MIC* Gastric-Jejunal Feeding Tube, Endoscopic/Radiologic… recallZ-1187-2024 · Avanos Medical, Inc. · initiated Jan 16, 2024
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Avanos Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
MIC* Gastric - Jejunal Feeding Tube Kits contain a sterile pre-filled syringe subsequently recalled by Nurse Assist.
FDA root cause category: Nonconforming Material/Component
What the company is doing
AVANOS MEDICAL, INC. issued an "URGENT MEDICAL DEVICE RECALL" notice to its consignees on 01/16/2024 via FedEx. The notice explained the issue, the potential risk to health, and requested the following actions be taken: " Please continue to perform normal post op care for any newly placed MIC GJ tube. Per the Daily Care & Maintenance Check List found in the product Instructions for Use: o Assess the patient for any signs of pain, pressure or discomfort. o Assess the stoma site for any signs of infection, such as redness, irritation, edema, swelling, tenderness, warmth, rashes, purulent, or gastrointestinal drainage. o Assess the patient for any signs of pressure necrosis, skin breakdown, or hypergranulation tissue. " Report any adverse events involving these products immediately to [email protected] " Check all storage facilities and warehouse locations for affected products and quarantine them. " Complete the Customer Response Form. which can be found online: https://iqvia-response.my.site.com/mt/fca?cid=avanos-fca-2024-001, or by using the QR code provided in the Field Correction Acknowledgement Instructions accompanying this letter. " Destroy all affected lots in your inventory per your facility s procedures o Alternatively if you are unable to destroy your affected inventory, please indicate so on the form and you will be provided information on how to return product for destruction. o A refund will be provided to you based on the quantity of affected product you have indicated on the Customer Response Form. " Distributors only: alert your affected customers of this recall. We ask that you work with them to destroy affected product and provide them refunds as per your internal processes. Provide to Avanos an updated response using the Customer Response Form to receive refunds or credit for your customers affected products. " Please reach out to [email protected] or call (855) 201-1355 with any questions or concerns regarding these instructions.
Recall details
- Recall number
- Z-1187-2024
- Status
- Open, Classified
- Initiated by the company
- Jan 16, 2024
- Posted by the FDA
- Mar 1, 2024
- Recalling firm
- Avanos Medical, Inc.
- Quantity
- 53126 Units
- Product code
- KNT
- FDA submissions
- K921370
- Distribution
- Worldwide distribution: US (nationwide) to states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY including PR; and OUS (foreign) countries of: Canada, Colombia, Costa Rica, UAB, Israel, Kuwait, and Panama.
Product description
MIC* Gastric-Jejunal Feeding Tube, Endoscopic/Radiologic Placement, GJ-Tube: a) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16; b) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-15; c) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-22; d) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 16 Fr, Product Code 0250-16-30; e) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18; f) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18-22; g) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 18 Fr, Product Code 0250-18-30; h) MIC* Gastric-Jejunal Feeding Tube Kit - Endoscopic/Radiologic Placement - 22 Fr, Product Code 0250-22; i) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 16 Fr, Product Code 0260-16; j) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 18 Fr, Product Code 0260-18; k) MIC* Gastric-Jejunal Feeding Tube Kit - Surgical Placement - 22 Fr, Product Code 0260-22;
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
a) Product Code 0250-16, UDI/DI 00350770954806, Batch Numbers: 30180718, 30181870, 30184758, 30186768, 30194327, 30226669, 30226817, 30229831, 30240299, 30244886, 30248036, 30248348, 30251106, 30251988, 30254746, 30258134, 30267592; b) Product Code 0250-16-15, UDI/DI 00350770954813, Batch Numbers: 30185370, 30194326, 30249243, 30256710; c) Product Code 0250-16-22, UDI/DI 00350770954820, Batch Numbers: 30226808, 30238975; d) Product Code 0250-16-30, UDI/DI 00350770954837, Batch Numbers: 30184813, 30186749, 30190699, 30226816, 30235093, 30245070; e) Product Code 0250-18, UDI/DI 00350770954844, Batch Numbers: 30181844, 30181869, 30184757, 30185369, 30186748, 30189492, 30190698, 30195869, 30226668, 30226815, 30228967, 30229830, 30238122, 30238885, 30239857, 30238982, 30240298, 30240451, 30243197, 30242224, 30244885, 30250139, 30251984, 30254745, 20108481, 30256210, 30256482, 30257069, 30258133, 30259563, 30261698, 30262037, 30264281, 30267591, 30268254, 30270503, 30270858; f) Product Code 0250-18-22, UDI/DI 00350770954851, Batch Numbers: 30186747, 30188504, 30226814, 30238981, 30248034, 30253948, 30256709; g) Product Code 0250-18-30, UDI/DI 00350770954868, Batch Numbers: 30181868, 30184756, 30185379, 30186767, 30189491, 30190697, 30194325, 30195868, 30200971, 30226813, 30228966, 30229829, 30238121, 30240450, 30244884, 30248033, 30248345, 30254744, 30259562; h) Product Code 0250-22, UDI/DI 00350770954875, Batch Numbers: 30178410, 30181843, 30181867, 30184755, 30185368, 30195736, 30192113, 30195735, 30198033, 30226667, 30226812, 30228965, 30231399, 30238120, 30238512, 30238980, 30240484, 30240297, 30240449, 30250138, 30250166, 30252990, 30253947, 30254743, 30256209, 30256481, 80402126, 80402127, 30257064, 30259561, 30261697, 30262036, 30268253, 30270502, 30270857; i) Product Code 0260-16, UDI/DI 00350770954882, Batch Numbers: 30195867, 30245069, 30258131, 30267590; j) Product Code 0260-18, UDI/DI 00350770954899, Batch Numbers: 30181880, 30185383, 30195858, 30238884, 30239856, 30244880, 30250593, 30250160, 30262027, 30290826; k) Product Code 0260-22, UDI/DI 00350770954905, Batch Numbers: 30180717, 30186745, 30236828, 30238511, 30250165, 30259560, 30265343, 30290827
Devices named in this recall
11 devices in the FDA device registry whose identifier appears in the recall's code information.
- AVANOS 95480Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall3 reports
- AVANOS 95487Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall3 reports
- AVANOS 95484Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall2 reports
- AVANOS 95486Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall1 report
- AVANOS 95481Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95482Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95483Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95485Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95488Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95489Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
- AVANOS 95490Avanos Medical, Inc. · Tubes, gastrointestinal (and accessories)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 93907).
- Z-1186-2024 · Open, ClassifiedMIC* Gastric-Jejunal Feeding Tube with ENFit Connectors: a) MIC* Gastric-Jejunal Feeding Tube Kit with ENFit Connector - Endoscopic/Radiologic Placement - 16 Fr, Product Code 82…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page