Skip to content

Endo-Model Knee System - Instruments recallZ-0749-2024 · Waldemar Link GmbH & Co. KG (Mfg Site) · initiated Nov 30, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Waldemar Link GmbH & Co. KG (Mfg Site) or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Mislabeled: Angle and height of the device marked as L(large) corresponds to the small/medium dimensions due to a product mix-up. This may result in too little bone material resected and not fit correctly leading to extended surgery time

FDA root cause category: Under Investigation by firm

What the company is doing

Link Bio Corp. issued Urgent Medical Device Recall Letter via email on 11/30/23. Letter states reason for recall, health risk and action to take: Please respond using the attached Distributor Reply form to indicate your understanding of this Device Recall and of the need to return the affected product. Pre-paid shipping labels will be provided once the Distributor Reply Form is returned. Please return the product as soon as possible to the following address: LinkBio Corporation 69 King St Dover, NJ 07801. Please contact LinkBio at 973-625-1333 or [email protected] for any questions.

Recall details

Recall number
Z-0749-2024
Status
Open, Classified
Initiated by the company
Nov 30, 2023
Posted by the FDA
Jan 18, 2024
Recalling firm
Waldemar Link GmbH & Co. KG (Mfg Site)
Quantity
1 unit
Product code
FZX
Distribution
US Nationwide distribution in the state of NH.

Product description

Link Patella Glide Resection Guide- For the implantation of the Endo-Model Knee System as part of a total knee arthroplasty procedure. Product Code: 15-2530/05

Code information (models, lots, serial numbers)

UDI-DI :04026575245253 Lot / Serial Number: C306165

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page