XScribe Cardiopulmonary Kit recallZ-0707-2024 · Baxter Healthcare Corporation · initiated Dec 8, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential distortion identified in electrocardiogram (ECG) readings when the Source Consistency Filter (SCF) is enabled.

FDA root cause category: Software design

What the company is doing

Baxter issued an Urgent Medical Device Correction notice to its consignees on 12/08/2023 via USPS first class mail. The notice explained the problem, hazard involved, requested the following actions be taken until a software update becomes available: - Clinicians should evaluate the potential impact of the SCF as described to determine if it should be enabled or disabled during stress testing and provided the method for disabling the setting. - If you distributed this product to other facilities or departments within your institution, please forward a copy of this communication to them, informing them of the requirement to evaluate the potential impact of the SCF during stress testing. - Respond using the customer portal: https://BaxterFieldActionCustomerPortal.onprocess.com. - If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please notify your customers of this Urgent Medical Device Correction in accordance with your customary procedures and check the associated box on the customer portal. - If you purchased this product from a distributor, please note that the Baxter reply form is not applicable. If a reply form is provided by your distributor or wholesaler, please return it to your distributor/wholesaler according to their instructions. For general questions regarding this communication, contact Baxter (legacy Welch Allyn and Hillrom) Customer Service at 800-535-6663, Option 1, or Option 2 for Technical Support, between the hours of 8:00 am and 8:00 pm Eastern Time, Monday through Friday.

Recall details

Recall number
Z-0707-2024
Status
Open, Classified
Initiated by the company
Dec 8, 2023
Posted by the FDA
Jan 10, 2024
Recalling firm
Baxter Healthcare Corporation
Quantity
1675 units
Product code
DPS
Distribution
United States (including Puerto Rico), Canada, Australia, Bangladesh, Brunei, Chia, Hong Kong, India, Indonesia, Korea, Malaysia, New Zealand, Philippines, Singapore, Sri Lanka, Thailand, Algeria, Angola, Austria, Bahrain, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Hungary, Iceland, Iran, Iraq, Ireland, Israel, Italy, Jordan, Kazakhstan, Kenya, Kuwait, Latvia, Lebanon, Libya, Morocco, Namibia, Netherlands, Norway, Oman, Pakistan, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Switzerland, Tunisia, Turkey, United Arab Emirates, United Kingdom, Argentina, Bahamas, Bolivia, Brazil, Chile, Colombia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Honduras, Mexico, Panama, Paraguay, Peru, and Trinidad and Tobago.

Product description

XScribe CP Cardiac Stress Testing System, version 6, with below product names and codes: a) X-SCRIBE CPI KIT WIRELESS USB 608 MHz, 41000-030-50; b) X-SCRIBE CPI KIT WIRELESS USB 2500 MHz, 41000-030-51; c) X-SCRIBE CPI KIT DIRECT CONNECT USB, 41000-030-52; d) X-SCRIBE CPI KIT QUICK CONNECT USB, 41000-030-53; e) XSCRIBE CP KIT WIRELESS, 41000-030-60; f) XSCRIBE CP KIT WIRED, 41000-030-62; g) XSCRIBE CP SW KIT, 41000-030-64

Code information (models, lots, serial numbers)

Show the full code information (24k characters)

