Skip to content

Box as labeled: Regard, Carter-Thomason CloseSure System… recallZ-0985-2024 · CooperSurgical, Inc. · initiated Aug 29, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from CooperSurgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Incorrect component descriptions within the "Contents" section on the Tyvek lid for individual kit trays.

FDA root cause category: Labeling Change Control

What the company is doing

An Urgent Medical Device Field Safety Notice dated August 29, 2023 was sent to the one distributor receiving shipments of the affected product lot. Customers are advised to complete the acknowledgement form, Affected product is to be returned to CooperSugical for credit. Contact CooperSurgical at +1-203-601-5200 extension 3300 during normal operating hours of 09:00 - 17:00 M-F EST, email [email protected].

Recall details

Recall number
Z-0985-2024
Status
Open, Classified
Initiated by the company
Aug 29, 2023
Posted by the FDA
Feb 5, 2024
Recalling firm
CooperSurgical, Inc.
Quantity
127 boxes, 3 kits per box (381 kits total)
Product code
GCJ
FDA submissions
K980123
Distribution
US Nationwide distribution in the state of MO.

Product description

Box as labeled: Regard, Carter-Thomason CloseSure System XL, REF 733776, Quantity 1 Box of 3 Systems. Kit as labeled: Regard, Carter-Thomason CloseSure System, REF 733776, Contents: One (1) Carter-Thomason Suture Passer, One (1) 5 mm Pilot Guide, One (1) 10/12 mm Pilot Guide.

Code information (models, lots, serial numbers)

UDI-DI: 60888937027053, lot# 326728.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K980123 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page