INTERVENTIONAL ANGIOGRAPHY SYSTEM recallZ-1155-2024 · Canon Medical System, USA, INC. · initiated Nov 15, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.

FDA root cause category: Component change control

What the company is doing

On 11/15/23 correction notices were distributed to customers who were asked to do the following: 1) Should this issue occur please attempt to restart the system and contact your service representative. 2) A recalling firm representative will contact you to schedule a time to replace the flat panel detector on the console. The parts required to address the defect are expected to be available no earlier that December 2023. 3) Complete and return the customer reply form via email to [email protected] For questions, please contact our InTouch team at [email protected] or 800-421-1968, or 800.521.1968.

Recall details

Recall number
Z-1155-2024
Status
Open, Classified
Initiated by the company
Nov 15, 2023
Posted by the FDA
Feb 16, 2024
Recalling firm
Canon Medical System, USA, INC.
Quantity
2
Product code
OWB
FDA submissions
K182415
Distribution
US: CA, LA

Product description

The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.

Code information (models, lots, serial numbers)

UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93844).

Sources and updates