Skip to content

Medline Kits, trays, and packs labeled as follows recallZ-1113-2024 · MEDLINE INDUSTRIES, LP - Northfield · initiated Nov 15, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline Industries, LP is recalling certain kits and trays that were manufactured using specific lots of Nurse Assist (as components): 0.9% Sodium Chloride Irrigation USP, Sterile Water for Irrigation USP and Saline Flush Syringe. Nurse Assist issued a recall due to the potential lack of sterility, which could result in the solution being non sterile.

FDA root cause category: Nonconforming Material/Component

What the company is doing

MEDLINE issued an IMMEDIATE ACTION REQUIRED NURSE ASSIST PRODUCT RECALL notice to its consignees on 11/16/2023 via USPS first class mail and email. The notice explained the problem, requested log on to their portal to identify the products and check their inventory so the product could be quarantined. Once Medline receives the completed response form, over-labels will be shipped to the consignee. Medline Industries, LP instructed customers to over-label the convenience kits with a warning label indicating to remove the sodium chloride irrigation or Sterile water and replace from a different sterile supply. Distributors or consignees who further distributed the affected products were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-1113-2024
Status
Open, Classified
Initiated by the company
Nov 15, 2023
Posted by the FDA
Feb 25, 2024
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Quantity
3900 units
Product code
LRS
Distribution
Distribution US nationwide including Puerto Rico, Canada, UAE, and Panama

Product description

Medline Kits, trays, and packs labeled as follows: a) PORT ACCESS TRAY, REF DYNDA2863; b) PORT ACCESS TRAY, REF DYNDA2863H; c) ULTRASOUND GUIDED IV START KIT, REF DYNDV2518; d) ULTRASOUND GUIDED IV START KIT, REF DYNDV2518H; e) KIT ULTRASOUND PIV START, REF DYNDV2445; f) NICU INSERTION & DRSG CHG TRAY, REF EBSI1306A; g) PIV BSI KIT, REF BSIPIV1010; h) ULTRASOUND GUIDED IV START KIT, REF DYNDV2465

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

a) REF DYNDA2863, UDI/DI 40195327008025 (case), 10195327008024 (unit), Lot Numbers: 22ABQ171, 22ABT619, 22IBG373; b) REF DYNDA2863H, UDI/DI 40195327008025 (case), 10195327008024 (unit), Lot Numbers: 22ABQ171, 22ABT619, 22IBG373; c) REF DYNDV2518, UDI/DI 40195327089079 (case), 10195327089078 (unit), Lot Numbers: 22GBA169, 22HBU208; d) REF DYNDV2518H, UDI/DI 40195327089079 (case), 10195327089078 (unit), Lot Numbers: 22GBA169, 22HBU208; e) REF DYNDV2445, UDI/DI 20193489110576 (case), 10193489110579 (unit), Lot Numbers: 2022012480; f) REF EBSI1306A, UDI/DI 00653160352122 (case), 10653160352129 (unit), Lot Numbers: 2022042850; g) REF BSIPIV1010, UDI/DI 00653160349153 (case), 10653160349150 (unit), Lot Numbers: 2022020750, 2022040480, 2022082980; h) REF DYNDV2465, UDI/DI 20193489110163 (case), 10193489110166 (unit), Lot Numbers: 2022031480, 2022041880

Devices named in this recall

7 devices in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 93800).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page