Product Code, UDI/DI Number, Serial Numbers a) 41000-030-50, UDI/DI None, Serial Numbers: 110120039437, 110120039438, 110120039439, 110190044556, 110190044557, 110190044558, 110190044559, 110190044560, 110190044561, 110190044562, 110190044563, 110190044564, 110190044565, 110190044566, 110190044567, 110200045831, 110200045832, 110200045833, 110200045834, 110200045835, 110260052818, 110260052819, 110260052820, 110260052821, 110260052822, 110260052823, 110420061166, 110420061167, 110420061168, 110420061169, 110460064753, 110460064754, 110460064755, 110460064697, 110460064698, 110460064699, 110460064700, 110460064702, 110460064703, 110460064704, 110460064705, 110460064706, 111150077803, 111150077804, 111150077805, 111150077806, 111150077807, 111150077808, 111150077809, 111200081334, 111200081335, 111200081336, 111200081337, 111200081338, 111200081339, 111200081340, 111200081341, 111300088621, 111300088622, 111300088623, 111300088624, 111300088625, 111300088626, 111300088627, 111340091242, 111340091243, 111360092140, 111360092141, 111360092142, 111360092143, 111360092144, 111340091246, 111340091247, 111340091248, 111340091249, 111340091250, 111420097839, 111420097840, 111420097841, 111420097842, 111420097843, 111500103083, 111500103084, 111500103085, 111500103086, 111530104948, 111530104949, 111530104950, 111530104951, 112020106363, 112020106364, 112020106365, 112020106366, 112020106367, 112020106368, 112020106369, 112020106370, 112180115629, 112180115630, 112180115631, 112180115632, 112180115633, 112180115634, 112180115635, 112200117094, 112200117095, 112200117096, 112200117097, 112200117098, 112200117099, 112200117100, 112200117101, 112200117102, 112200117103, 112320128027, 112320128028, 112320128029, 112320128030, 112320128031, 112350130930, 112350130931, 112350130932, 112350130933, 112350130934, 112350130935, 112350130936, 112350130937, 112440138921, 112440138922, 112440138923, 112440138924, 112440138925, 112440138926, 112440138927, 112440138928, 112440138929, 112440138930, 112440138931, 112440138932, 113010147090, 113010147091, 113010147092, 113010147093, 113070153116, 113070153117, 113070153118, 113070153119, 113070153120, 113070153121, 113070153122, 113070153123, 113070153124, 113070153125, 113070153126, 113070153127, 113070153128, 113070153129, 113270168996, 113270168997, 113270168998, 113270168999, 113270169000, 113310172113, 113310172114, 113310172115, 113310172116, 113350174038, 113350174039, 113350174040, 113350174041, 113350174046, 113350174047, 113350174042, 113350174043, 113350174044, 113350174045, 113410177775, 113410177776, 113410177777, 113410177778, 113410177779, 113410177780, 113410177781, 113440180951, 113440180952, 113440180953, 113470184299, 113470184300, 113470184301, 113530189429, 113530189430, 113530189431, 113530189432, 113530189433, 113530189434, 114050192901, 114050192902, 114050192903, 114090196065, 114090196066, 114090196067, 113460182788, 113460182789, 113460182790, 113460182791, 113460182792, 113460182793, 114420232656, 114420232657, 114420232658, 114420232659, 114420232660, 114420232661, 114480237409, 114480237410, 114420232662, 114420232663, 114420232664, 114420232665, 114520240984, 114520240985, 114520240986, 114520240987, 114520240988, 114520240989, 115050243901, 115050243902, 115050243903, 114470236646, 114470236647, 114470236648, 115230258572, 115230258573, 115230258574, 115230258575, 115230258576, 115230258424, 115230258425, 115230258426, 115280263454, 115280263455, 115280263456, 115280263457, 115280263458, 115280263459, 115280264206, 115280264207, 115280264208, 115280264209, 115280264210, 115280264211, 115330268899, 115330268900, 115330268901, 115330268902, 115330268903, 115330268904, 115400275479, 115400275480, 115400275481, 115400275482, 115400275483, 115400275484, 115400275485, 115400275486, 115400275487, 115400275488, 116020287198, 116020287199, 116020287200, 116100335156, 116100335157, 116100335158, 116130382423, 116130382424, 116130382425, 116130382426, 116130382427, 116180422043, 116180422044, 116180422045, 116180422046, 116180422047, 116240477718, 116240477719, 116240477720, 116240477721, 116270497993, 116270497994, 116270497995, 116270497996, 116380527123, 116380527124, 116380527125, 116380527126, 116380527127, 116380527128, 116380527129, 116380527130; b) 41000-030-51, UDI/DI None, Serial Numbers: 110120039425, 110120039426, 110120039427, 110190044568, 110190044569, 110190044570, 110200045836, 110200045837, 