Skip to content

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part… recallZ-0642-2024 · Globus Medical, Inc. · initiated Dec 7, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Globus Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.

FDA root cause category: Device Design

What the company is doing

The firm notified affected customers through URGENT Medical Device Recall Notificatnoi letters dated December 7, 2023. Customers are instructed to immediately return the Flat Panel Fluoroscopy Registration Fixture. For questions related to this recall, please contact Globus Medical Complaint Handling Unit at [email protected].

Recall details

Recall number
Z-0642-2024
Status
Open, Classified
Initiated by the company
Dec 7, 2023
Posted by the FDA
Jan 18, 2024
Recalling firm
Globus Medical, Inc.
Quantity
9
Product code
OLO
FDA submissions
K211616
Distribution
Domestic distribution to the following states: CA, CO, MT, SC, WA. International distribution to Germany and UK.

Product description

ExcelsiusGPS Flat Panel Fluoroscopy Fixture Kit, Part number 6203.2100

Code information (models, lots, serial numbers)

UDI-DI 00193982465155 Serial Numbers FPFF-0101 FPFF-0102 FPFF-0103 FPFF-0104 FPFF-0105 FPFF-0106 FPFF-0107 FPFF-0108 FPFF-0109

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates