Skip to content

QC Sets and Panels: Helix Elite recallZ-1075-2024 · Microbiologics Inc · initiated Jan 8, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Microbiologics Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The negative control was contaminated with one of the positive control pathogens (Norovirus). This would result in users getting a positive call for Norovirus when running the negative control.

FDA root cause category: Process control

What the company is doing

The recalling firm issued letters dated 1/8/2024 beginning 1/8/2024 informing the consignee of the issue and providing the following instructions: (1) Review lab procedures to understand how this information affects your usage; (2) Use or discard depending on your lab procedures and how this information affects your usage; (3) Complete the response form provided; (4) Return the response form to the firm; (5) keep the letter for your records; and (6) contact Microbiologics if a replacement kit is needed. On 1/25/2024, the firm issued letters dated 1/24/2024 via email to notify consignees they have expanded the recall to include additional lot numbers. Other than providing additional lot numbers, the letter was the same as the initial letter issued on 1/8/2024. The firm reported the email that had the letter attached alerted the consignee that the recall was expanded to include additional lot numbers.

Recall details

Recall number
Z-1075-2024
Status
Open, Classified
Initiated by the company
Jan 8, 2024
Posted by the FDA
Feb 12, 2024
Recalling firm
Microbiologics Inc
Quantity
580 packages
Product code
OHQ
Distribution
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, MA, MI, MN, MO, MS, MT, NE, NH, NJ, NY, OH, PA, RI, TN, TX, VA, WA, WI, WV, and WY. There was government distribution and no military distribution. Foreign distribution was made to France.

Product description

Microbiologics Enteric Pathogens Control Panel (Inactivated Pellet), REF 8184.

Code information (models, lots, serial numbers)

Lot #8184-68, Exp. 5/31/2025, UDI-DI 70845357041448. UPDATE: The recall was expanded on 1/23/2024 to include the following additional lot numbers: 8184-65, exp. 3/31/2025; 8184-66, exp. 4/30/2025; 8184-67, exp. 5/31/2025; 8184-69, exp. 6/30/2025; 8184-70, exp. 8/31/2025; and 8184-71, exp. 8/31/2025.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page