110200045838, 110330056675, 110330056676, 110370059009, 110370059010, 110370059011, 110390059824, 110420060941, 110420060942, 110500066898, 110500066899, 110500066900, 111060071714, 111060071715, 111060071716, 111150077462, 111150077463, 111200081342, 111200081343, 111270086404, 111270086405, 111270086406, 111270086407, 111380093553, 111380093554, 111380093555, 111380093556, 111380093557, 111470101214, 111470101215, 111500103087, 111500103088, 111500103089, 112030106627, 112030106628, 112050108028, 112050108029, 112100110764, 112100110765, 112050108026, 112050108027, 112220118711, 112220118712, 112220118713, 112270124286, 112270124287, 112270124288, 112320128032, 112320128033, 112320128034, 112410137135, 112410137136, 112410137137, 112440139667, 112440139668, 112440138933, 112440138934, 112440138935, 112440138936, 112440138937, 112440138938, 113020147730, 113020147731, 113020147732, 113020147733, 113020147734, 113020147735, 113020147736, 113020147737, 113130159216, 113130159217, 113130159218, 113130159219, 113130159220, 113130159221, 113130159222, 113130159223, 113130159224, 113130159225, 113130159226, 113130159227, 113130159228, 113130159229, 113130159230, 113450181907, 113450181908, 113480185307, 113480185308, 113450181644, 113450181645, 113450181646, 113450181647, 113450181648, 113450181649, 113450181650, 113450181651, 113450181652, 113450181653, 114380229638, 114380229639, 114440234317, 114440234325, 114470236649, 114470236650, 114480237351, 114480237352, 114480237353, 114480237354, 114480237349, 114480237350, 115060244042, 115060244043, 115100247551, 115100247552, 115100247553, 115140251901, 115140251902, 115140251903, 115140251904, 115140251905, 115140251906, 115140251907, 115140251908, 115200256104, 115200256105, 115230258565, 115230258566, 115230258567, 115230258568, 115240258971, 115240258972, 115240258973, 115240258974, 115200256100, 115200256101, 115200256102, 115200256103, 116020287201, 116020287202, 116050291262, 116050291263, 116100335159, 116100335160, 116140387942, 116140387943, 116140387944, 116180423551, 116180423552; c) 41000-030-52, UDI/DI None, Serial Numbers: 110120039363, 110120039364, 110230049218, 110230049219, 110230049220, 110230049221, 110280053721, 110280053722, 110280053723, 110280053724, 110280053725, 110330056793, 110330056794, 110370059012, 110370059013, 110370059014, 110410060508, 110410060509, 110410060510, 110410060511, 110440063042, 110440063043, 110440063044, 110440063045, 110460064750, 110460064751, 110460064752, 110460064707, 110460064708, 110460064709, 110460064710, 110460064711, 110460064712, 110460064713, 110460064714, 111060071710, 111060071711, 111060071712, 111060071713, 111200081535, 111200081536, 111200081537, 111230083963, 111230083964, 111230083965, 111330090559, 111330090560, 111330090561, 111330090562, 111330090563, 111330090564, 111330090565, 111450099468, 111450099469, 111450099470, 111450099471, 111450099472, 111450099473, 111450099474, 111450099475, 111460100779, 111460100780, 111460100781, 111490102399, 111490102400, 111490102401, 111460100426, 111460100427, 111460100428, 112020105816, 112020105817, 112020105818, 112020105819, 112020105820, 112020105821, 112130112152, 112130112153, 112130112154, 112130112155, 112130112156, 112130112157, 112280125032, 112280125033, 112280125034, 112280125035, 112280125036, 112280125037, 112250123314, 112250123315, 112250123316, 112250123317, 113100155571, 113100155572, 113130159201, 113130159202, 113130159203, 113130159204, 113130159205, 113130159206, 113130159207, 113130159208, 113130159209, 113130159210.113130159211, 113130159212, 113130159213, 113130159214, 113130159215, 114020190211, 114020190212, 114020190213, 114020190214, 114020190215, 114020190216, 114020190217, 114180209245, 114180209246, 114020190218, 114020190219, 114180209247, 114180209248, 114180209249, 114180209250, 114180209251, 114180209252, 114250218419, 114460236532, 