BSX000331, BSX000332, BSX000333, BSX000334, BSX000335, BSX000336, BSX000337, BSX000338, BSX000339, BSX000340, BSX000341, BSX000342, BSX000343, BSX000344, BSX000345, BSX000346, BSX000347, BSX000348, BSX000349, BSX000350, BSX000351, BSX000352, BSX000353, BSX000354, BSX000355, BSX000356, BSX000357, BSX000358, BSX000359, BSX000360, BSX000361, BSX000362, BSX000363, BSX000364, BSX000365, BSX000366, BSX000367, BSX000368, BSX000369, BSX000370, BSX000371, BSX000372, BSX000373, BSX000374, BSX000375, BSX000376, BSX000377, BSX000378, BSX000379, BSX000380, BSX000381, BSX000382, BSX000383, BSX000384, BSX000385, BSX000386, BSX000387, BSX000388, BSX000389, BSX000390, BSX000391, BSX000392, BSX000393, BSX000394, BSX000395, BSX000396, BSX000397, BSX000398, BSX000399, BSX000400, BSX000401, BSX000402, BSX000403, BSX000404, BSX000405, BSX000406, BSX000407, BSX000408, BSX000409, BSX000410, BSX000411, BSX000412, BSX000413, BSX000414, BSX000415, BSX000416, BSX000417, BSX000418, BSX000419, BSX000420, BSX000421, BSX000422, BSX000423, BSX000424, BSX000425, BSX000426, BSX000427, BSX000428, BSX000429, BSX000430, BSX000431, BSX000432, BSX000433, BSX000434, BSX000435, BSX000436, BSX000437, BSX000438, BSX000439, BSX000440, BSX000441, BSX000442, BSX000443, BSX000444
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Recanalization System BDRECANSYSTEMBard Peripheral Vascular, Inc. · Catheter for crossing total occlusions1 recall8 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page