Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296 recallZ-0945-2024 · Philips North America · initiated Jan 3, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The Quadrature Body Coil (QBC) seal adhesive may fail, creating sharp edges that may come in contact with patients. If the QBC seal becomes loose during the scanning process, the risk to the patient may include skin abrasions, bruises, lacerations, hair loss/entanglement, and tissue injury.
FDA root cause category: Process control
What the company is doing
On January 4, 2024, the firm began notification of affected customers via URGENT Medical Device Correction letters. Customers were provided with precautions to take when using the device, including inspection of the QBC seal for separation. If the QBC seal is found to be loose, or becomes loose during a patient scan, customers should immediately stop use and contact their local Philips service representative. Customers should also ensure that all users of the device are aware of the issue and retain the letter with affected systems until a solution is installed. Philips will contact customers to schedule a time for a Field Service Engineer (FSE) to visit your site and replace your system's QBC Seal. Philips plans to start implementing corrections in Q3 2024. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377). Revision B of the URGENT Medical Device Correction letter was distributed to the additional US consignees on 24-Mar-2025. It was later identified that UMDC Rev B contained three MR systems with a 70cm bore (Ingenia 1.5T: 781341, Ingenia 3.0T: 781377, Ingenia Ambition X: 782109) which were incorrectly included, as the scope of this field action is MR systems with 60cm bore. The URGENT Medical Device Correction letter was updated to Rev C to remove these systems. Revision C of the UMDC was distributed to the additional US consignees on 09-Apr-2025.
Recall details
- Recall number
- Z-0945-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 3, 2024
- Posted by the FDA
- Feb 1, 2024
- Recalling firm
- Philips North America
- Quantity
- 119 US; 1,807 OUS
- Product code
- LNI
- FDA submissions
- K193215
- Distribution
- Worldwide - US Nationwide distribution.
Product description
Achieva 1.5T. Model (REF) Numbers 781196, 781343, 781296.
Code information (models, lots, serial numbers)
a. Model No. 781196; UDI-DI N/A. b. Model No. 781343; UDI-DI 00884838055292. c. Model No. 781296; UDI-DI 00884838004108. All units are affected.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Achieva Achieva 1.5TPhilips Medical Systems Nederland B.V. · System, nuclear magnetic resonance spectroscopic3 recalls10 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93787).
- Z-0950-2024 · Open, ClassifiedIngenia 1.5T CX. Model (REF) Numbers 781262, 781261.
- Z-0951-2024 · Open, ClassifiedIngenia 3.0T CX. Model (REF) Numbers 781271, 782105.
- Z-0955-2024 · Open, ClassifiedSmartPath to dStream for 1.5T. Model (REF) Numbers 781260, 782112.
- Z-0956-2024 · Open, ClassifiedSmartPath to dStream for XR and 3.0T. Model (REF) Numbers 781270, 782113, 782129.
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page