114460236533, 114460236534, 114460236529, 114460236530, 114460236531, 114460236526, 114460236527, 114460236528, 115260260562, 115260260563, 115260260564, 115360271118, 115360271119, 115360271120, 115440278666, 115440278667, 115440278668, 115500283264, 115500283265, 115500283266, 116050290129, 116050290130, 116050290131, 116050290132, 116110349285, 116110349286, 116110349287, 116110349288, 116110349289, 116200433074, 116200433075, 116230452680, 116230452681, 116230452682, 116280504772, 116280504773, 116280504774; d) 41000-030-53, UDI/DI None, Serial Numbers: 110120039428, 110120039429, 110120039430, 110120039431, 110120039432, 110120039433, 110120039434, 110120039435, 110120039436; e) 41000-030-60, UDI/DI 00817655020051, Serial Numbers: 118060611436, 118060611438, 118060611439, 118080615013, 118080615022, 118180634934, 118080615011, 118180634932, 118180634933, 118180634936, 118060611447, 118080615017, 118080615018, 118080615021, 118080615019, 118080615020, 118180634938, 118180634939, 118180634940, 118160628568, 118160628569, 118160628570, 118180634931, 118180634935, 118160628577, 118160628578, 118160628579, 118160628580, 118160628581, 118160628583, 118160628564, 118160628565, 118160628566, 118160628567, 118160628571, 118160628572, 118160628573, 118160628582, 118160628574, 118160628575, 118160628576, 118340001391, 118340001392, 118340001393, 118340001394, 118340001395, 118340001396, 118340001397, 118340001388, 118340001389, 118340001390, 119110000129, 119110000130, 119110000132, 119110000133, 119110000134, 119110000135, 119110000136, 119110000137, 119130001327, 119130001328, 119130001331, 119110000128, 119110000131, 119130001329, 119130001330, 119170000064, 119170000066, 119170000067, 119170000068, 119190000633, 119190000634, 119190000635, 119190000629, 119190000630, 119190000636, 119190000637, 119170000054, 119170000061, 119190000628, 119190000631, 119190000632, 119170000055, 119170000056, 119170000062, 119170000063, 119170000065, 119170000057, 119170000058, 119170000059, 119170000060, 119240001197, 119240001198, 119340000071, 119340000072, 119340000073, 119340000074, 119340000075, 119340000076, 119340000077, 119340000078, 119340000079, 119340000080, 119350002675, 119350002676, 119350002677, 119350002678, 119350002679, 119350002680, 119350002681, 119350002682, 119350002683, 119350002684, 119380001144, 119380001145, 119380001146, 119380001147, 119380001148, 119380001149, 119430001324, 119430001325, 119430001326, 119430001327, 119430001328, 119430001329, 119430001330, 119430001331, 119430001332, 119430001333, 119470001107, 119470001108, 119470001109, 119470001110, 119470001111, 119470001112, 119470001113, 119470001114, 119470001115, 119470001116, 119500000427, 119500000428, 119500000429, 119500000430, 119500000431, 119500000432, 119500000433, 119500000434, 120080000436, 120080000437, 120080000438, 120080000439, 120080000440, 120080000441, 120080000442, 120080000443, 120080000444, 120080000445, 120080000446, 120080000447, 120080000448, 120080000449, 120080000450, 120080000451, 120080000452, 120080000453, 120080000454, 120080000455, 120260000436, 120260000437, 120260000438, 120370001869, 120370001870, 120370001872, 120260000439, 120260000440, 120370001871, 120370001873, 120370001874, 120370001875, 120370001876, 120370001877, 120370001878, 120410001420, 120410001421, 120410001422, 120410001423, 120410001424, 120410001425, 120410001426, 120410001427, 120480000382, 120480000383, 120480000384, 120480000385, 120480000386, 120480000387, 120480000388, 120480000389, 120480000390, 120480000391, 120480000392, 120480000393, 120480000394, 120480000395, 120520000178, 120520000179, 120520000180, 120520000181, 120520000182, 120520000183, 120520000184, 120520000185, 120520000186, 120520000187, 120520000188, 120520000189, 121060001022, 121060001023, 121060001024, 121060001025, 121060001026, 121060001027, 121060001028, 121060001029, 121060001030, 121060001031, 