V. Mueller recallZ-0737-2024 · Steris Corporation · initiated Nov 16, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Steris Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Bandage Scissors handle cover component contains an incorrect color masterbatch resin carrier. This may allow a pink dye/residue to appear on the Bandage Scissors and/or sterilization pouches after processing.
FDA root cause category: Under Investigation by firm
What the company is doing
Steris issued Urgent Medical Device Recall Letter to Distributors and Customers on November 16, 2023. Letter states reason for recall, health risk and action to take: Please ensure the following steps are completed: 1. Please immediately inspect your on-hand inventory for product affected by this recall. 2. Please complete the Medical Device Recall Response Form included with this letter and destroy any remaining product in your inventory. STERIS will coordinate credit issuance upon receipt of the completed Recall Response Form. 3. Return the completed Recall Response Form via email to: [email protected] or via fax to (440) 392-8963. If you have any questions regarding this recall, please contact [email protected].
Recall details
- Recall number
- Z-0737-2024
- Status
- Open, Classified
- Initiated by the company
- Nov 16, 2023
- Posted by the FDA
- Jan 17, 2024
- Recalling firm
- Steris Corporation
- Quantity
- 4,214 units
- Product code
- LRW
- Distribution
- US Nationwide distribution.
Product description
V. Mueller Universal Bandage Scissors Black Plastic Handles Heavy-Duty, Autoclavable, Overall Length 7 (18 cm)- Intended to cut through tape and bandages Model: SU2014-001
Code information (models, lots, serial numbers)
UDI-DI: 10885403065866 Lot Numbers: L22XRD, C23XRD, E23XRD, and F23XRD.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- V. Mueller SU2014-001Steris Corporation · SCISSORS, GENERAL, SURGICAL1 recall0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page