Skip to content

VITROS recallZ-1221-2024 · Ortho-Clinical Diagnostics, Inc. · initiated Jan 22, 2024

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Ortho-Clinical Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential For Aspiration from Unintended Sample Container During Sampling Center Processing on VITROS 3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems which may lead to erroneous test results.

FDA root cause category: Component change control

What the company is doing

An URGENT PRODUCT CORRECTION NOTIFICATION letter dated 01/22/24 was sent to customers. REQUIRED ACTIONS: To reduce the risk of unintended sample aspiration, refer to the procedure provided in this notification titled, Temporary TRAY Loading Instructions . QuidelOrtho recognizes that the Temporary TRAY Loading Instructions procedure may not be tenable for all laboratories due to testing volumes. Review your laboratory risk management plan(s) to assess the need for this further mitigation. Acknowledge your understanding of this notification by completing the enclosed Confirmation of Receipt form no later than January 30, 2024. Save this notification with your User Documentation or post this notification by each VITROS 3600/5600/XT 7600 System until this issue has been resolved. If your laboratory has experienced the anomaly described in this notification and you have not already done so, please report the occurrence to your local Global Services Organization (formerly Ortho Care). QuidelOrtho has identified root cause however, our investigation is still ongoing. We are currently working on a Hardware Modification (MOD) to resolve this issue and will communicate again once the MOD or additional information is available. Current estimated availability date of the MOD is 2nd quarter 2024. If you have further questions, please contact our Global Services Organization at 1-800-421-3311.

Recall details

Recall number
Z-1221-2024
Status
Open, Classified
Initiated by the company
Jan 22, 2024
Posted by the FDA
Mar 1, 2024
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Quantity
1460 units (506 US, 954 OUS)
Product code
JJE
FDA submissions
K081543
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, United Kingdom.

Product description

VITROS 5600 Integrated System (New and Refurbished)

Code information (models, lots, serial numbers)

UDI/DI: (New) 10758750002740, (Refurbished) 10758750007110; Serial Numbers: J56000280, J56000330, J56000415, J56000631, J56000634, J56000721, J56000839, J56000871, J56001088, J56001209, J56001228, J56001265, J56001305, J56001355, J56001834, J56001855, J56001941, J56001998, J56002111, J56002239, J56002476, J56002492, J56002531, J56002561, J56002649, J56002655, J56002658, J56002691 and J36001164 through J36001692 (inclusive).

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 93892).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page