121080000918, 121080000919, 121080000920, 121080000921, 121080000922, 121080000923, 121080000924, 121130001105, 121130001106, 121130001107, 121130001108, 121130001109, 121130001110, 121130001111, 121130001112, 121130001113, 121130001114, 121130001115, 121130001116, 121130001117, 121130001118, 121130001119, 121140001520, 121140001521, 121140001522, 121140001523, 121140001524, 121170000316, 121170000317, 121170000318, 121170000319, 121170000320, 121170000321, 121170000322, 121170000323, 121170000324, 121170000325, 118050609197, 121210002553, 121210002554, 121210002555, 121210002556, 121210002557, 121210002558, 121210002559, 121210002560, 121210002561, 121210002562, 121250000651, 121250000652, 121250000653, 121250000654, 121250000655, 121250000656, 121250000657, 121250000658, 121250000659, 121250000660, 121250000661, 121250000662, 121300001879, 121300001880, 121300001881, 121300001882, 121300001883, 121300001884, 121300001885, 121300001886, 121300001887, 121300001888, 121320000722, 121320000723, 121320000724, 121320000725, 121320000726, 121320000727, 121320000728, 121320000729, 121390000580, 121390000581, 121390000582, 121390000583, 121390000584, 121390000585, 121390000586, 121390000587, 121390000588, 121390000589, 121390000590, 121390000591, 121430000457, 121430000458, 121430000459, 121430000460, 121430000461, 121430000462, 121430000463, 121430000464, 121430000465, 121430000466, 121430000467, 121430000468, 121450000127, 121450000128, 121460000618, 121490000525, 121490000526, 121490000527, 121490000528, 121490000529, 121490000530, 121490000531, 121490000532, 121490000533, 121490000534, 121490000535, 121490000536, 122130000539, 122130000540, 122130000541, 122130000542, 122130000543, 122130000544, 122130000545, 122130000546, 122190000029, 122190000030, 122190000031, 122190000032, 122190000033, 122190000034, 122200001921, 122200001922, 122200001923, 122200001924, 122200001925, 122200001926, 122200001927, 122200001928, 122200001929, 122200001930, 122200001931, 12200001930, 122240001466, 122240001467, 122240001468, 122240001469, 122240001470, 122240001471, 122230000033, 122230000036, 122230000039, 122230000040, 122230000042, 122230000044, 122230000037, 122230000038, 122230000041, 122230000043, 122230000034, 122260001174, 122260001175, 122260001176, 122260001177, 122260001178, 122260001179, 122280000876, 122280000877, 122280000878, 122280000879, 122280000880, 122280000881, 122340001204, 122340001205, 122340001206, 122340001207, 122390000370, 122390000371, 122390000372, 122390000373, 122390000374, 122390000375, 122420000196, 122420000197, 122420000198, 122420000199, 122420000200, 122420000203, 122420000192, 122420000193, 122420000194, 122420000195, 122420000201, 122420000202, 122450001786, 122450001787, 122450001788, 122450001789, 122450001790, 122450001791, 123100000063, 123100000069, 123100000070, 123100000071, 123100000064, 123100000065, 123100000066, 123100000067, 123100000068, 123130001066, 123130001067, 123210000710, 123210000711, 123340001553, 123340001554, 123320000551, 123320000552, 123370000077, 123370000078, 123370000079, 123370000080, 118050610039, 118180634937, 118050609198, 118180634937, 121450000127, 121450000128, 121460000618, 122230000033, 122230000034, 122230000035, 122230000036, 122230000037, 122230000038, 122230000039, 122230000040, 122230000041, 122230000042, 122230000043, 122230000044, 122500000829, 122500000830, 122500000831, 116480540886, 116480540887, 116480540885, 116480540888, 116480540889, 116480540890, 116480540891, 116500543972, 116530547195, 116530547196, 116530547197, 116530547198, 116530547199, 116530547200, 116530547201, 116530547207, 116530547208, 116530547209, 116530547202, 116530547203, 116530547204, 116530547205, 116530547206, 117050552991, 117050552992, 117140561292, 117140561293, 117140561294, 117140561295, 117140561296, 117140561297, 117140561299, 117140561300, 117140561301, 117140561302, 117190567099, 117190567100, 117140561298, 117190567101, 117190567102, 117190567103, 117190567104, 117190568321, 117190568322, 117190568323, 117210570066, 117250575592, 117250575593, 117250575594, 117250575595, 117250575596, 117250575597, 117260576261, 117260576262, 117260576263, 117250575598, 117250575599, 117250575600, 117250575601, 117310581178, 117310581179, 117350584646, 117350584647, 117350584648, 117350584649, 117350584650, 117350584642, 117350584643, 117400589264, 117400589265, 117400589266, 117400589267, 117400589268, 117400589293, 117400589294, 117400589295, 117400589296, 117400589297, 117400589298, 117400589299, 117400589269, 117400589270, 117400589271, 117400589272, 117400589273, 117400589274, 117430591637, 117430591638, 117430591639, 117430591640, 117430591980, 117430591981, 117490599622, 117490599623, 117490599624, 117490599625, 117490599627, 117490599628, 117490599630, 117490599619, 117490599620, 117490599621, 117490599626, 117490599629, 117490599631, 117490599632, 117490599633, 117510602851, 117510602852, 117510602853, 117510602854, 117520603151, 117520603152, 117520603153, 117520603154, 118030606983, 118030606984, 118030606985, 118030606986, 118030606987, 118030606988, 118030606989, 118030606990, 118030606991, 118030606992, 118050609193, 118050609194, 118050609195, 118050609196, 118050609199, 118050609200, 118050609201, 118050609202, 118060611433, 118060611429, 118060611430, 118060611431, 118060611432, 118060611434, 118060611435, 118060611437, 118060611440, 118080615016, 118080615015, 118080615014, 118080615012, 118080615010, 118080615009, 118080615008, 118080615007, 118080615006, 118080615005, 118080615004, 118080615003, 118060611448, 118060611446, 118060611445, 118060611444, 118060611443, 118060611442, 118060611441, 41000-030-60, 118180634941, 118180634942, 118340001398, 118340001399, 118340001400, 118340001401, 118340001402, 118340001403, 118340001404, 118340001405, 118340001406, 118340001407, 118340001408, 118340001409, 118340001410, 118340001411, 118340001412, 120180001027, 120180001028, 120180001029, 120180001030, 120180001031, 10081765502005100, 121040002235, 121040002236, 121040002237, 121040002238, 121040002239, 121040002240, 121040002241, 121040002242, 121040002243, 121040002244, 121050001674, 121050001675, 121050001676, 121050001677, 121050001678, 121050001679, 121050001680, 121050001681, 121050001682, 121050001683, 121080000925, 121080000926, 121080000927, 122200001932, 10081765502005100, 123390000815, 123390000816, 123390000817, 123390000818, 123400000181, 123400000182, 123400000183, 123400000184, 123420000969, 123420000970, 123420000971, 123420000972, 123430002087, 123440000214, 123440000215, 123440000216, 123440000217, 123450001456, 123450001457, 123450001458, 123450001459, 123480000269, 123480000270, 123480000271, 123480000272; f) 41000-030-62, UDI/DI 00817655020068, Serial Numbers: 118160628584, 118160628585, 118320000033, 118330000323, 118330000325, 118320000032, 118330000324, 118330000326, 118330000320, 118330000321, 118330000322, 118450000081, 118450000082, 118330000319, 118470000102, 118470000406, 118470000407, 118470000408, 118480000446, 118480000447, 118480000448, 118480000449, 118500001826, 118500001827, 118500001834, 118500001835, 118500001828, 118500001830, 118500001829, 118500001831, 118500001832, 119190000641, 119190000642, 119190000638, 119190000639, 119190000640, 118500001825, 118500001833, 119250001407, 119320001005, 119320001006, 119320001007, 119320001002, 119320001003, 119320001004, 119350002522, 119350002523, 119350002524, 119350002525, 119430001320, 119430001321, 119430001322, 119430001323, 119430001318, 119430001319, 119470002142, 119470002144, 119470002146, 119470002143, 119470002145, 119470002147, 120040000343, 120040000344, 120050001743, 120050001744, 120050001745, 120080000456, 120080000457, 120080000458, 120080000459, 120080000460, 120080000461, 120080000462, 120080000463, 120080000464, 