EXCELSIUS GPS recallZ-0997-2024 · Globus Medical, Inc. · initiated Dec 11, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Globus Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Sterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
FDA root cause category: Package design/selection
What the company is doing
Globus Medical issued Urgent Medical Device Recall Notification letter on 12/11/23. Letter states reason for recall, health risk and action to take: Customers are instructed to immediately return all affected sterile-packed Cranial Drills to Globus Medical and complete the attached Response Card. Report any adverse events believed to be associated with the affected sterile-packed Cranial Drills to Globus Medical immediately. For Questions and Support: Globus Medical Contact Contact Information Areas of Support Globus Medical Complaint Handling Unit Fax: 610-300-1342 Email: [email protected] Recall Related Questions
Recall details
- Recall number
- Z-0997-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 11, 2023
- Posted by the FDA
- Feb 6, 2024
- Recalling firm
- Globus Medical, Inc.
- Product code
- HAW
- FDA submissions
- K200047
- Distribution
- Worldwide - US Nationwide and the countries of Czech Republic, Poland.
Product description
Globus Medical Cranial Drill, 6.00mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1870S
Code information (models, lots, serial numbers)
Lot Number UDI: GBY094NA (01)00193982302320(17)280311(10)GBY094NA; GBY094MA (01)00193982302320(17)280311(10)GBY094MA; GBY094JA (01)00193982302320(17)280311(10)GBY094JA; GBY094HA (01)00193982302320(17)280311(10)GBY094HA; GBY093XA (01)00193982302320(17)280311(10)GBY093XA; GBY093VA (01)00193982302320(17)280311(10)GBY093VA; GBY094NA-RB (01)00193982302320(17)280311(10)GBY094NA-RB; GBY094MA-RB (01)00193982302320(17)280311(10)GBY094MA-RB; GBY094JA-RB (01)00193982302320(17)280311(10)GBY094JA-RB; GBY094HA-RB (01)00193982302320(17)280311(10)GBY094HA-RB; GBY093XA-RB (01)00193982302320(17)280311(10)GBY093XA-RB; GBY093VA-RB (01)00193982302320(17)280311(10)GBY093VA-RB
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EXCELSIUS GPS 6143.1870SGlobus Medical, Inc. · Neurological stereotaxic Instrument1 recall0 reports
Product lines
Other devices under the same FDA submission
3 devices cleared under K200047 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EXCELSIUS GPS 6143.1834SGlobus Medical, Inc. · Orthopedic stereotaxic instrument1 recall0 reports
- EXCELSIUS GPS 6143.1842SGlobus Medical, Inc. · Orthopedic stereotaxic instrument1 recall0 reports
- EXCELSIUS GPS 6143.2505Globus Medical, Inc. · Neurological stereotaxic Instrument1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 93742).
- Z-0995-2024 · Open, ClassifiedGlobus Medical Cranial Drill with Stop, 2.40mm - Instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref : 6143.1834S
- Z-0996-2024 · Open, ClassifiedGlobus Medical Cranial Drill with Stop, 3.20mm - instruments for access, preparation, and placement of cranial devices with the ExcelsiusGPS. Ref: 6143.1842S
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page