Zyno Medical Z800 Large Volume Infusion Pumps -Intended to… recallZ-1183-2024 · Zyno Medical LLC · initiated Dec 20, 2023

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Zyno Medical LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When utlizing the patient query feature on the Zyno Medical Z-800WF pumps with software version 5.2.05, the alarm volume may inadvertently revert from a higher volume setting to a low one, causing potential delays in therapy and in extreme cases, associated risk of organ failure or death

FDA root cause category: Software design

What the company is doing

Zyno Medical LLC issued URGENT: Medical Device Correction Zyno Medical Z-800WF Infusion Pump -Alarm Volume Reversion Issue letter on 12/20/23 via Certified Mail. Letter states reason for recall, health risk and action to take: To address this issue, users can reset the alarm volume by accessing the Factory Settings (2) via the configuration menu on the pump and selecting the desired alarm volume. The chosen setting will remain on the pump until it is manually altered or until the initial workflow that triggered the reversion is repeated. Zyno Medical LLC is mitigating this alarm issue by conducting a voluntary medical device correction to downgrade the software from version 5.2.05 to version 5.1.87. Please note that the low volume alarm issue occurred with a specific software change that occurred in version 5.2.05 and the issue does not exist. Complete the attached Zyno Medical Z-800WF Infusion Pump - Alarm Volume Reversion Issue -Response Verification Form. This will include checking your inventory for affected products, verifying serial numbers, confirming product location, reporting any transferred product, and signing and dating the completed form. 3. A representative from Zyno Medical's distribution partner, lntuvie LLC, will be in contact with you to coordinate instructions on exchanging devices and the timing to remediate this action. 4. Pass this notice to the appropriate healthcare professional within your facility who needs to be informed and to any facility where the potentially affected devices have been transferred (if appropriate). Provide Zyno Medical with details of any affected devices that have been transferred to other facilities. If you have any questions or would like assistance, please contact your local Business Development Manager

Recall details

Recall number
Z-1183-2024
Status
Open, Classified
Initiated by the company
Dec 20, 2023
Posted by the FDA
Feb 23, 2024
Recalling firm
Zyno Medical LLC
Quantity
574 units
Product code
FRN
FDA submissions
K130690
Distribution
Nationwide

Product description

Zyno Medical Z800 Large Volume Infusion Pumps -Intended to provide intravenous infusion of parenteral fluids, blood and blood products to a patient under the direction or supervision of physician or other certified health care professional Model: Z-800WF

Code information (models, lots, serial numbers)

UDI: 00814371020037 software version 5.2.05

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K130690 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page