REV 2 Flowcell Capsule Spare, Part Number: C04154 recallZ-0786-2024 · Beckman Coulter, Inc. · initiated Dec 19, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Beckman Coulter has become aware that the cannula, which is found in the Flowcell and the iQClear Rinse Waste Well Assembly, may rust. Rust from the cannula has the potential to generate erroneous results.
FDA root cause category: Other
What the company is doing
Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/19/2023 via first class mail and email. The notice explained the issue, its impact on test results, and requested the following actions be taken: - Continue to run a complete Quality Control (QC) rack [including Iris Cleanser, Iris Diluent, iQ Focus and iQ Controls] as directed per iQ200 Series Instructions for Use (IFU) (300-4320 CE or 300-4321 EE) and DxU 850m and 840m Iris IFU (C49320 AB) - If the QC passes: o Run patient samples as routine - If the QC fails: o Refer to the IFU, Chapter 10 Troubleshooting; iQ200 Series IFU or DxU 850m and 840m Iris IFU o If the failure persists, contact your Beckman Coulter Representative - If a "Sequential Flags" Alarm displays: Refer to the IFU, Chapter 10; Alarm #25; iQ200 Series IFU or DxU 850 and 840m Iris IFU o If the alarm persists, contact your Beckman Coulter Representative Additional weekly action if using DxU 850m/840m Microscopy Analyzers: Visually inspect the iQClear Rinse Waste Well jets weekly for rust o If the iQClear shows rust on the rinse jets, contact your Beckman Coulter Representative Consult with your laboratory management to determine whether a retrospective review of results is necessary. For questions regarding this notice, please contact Customer Support Center or your local representative; From our website: http://www.beckmancoulter.com By phone: call 800-526-7694 in the United States and Canada. o Outside the United States and Canada, contact your local Beckman Coulter representative.
Recall details
- Recall number
- Z-0786-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 19, 2023
- Posted by the FDA
- Jan 25, 2024
- Recalling firm
- Beckman Coulter, Inc.
- Quantity
- 16 units
- Product code
- LKM
- FDA submissions
- K022774
- Distribution
- US and Australia, Brazil, Canada, China, France, Germany, Ghana, India, Israel, Italy, Lebanon, Lithuania, Mexico, Panama, Qatar, Romania, Saudi Arabia, Spain, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland,
Product description
REV 2 Flowcell Capsule Spare, Part Number: C04154
Code information (models, lots, serial numbers)
UDI/DI15099590689452, All lot numbers
Other devices under the same FDA submission
4 devices cleared under K022774 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- iQ200 Elite Plus (US) (2008 Version) (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls2 reports
- iQ200 Sprint Pro (NA)/SPRINT w load/unload (NA)/SPRINT Pro w load/unload (NA)/Sprint PBeckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls1 report
- iQ200 Select/Elite Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System3 recalls0 reports
- iQ200 Select Plus Kit 2008 Analyzer (NA)/ Select Pro (NA)Beckman Coulter Ireland Inc. · Automated Urinalysis System2 recalls4 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93655).
- Z-0784-2024 · Open, ClassifiedBeckman Coulter DxU 850m/840m Microscopy Analyzers, Part Numbers: C49513, C76947
- Z-0785-2024 · Open, ClassifiediQ200 Series Urine Microscopy Analyzers, Part Numbers: a) 700-3320, b) 700-3370, c) 700-3345, d) 700-3347, e) C10683, f) 700-3375, g) 700-3325, h) C10684
- Z-0787-2024 · Open, ClassifiediQ Waste Well Adapter, Part Number: 700-3393
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page