120080000465, 120080000466, 120080000467, 120080000468, 120080000469, 120080000470, 120080000471, 120080000472, 120080000473, 120080000474, 120330000674, 120330000676, 120050001744, 120050001745, 120330000671, 120330000672, 120330000673, 120330000675, 120330000677, 120330000678, 120330000679, 120330000680, 120050001743, 120480000397, 120480000396, 120520000680, 120520000681, 120520000682, 121060001019, 121060001020, 121060001021, 121080000928, 121080000929, 121130001120, 121130001121, 121130001122, 121170000326, 121170000327, 121170000328, 121210002563, 121210002564, 121210002565, 121250000663, 121250000664, 121250000665, 121300001889, 121300001890, 121300001891, 121340000858, 121340000859, 121340000860, 121420001162, 121420001163, 121420001164, 121420001165, 121470000236, 121470000237, 121470000238, 121510001415, 121510001416, 121510001417, 122050001914, 122050001915, 122050001916, 122080000224, 122080000225, 122080000226, 122130000547, 122130000548, 122130000549, 122240001057, 122240001061, 122240001062, 122240001058, 122240001059, 122240001060, 122240001063, 122310000395, 122310000396, 122310000397, 122320000145, 122320000146, 122320000147, 122320000881, 122320000882, 122320000883, 122380000624, 122380000625, 122380000626, 122450001792, 122450001793, 122450001794, 122480001906, 122480001909, 123020001482, 123020001483, 123070001510, 123070001513, 123070001514, 123070001515, 123100001935, 123100001936, 123070001509, 123070001511, 123070001512, 123190000747, 123190000749, 123150002638, 123190000748, 123200000231, 123150002639, 123220001435, 123220001436, 123300001463, 123300001464, 123300000098, 123300000099, 123370000075, 123370000076, 122480001907, 122480001908, 123320000988, 123300001462, 123320000989, 123320000986, 123320000987, 120050001743, 120050001744, 120050001745, 120330000674, 120330000676, 120480000396, 113400177568, 116480540898, 116530547192, 116530547193, 116530547194, 117010549589, 117010549590, 117110556514, 117110556515, 117110556516, 117130560456, 117130560457, 117130560458, 117150562750, 117150562748, 117150562749, 117180567076, 117180567077, 117190568320, 117210570063, 117210570064, 117210570065, 117260575996, 117260575997, 117260575998, 117310581180, 117310581181, 117350584629, 117350584630, 117350584631, 117350584632, 117350584633, 117400589290, 117400589291, 117400589292, 117430592542, 117430592543, 117430592544, 117490599636, 117490599637, 117490599639, 117490599634, 117490599635, 117490599638, 118070612685, 118070612687, 118070612686, 118080614439, 118080614440, 118090617067, 118090617068, 118090617071, 118090617070, 118090617069, 118160628588, 118160628587, 118160628586, 118160628589, 118160628590, 118320000034, 118320000035, 118330000327, 118330000328, 118330000329, 120040000345, 120040000346, 120040000347, 120480000398, 120480000399, 120480000400, 120480000401, 120480000402, 120480000403, 120480000404, 120480000405, 120480000406, 120480000407, 120480000408, 123400000068, 123400000069; g) 41000-030-64, UDI/DI 00817655021423, Serial Numbers: 119490000440, 119500001335, 121350001043, 120070000730, 119490000439, 120070000732, 118380003022, 118380003023, 118470000103, 118470000104, 119500001336, 118380003022, 118380003023, 118470000103, 118470000104, 120070000731, 123150002068, 123150002069, 123150002070, 123160000556, 123150002055, 123200001669, 123150002057, 123150002058, 123200001670, 123200001671, 123150002066, 123150002067, 120070000733, 123150002056, 123150002061, 123150002062, 123150002063, 123150002064, 123150002065, 123150002059, 123150002060, 123160000552, 123160000553, 123160000554, 123160000555, 119500001335, 121350001043, 120070000730, 119490000439, 120070000732, 117320582006, 120070000734, 120070000735, 120070000736, 120070000737, 11847000104, 120070000738, 120070000739, 121350001044, 121350001045

Devices named in this recall

3 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93209).