EXACTAMIX recallZ-1001-2024 · Baxter Healthcare Corporation · initiated Dec 22, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
An error was identified in software versions 2.0.8 and 2.1.8 while using the "Use Some Overfill" feature which may lead to over-delivery of an ingredient. This issue can result in redundant ingredient delivery of the requested overfill volume.
FDA root cause category: Device Design
What the company is doing
Baxter issued an Urgent Medical Device Correction notice to its consignees on 12/22/2023 via USPS First Class mail. The notice explained the issue with the device, potential risks, and requested the following actions be taken for the safe operation until a software upgrade is released: "1. Operators may continue to use the ExactaMix Pro compounders, but when the system prompts the operator that ingredients are depleted, operators should choose to Swap Container or Use a Different Container and not utilize the Use Some Overfill feature until further notice by Baxter. For more information on alternatives to using overfill, refer to the information on Replacing a Source Container on pages 82-83 in the ExactaMix Pro 2400 Operator Manual and pages 83-84 in the ExactaMix Pro 1200 Operator Manual. 2. If the Use Some Overfill feature is inadvertently used, operators must ensure that: o The Use Some Overfill feature is used only ONCE when an ingredient is depleted. o When prompted to enter the volume of overfill to use, enter LESS THAN the volume left to run stated in the Swap Container window. o On the MixCheck report, verify that all ingredient deliveries that used overfill volume delivered the expected ingredient volume and that the final total volume for the bag aligns with the expected volume. o If you observe redundant ingredient delivery using the overfill volume per the MixCheck report, discard the bag. 3. Always review the final MixCheck report as described in the ExactaMix Pro Operator Manual. Please review the information on this feature on pages 167-171 in the ExactaMix Pro 2400 Operator Manual and pages 164-168 in the ExactaMix Pro 1200 Operator Manual. 4. A Baxter representative will contact your facility when the software upgrade becomes available. Please note you will be receiving this upgrade from Baxter at no charge." If the affected product was distributed to other facilities or departments within your institution, please forward a copy of
Recall details
- Recall number
- Z-1001-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 22, 2023
- Posted by the FDA
- Feb 16, 2024
- Recalling firm
- Baxter Healthcare Corporation
- Quantity
- 8 devices
- Product code
- LHI
- FDA submissions
- K002705
- Distribution
- US: AL, AZ, CA, FL, GA, IA, KY, MD, MI, MN, MO, NC, NE, NH, NJ, NV, NY, OR, SD, TX, WA, WI,
Product description
Baxter Exactamix Pro 1200, REF EXM12DY
Code information (models, lots, serial numbers)
UDI/DI 05413765588150, Serial Numbers: PLD2314002, PLD2328001, PLD2328004, PLD2328042, PLD2314006, PLD2328002, PLD2328005, LD2339005
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EXACTAMIX EXM12DYBaxter Healthcare Corporation · System/device, pharmacy compounding1 recall0 reports
Product lines
Other devices under the same FDA submission
7 devices cleared under K002705 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EXACTAMIX H938724Baxter Healthcare Corporation · Set, i.V. Fluid transfer1 recall513 reports
- EXACTAMIX H938175Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING3 recalls263 reports
- EXACTAMIX H938173Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls260 reports
- EXACTAMIX H938174Baxter Healthcare Corporation · SYSTEM/DEVICE, PHARMACY COMPOUNDING2 recalls238 reports
- EXACTAMIX H938176Baxter Healthcare Corporation · Set, i.V. Fluid transfer2 recalls232 reports
- EXACTAMIX H938724EBaxter Healthcare Corporation · Set, i.V. Fluid transfer105 reports
- EXACTAMIX H938792Baxter Healthcare Corporation · Set, i.V. Fluid transfer12 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93700).
- Z-1002-2024 · Open, ClassifiedBaxter Exactamix Pro 2400, REF EXM24DY
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page