Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS recallZ-0699-2024 · Medtronic Navigation, Inc. · initiated Nov 28, 2023
Class II Open, Classified All recalls posted in January 2024
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Navigation, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Their is the potential that the percutaneous pin may have a cross-pin that may render the pin unable to fit into tap cap, or too tight to remove.
FDA root cause category: Device Design
What the company is doing
On December 6, 2023, Medtronic Navigation, Inc. issued a Urgent: Medical Device Recall Notification via: UPS . Medtronic ask consignees to take the following actions: 1. Immediately locate and quarantine all unused impacted product(s). Refer to the affected lot numbers identified in Table 1 below. 2. Return the impacted product(s) to Medtronic following the instructions in the enclosed Customer Confirmation Form 3. Complete the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and email the completed form to Medtronic at [email protected]. If the affected devices have already been utilized and/or discarded, we still ask that you complete and return the Customer Confirmation Form detailing that information. 4. This notice should be distributed to all others in your organization who should be aware, or to any organization where the potentially affected devices have been transferred. Please maintain a copy of this notice in your records.
Recall details
- Recall number
- Z-0699-2024
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Nov 28, 2023
- Posted by the FDA
- Jan 10, 2024
- Recalling firm
- Medtronic Navigation, Inc.
- Quantity
- 4,224
- Product code
- OLO
- FDA submissions
- K131425
- Distribution
- Worldwide - US Nationwide distribution in the states of AK, AL, AZ, CA and the countries of Armenia, Australia, Austria, Belgium, Canada, China, Denmark, Finland, France, Germany, Hong Kong, Hungary, Ireland, Israel, Italy, Japan, Luxembourg, Netherlands, New Zealand, Oman, Portugal, Spain, Switzerland, United Kingdom.
Product description
Medtronic, REF 9733236, Pin, 150mm, STERILE, PERCUTANEOUS
Code information (models, lots, serial numbers)
UDI: 00613994247865 and 00643169105669(Japan)/ Serial Numbers: 2023041143, 2023051122, 2023051457, 2023051458, 2023060918, 2023051459, 2022030438, 2023041136, 2023041141
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- N/A 9733236Medtronic Navigation, Inc. · Orthopedic stereotaxic instrument2 recalls27 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93611).
- Z-0698-2024 · Open, ClassifiedMedtronic, REF 9733235, Pin, 100mm, STERILE, PERCUTANEOUS
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