BD Recanalization System, REF: BDRECANSYSTEM recallZ-0676-2024 · Bard Peripheral Vascular Inc · initiated Nov 14, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Bard Peripheral Vascular Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

FDA root cause category: Process design

What the company is doing

On 11/14/23 recall notices were mailed and emailed to customers who were asked to do the following: 1) Discontinue use of affected product and immediately quarantine until ready to return for service activities. 2) Share this notification with all device users within your facility and network to ensure they are aware of this field action. This notification should also be shared with any organization where the potentially affected devices have been transferred. 3) Complete and return the response form via email to [email protected] 4) Recalling firm will contact customers to coordinate product return for service. Customers with questions can call the firm's North American Regional Complaint Center at 1-844-823-5433, Say "Recall" when prompted M-F 8am - 5pm CT, or email [email protected]

Recall details

Recall number
Z-0676-2024
Status
Open, Classified
Initiated by the company
Nov 14, 2023
Posted by the FDA
Jan 5, 2024
Recalling firm
Bard Peripheral Vascular Inc
Quantity
403
Product code
PDU
FDA submissions
K203363
Distribution
US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

Product description

BD Recanalization System, REF: BDRECANSYSTEM

Code information (models, lots, serial numbers)

Show the full code information (4k characters)

UDI-DI: 00801741202421. Serial Numbers: BSX000001, BSX000002, BSX000003, BSX000004, BSX000005, BSX000006, BSX000007, BSX000008, BSX000009, BSX000010, BSX000011, BSX000012, BSX000013, BSX000014, BSX000015, BSX000016, BSX000017, BSX000018, BSX000019, BSX000020, BSX000021, BSX000022, BSX000023, BSX000024, BSX000025, BSX000026, BSX000027, BSX000028, BSX000029, BSX000030, BSX000031, BSX000032, BSX000033, BSX000034, BSX000035, BSX000036, BSX000037, BSX000038, BSX000039, BSX000040, BSX000041, BSX000042, BSX000043, BSX000044, BSX000045, BSX000046, BSX000047, BSX000048, BSX000049, BSX000050, BSX000051, BSX000052, BSX000053, BSX000054, BSX000055, BSX000056, BSX000057, BSX000058, BSX000059, BSX000060, BSX000061, BSX000062, BSX000063, BSX000064, BSX000065, BSX000066, BSX000067, BSX000068, BSX000109, BSX000110, BSX000111, BSX000112, BSX000113, BSX000114, BSX000115, BSX000116, BSX000117, BSX000118, BSX000119, BSX000120, BSX000121, BSX000122, BSX000123, BSX000124, BSX000125, BSX000126, BSX000127, BSX000128, BSX000129, BSX000130, BSX000131, BSX000132, BSX000133, BSX000134, BSX000135, BSX000136, BSX000137, BSX000138, BSX000139, BSX000140, BSX000141, BSX000142, BSX000143, BSX000144, BSX000145, BSX000146, BSX000147, BSX000148, BSX000149, BSX000150, BSX000151, BSX000152, BSX000153, BSX000154, BSX000155, BSX000156, BSX000157, BSX000158, BSX000159, BSX000160, BSX000161, BSX000162, BSX000163, BSX000164, BSX000165, BSX000166, BSX000167, BSX000168, BSX000169, BSX000170, BSX000171, BSX000172, BSX000173, BSX000174, BSX000175, BSX000176, BSX000177, BSX000178, BSX000179, BSX000180, BSX000181, BSX000182, BSX000183, BSX000184, BSX000185, BSX000186, BSX000187, BSX000188, BSX000189, BSX000190, BSX000191, BSX000192, BSX000193, BSX000194, BSX000195, BSX000196, BSX000197, BSX000198, BSX000199, BSX000200, BSX000201, BSX000202, BSX000203, BSX000204, BSX000205, BSX000206, BSX000207, BSX000208, BSX000209, BSX000210, BSX000211, BSX000212, BSX000213, BSX000214, BSX000215, BSX000216, BSX000217, BSX000218, BSX000219, BSX000220, BSX000221, BSX000222, BSX000223, BSX000224, BSX000225, BSX000226, BSX000227, BSX000228, BSX000229, BSX000230, BSX000231, BSX000232, BSX000233, BSX000234, BSX000235, BSX000236, BSX000237, BSX000238, BSX000239, BSX000240, BSX000241, BSX000242, BSX000243, BSX000244, BSX000245, BSX000246, BSX000247, BSX000248, BSX000249, BSX000250, BSX000251, BSX000252, BSX000253, BSX000254, BSX000255, BSX000256, BSX000257, BSX000258, BSX000259, BSX000260, BSX000261, BSX000262, BSX000263, BSX000264, BSX000265, BSX000266, BSX000267, BSX000268, BSX000269, BSX000270, BSX000271, BSX000272, BSX000273, BSX000274, BSX000275, BSX000276, BSX000277, BSX000278, BSX000279, BSX000280, BSX000281, BSX000282, BSX000283, BSX000284, BSX000285, BSX000286, BSX000287, BSX000288, BSX000289, BSX000290, BSX000291, BSX000292, BSX000293, BSX000294, BSX000295, BSX000296, BSX000297, BSX000298, BSX000299, BSX000300, BSX000301, BSX000302, BSX000303, BSX000304, BSX000305, BSX000306, BSX000307, BSX000308, BSX000309, BSX000310, BSX000311, BSX000312, BSX000313, BSX000314, BSX000315, BSX000316, BSX000317, BSX000318, BSX000319, BSX000320, BSX000321, BSX000322, BSX000323, BSX000324, BSX000325, BSX000326, BSX000327, BSX000328, BSX000330, BSX000331, BSX000332, BSX000333, BSX000334, BSX000335, BSX000336, BSX000337, BSX000338, BSX000339, BSX000340, BSX000341, BSX000342, BSX000343, BSX000344, BSX000345, BSX000346, BSX000347, BSX000348, BSX000349, BSX000350, BSX000351, BSX000352, BSX000353, BSX000354, BSX000355, BSX000356, BSX000357, BSX000358, BSX000359, BSX000360, BSX000361, BSX000362, BSX000363, BSX000364, BSX000365, BSX000366, BSX000367, BSX000368, BSX000369, BSX000370, BSX000371, BSX000372, BSX000373, BSX000374, BSX000375, BSX000376, BSX000377, BSX000378, BSX000379, BSX000380, BSX000381, BSX000382, BSX000383, BSX000384, BSX000385, BSX000386, BSX000387, BSX000388, BSX000389, BSX000390, BSX000391, BSX000392, BSX000393, BSX000394, BSX000395, BSX000396, BSX000397, BSX000398, BSX000399, BSX000400, BSX000401, BSX000402, BSX000403, BSX000404, BSX000405, BSX000406, BSX000407, BSX000408, BSX000409, BSX000410, BSX000411, BSX000412, BSX000413, BSX000414, BSX000415, BSX000416, BSX000417, BSX000418, BSX000419, BSX000420, BSX000421, BSX000422, BSX000423, BSX000424, BSX000425, BSX000426, BSX000427, BSX000428, BSX000429, BSX000430, BSX000431, BSX000432, BSX000433, BSX000434, BSX000435, BSX000436, BSX000437, BSX000438, BSX000439, BSX000440, BSX000441, BSX000442, BSX000443, BSX000444

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.