Accure Laser System, Model: PFMS00004 recallZ-0778-2024 · Accure Acne, Inc · initiated Dec 6, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Accure Acne, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Laser system's spot tracking system may become misaligned resulting in the possible risk of delivery of higher-than-expected laser energy to epidermal tissue, which may cause blistering.

FDA root cause category: Use error

What the company is doing

On 12/6/23, "Field Safety Alert(s)" were emailed to customers informing them of the following: 1) Following these steps will minimize the likelihood of Region of Interest (ROI) misalignment: Make sure that during treatment that the handpiece tip movement is minimized and to not lift off the tip until the emission is complete and foot pedal released. Remember this order of operations to reduce the potential of misalignment: "Tip down "Pedal down "Emission completes "Pedal up "Tip up 2) Monitor the patient during treatment for unexpected levels of pain. If higher than usual stop treatment and contact the recalling firm. 3) Immediately discontinue therapy if any Skin Overtemperature alarms are observed while treating with the device in Boost mode. This could be an indication that the ROI is misaligned. Contact your Accure representative before any further use of the device. 4) Immediately discontinue therapy if any PETs are observed that are above the green range displayed on the device touchscreen. This could be an indication that the ROI is misaligned. Contact your recalling firm service representative. 5) Ask your patients to contact you immediately if they experience prolonged erythema (longer than 2 hours) or blistering. Blistering is a potential residual risk and is not expected. If it occurs, it may require medical intervention. It could also be an indication that the ROI is misaligned. 6) If ROI misalignment is suspected suspend all further use of the laser and contact the recalling firm. The firm is releasing a software upgrade to maintain the ROI in the intended monitored location regardless of movement of the handpiece tip, and will safely detect and disable the system in the event that the ROI has become misaligned. Acknowledge, and send your response to the firm, see Field Safety Notice. Customers with questions can contact the firm via email at [email protected]

Recall details

Recall number
Z-0778-2024
Status
Open, Classified
Initiated by the company
Dec 6, 2023
Posted by the FDA
Jan 20, 2024
Recalling firm
Accure Acne, Inc
Quantity
22
Product code
GEX
FDA submissions
K222109
Distribution
US distribution to states of: CO, CA, CT, NY, NJ, MN, LA, OH, MA, TX

Product description

Accure Laser System, Model: PFMS00004

Code information (models, lots, serial numbers)

UDI-DI: 08059173392230, Serial Numbers: LAD0000-0004, LAD0000-0014, PROT01-0720, LAD0000-0012, LAD0000-0005, LAD2216-0623, LAD2356-0723, LAD0000-0003, PROTO07-1120, LAD2076-0623, LAD0000-0011, LAD0000-0008, LAD0000-0010, LAD2357-0723, LAD2075-0623, LAD2232-0723, LAD0000-0013, LAD0000-0009, LAD0000-0006, LAD0000-0001, PROT06-1120, LAD2507-0723