Olympus recallZ-1167-2024 · Olympus Corporation of the Americas · initiated Dec 15, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The jaw may fracture prior to the procedure during the inspection instructed per the IFU, or during the procedure.
FDA root cause category: Under Investigation by firm
What the company is doing
On January 15, 2024, the firm notified customers of the issue via URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect your inventory and identify any products of the model and lots subject to this action. Please check all areas of the hospital to determine if any of these devices remain in inventory. Quarantine and cease use of the affected model/lots. Customers were given detailed instructions on how to identify affected product. Please contact your Olympus customer service representative at 1-800-848-9024, option 2 with the quantity, model and lot number of the affected device(s). Olympus will issue a Return Material Authorization to return any affected product at no charge. Olympus will issue a credit to your facility upon return of the affected product. Customers who have further distributed product should notify downstream accounts at once. If you require assistance with identifying affected product, please contact the Olympus Technical Assistance Center at 1- 800-848-9024, option 1, and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. If you have questions about this recall, please contact the recalling firm at (647) 999-3203 or by e-mail at [email protected].
Recall details
- Recall number
- Z-1167-2024
- Status
- Open, Classified
- Initiated by the company
- Dec 15, 2023
- Posted by the FDA
- Feb 20, 2024
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 11,123 units
- Product code
- GEI
- FDA submissions
- K203277
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Mexico, P.R. China, Singapore, South Korea.
Product description
ESG PK CUTTING FORCEPS, 5MM, 33CM
Code information (models, lots, serial numbers)
Model: PK-CF0533; UDI DI: 00821925035867; Lot Numbers: FR258486, FR261738, FR259544, FR263183, FR276278, FR279757, FR279625, FR279658, FR286369, FR287112, FR374622, FR207935, FR212413, FR212735, FR215061, FR216817, FR220525, FR222456, FR227254, FR227774, FR227781, FR232882, FR234845, FR247614, FR240430, FR246627, FR247039, FR248008, FR255679, FR248312, FR263636, FR263244, FR269280, FR269267, FR269269, FR287114, FR303799, FR297022, FR301062, PWO-308561, PWO-308562, PWO-308563, PWO-308560, PWO-308567, PWO-308566, PWO-308565, FR303864, FR313639, FR316869, FR319703, FR316872, FR327890, PWO-308972, FR335984, FR335990, FR337947, FR373912, FR401646, FR379790, FR347006;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Olympus PK-CF0533Gyrus ACMI, LLC · ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES2 recalls67 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.