Skip to content

Actim PROM dipstik, REF 30831ETUS recallZ-1592-2024 · Actim Oy · initiated Nov 23, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Actim Oy or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Interference testing has shown that the presence of personal lubricants in the vaginal specimen can interfere with the PROM test result by providing a false positive result.

FDA root cause category: Under Investigation by firm

What the company is doing

Actim Oy issued a MEDICAL DEVICE RECALL notice to its sole distributor on 11/23/2023 via email. The notice explained the issue, potential risk, and requested that users refrain from using surgical lubricants when collecting the vaginal specimen and that vaginal specimens do not contain significant amounts of personal lubricant from patient's recent personal use. The instructions for use will be updated according to the latest test results of surgical and personal lubricant interference. In the interim, a separate note will be included in the test kit regarding the correction of the interference claims and guidance of lubricant use.

Recall details

Recall number
Z-1592-2024
Status
Open, Classified
Initiated by the company
Nov 23, 2023
Posted by the FDA
Apr 17, 2024
Recalling firm
Actim Oy
Quantity
6471 kits
Product code
OAM
FDA submissions
K123986
Distribution
US Nationwide distribution in the state of Connecticut.

Product description

Actim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membrane

Code information (models, lots, serial numbers)

UDI/DI 06430030990608, All serial numbers in distribution

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page