Flow-e Anesthesia System, model 6887900 recallZ-0991-2024 · Getinge Usa Sales Inc · initiated Dec 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Getinge Usa Sales Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential cybersecurity vulnerability that could lead to a Denial of Service (DoS) attack, tampering, or remote code execution via remote login to the Flow Anesthesia System, any of which could lead to a hazardous situation.
FDA root cause category: Software design
What the company is doing
An Urgent Medical Device Correction notice dated December 21, 2023 was issued via FedEx. The letter advises customers to examine inventory immediately to determine if you have any affected product. Directions are outlined for 1) Users with system version 4.8 and above with a Flow Anesthesia System connected to Connected Services directly via the Ethernet port on the Flow Anesthesia System (i.e., without a Getinge Connect module) and for 2) Users with system version 4.7 and below with Flow Anesthesia System connected to Remote Services via ethernet cable. Customers are asked to complete and return the MEDICAL DEVICE - CORRECTION RESPONSE FORM. Information should be forwarded to all current and potential system users with in your hospital / facility. Distributors are to forward the notification to customers for appropriate action. Contact your Getinge representative or call Getinge Customer Service at (888) 943-8872 (press option 2), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone) with questions.
Recall details
- Recall number
- Z-0991-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 11, 2023
- Posted by the FDA
- Feb 6, 2024
- Recalling firm
- Getinge Usa Sales Inc
- Quantity
- 86 (US)
- Product code
- BSZ
- FDA submissions
- K133958
- Distribution
- Worldwide - US Nationwide distribution in the states of AL, AR, AZ, CA, CO, DC, FL, GA, IN, MA, MI, MO, NY, OH, OR, PA, TX, VA, WA & WV.
Product description
Flow-e Anesthesia System, model 6887900
Code information (models, lots, serial numbers)
UDI-DI: 07325710010457, serial numbers (US): 50012, 50030, 50031, 50033, 50034, 50035, 50036, 50037, 50038, 50039, 50040, 50041, 50042, 50043, 50046, 50047, 50048, 50051, 50052, 50053, 50054, 50055, 50056, 50057, 50058, 50059, 50060, 50069, 50070, 50071, 50072, 50073, 50074, 50084, 50085, 50086, 50087, 50088, 50089, 50090, 50091, 50092, 50093, 50094, 50096, 50097, 50108, 50109, 50110, 50111, 50132, 50154, 50155, 50193, 50194, 50195, 50196, 50197, 50199, 50211, 50212, 50213, 50214, 50215, 50216, 50217, 50218, 50219, 50220, 50221, 50223, 50224, 50225, 50244, 50245, 50246, 50247, 50248, 50249, 50250, 50251, 50252, 50253, 50254, 50276, 50277.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Flow-e Anesthesia System 6887900Maquet Critical Care AB · Gas-Machine, Anesthesia5 recalls19 reports
Other devices under the same FDA submission
2 devices cleared under K133958 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vaporizer Sevoflur. Maquet filling 6682282Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
- Vaporizer Sevoflurane, QUIK FIL 6682285Maquet Critical Care AB · Gas-Machine, Anesthesia1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93736).
- Z-0987-2024 · Open, ClassifiedFlow-i C20 Anesthesia System, model 6677200
- Z-0988-2024 · Open, ClassifiedFlow-i C30 Anesthesia System, model 6677300
- Z-0989-2024 · Open, ClassifiedFlow-i C40 Anesthesia System, model 6677400
- Z-0990-2024 · Open, ClassifiedFlow-c Anesthesia System, model 6887700
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