Skip to content

Ingenia recallZ-0974-2024 · Philips North America · initiated Jan 6, 2022

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Magnetic materials were used for the wheels of trolley. Wheels with magnetic material might be attracted to MRI systems if the wheels are in the magnetic field.

FDA root cause category: Device Design

What the company is doing

Beginning in January 2022, Philips Field Service Engineers (FSE) visited impacted customer sites to replace the magnetic castor wheels of the HA Flextrack Trolley II with non-magnetic castor wheels. All reachable systems were addressed as of August 10, 2023.

Recall details

Recall number
Z-0974-2024
Status
Open, Classified
Initiated by the company
Jan 6, 2022
Posted by the FDA
Feb 2, 2024
Recalling firm
Philips North America
Quantity
2
Product code
LNH
FDA submissions
K173356, K173451
Distribution
Worldwide - US Nationwide distribution in the states of FL and NY. The countries of Denmark, Norway, and Sweden.

Product description

Ingenia Elition X (Product Number 781358) - HA FlexTrak II (Accessory Number 989710008732), patient transportation system

Code information (models, lots, serial numbers)

UDI-DI 00884838088115 Ingenia Serial number / Accessory Serial Number 45294 100029 45294 100030

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93737).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page