Ingenia recallZ-0974-2024 · Philips North America · initiated Jan 6, 2022
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.
FDA root cause category: Device Design
What the company is doing
Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.
Recall details
- Recall number
- Z-0974-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 6, 2022
- Posted by the FDA
- Feb 2, 2024
- Recalling firm
- Philips North America
- Quantity
- 2
- Product code
- LNH
- FDA submissions
- K173356, K173451
- Distribution
- Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.
Product description
Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Code information (models, lots, serial numbers)
UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Ingenia Ingenia Elition XPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance imaging14 recalls12 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 93737).
- Z-0973-2024 · Open, ClassifiedIngenia 1.5T (Product Number 781396) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
- Z-0975-2024 · Open, ClassifiedIngenia Ambition X (Product Number 781356) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.