WATER FILTER recallZ-1009-2024 · Olympus Corporation of the Americas · initiated Jan 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.
FDA root cause category: Process control
What the company is doing
On January 16, 2024, the firm notified customers of the issue via URGENT MEDICAL DEVICE FIELD CORRECTIVE ACTION letters. Customers were instructed to immediately assess all MAJ-2318 at their facility to check the model and lot number. Customers were given detailed instructions on how to identify affected product, including for MAJ-2318 installed in the OER machine. Please contact your Olympus customer service representative at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return any affected product. Customers who have further distributed product should notify downstream accounts at once. If you require assistance with identifying affected product, please contact the Olympus Technical Assistance Center at 1- 800-848-9024, option 1. If you have questions about this recall, please contact the recalling firm at (647) 999-3203 or by e-mail at [email protected].
Recall details
- Recall number
- Z-1009-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 16, 2024
- Posted by the FDA
- Feb 7, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 877 MAJ-2318, 119 OER-Elite, 3 OER-MINI
- Product code
- FEB
- FDA submissions
- K103264
- Distribution
- Domestic US distribution nationwide.
Product description
Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope reprocessor for prevent contamination of the rinse water. The MAJ-2318 water filter is compatible with OER-ELITE and OER-MINI.
Code information (models, lots, serial numbers)
UDI-DI 04953170401473 Lots 36A, 37A, 38A Affected product may have also been packaged within OER boxes (OER-Elite and OER-MINI)
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- WATER FILTER MAJ-2318Olympus Medical Systems Corp. · Accessories, Cleaning, For Endoscope1 recall33 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.