Sensis Vibe Hemo recallZ-0683-2024 · Siemens Medical Solutions USA, Inc · initiated Nov 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Medical Solutions USA, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The possibility of the Sensis documentation functionality application to crash.

FDA root cause category: Software design

What the company is doing

The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 11/29/23 by mail to consignees to notify them of this recall event. In the case this failure mode occurs during use, consignees are asked to continue the exam by opening a new study for the same patient without adding any Type 1 reporting events. It is instructed that consignees can compensate this by adding free text comments (procedure notes) only and manually inserting the desired information in those comments. The recalling firm will deploy a software update via Update Instruction AX010/23/S to correct the error. A member of the service organization will contact consignees to schedule the update. Consignees can contact the service organization at 1-800-888-7436 to schedule an earlier appointment. The recall notice is to be disseminated to all users of affected products within consignee organizations and others who may need to be informed. Consignees are to return the provided response form by email to [email protected] within 14 days of receipt.

Recall details

Recall number
Z-0683-2024
Status
Open, Classified
Initiated by the company
Nov 29, 2023
Posted by the FDA
Jan 9, 2024
Recalling firm
Siemens Medical Solutions USA, Inc
Quantity
331 units
Product code
DQK
FDA submissions
K223812
Distribution
Domestic: CA, CT, & FL.

Product description

Sensis Vibe Hemo

Code information (models, lots, serial numbers)

Model No. 11007641; UDI-DI 04056869010199; Serial No. 18009, 106245, 18016, 18014, 18011, 18010, 19001, 106244, 18015.