Skip to content

SUR-FIT NATURA recallZ-1276-2024 · ConvaTec, Inc · initiated Jan 17, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ConvaTec, Inc or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The kit contains a flange that has a 45mm coupling ring, however, the pouch within the kit is for a 57mm coupling ring.

FDA root cause category: Labeling Change Control

What the company is doing

The recalling firm issued letters dated Jan-2024 via email on 1/17/2024. The letter explained the issue, contained an embedded photo of the label for ease in identifying the product, and requested the consignee immediately examine their inventory and quarantine the affected product. If the consignee has further distributed the affected product, they are requested to notify them at once of the recall. A response form was enclosed for completion and return as soon as possible or within 30 days of receipt. A section of the letter listed the consignee's responsibilities. The responsibilities were: (1) Review this notification and ensure that all relevant stakeholders are aware of this recall communication; (2) Immediately locate and quarantine affected product in your inventory; (3) Complete APPENDIX 1 (Response Form) and return to Convatec as soon as possible or within 30 calendar days of receipt; (4) A review of the information you provide in accordance with APPENDIX 1 will be completed; (5) If the information from Step 4 is satisfactory, formal authorization from a Convatec Representative to proceed with product destruction will be provided; (6) Convatec will provide you with a Certificate of Destruction (COD) to complete; (7) Immediately destroy all affected product and provide Convatec with a signed Certificate of Destruction (COD) as evidence to support reconciliation and regulatory reporting; (8) Your account will be credited for all destroyed product upon receipt of a signed form as per APPENDIX 1 and a Certificate of Destruction (COD); and (9) If you are in receipt of this recall communication and are still unsure how to proceed, please contact Convatec via [email protected] or phone 800-582-6514.

Recall details

Recall number
Z-1276-2024
Status
Open, Classified
Initiated by the company
Jan 17, 2024
Posted by the FDA
Sep 30, 2024
Recalling firm
ConvaTec, Inc
Quantity
1,715 units
Product code
FON
Distribution
Distribution was made to AK, CA, FL, IL, IN, MI, NE, NJ, NY, OH, PA, and TX. There was government distribution but no military distribution. Foreign distribution was made to Canada.

Product description

ConvaTec Natura Stomahesive Post-Operative / Surgical Ostomy System, REF 416938, 1 3/4 in. 45mm, Sterile.

Code information (models, lots, serial numbers)

Lot #2C02496, UDI-DI #00768455116096.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.